PARADIGM® REAL-time MMT-722NAP (UDI 00613994312600)Recalls & reported problems

Medtronic Minimed, Inc. · PUMP MMT-722NAP PRDGM V2.2 PL EN PR LK · Class 3

All models of PARADIGM REAL-time (95)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00613994312600), received between Jul 14, 2017 and Oct 30, 2019 — classified as 1 injury, 9 malfunctions.

The most frequently reported specific device problem is “Device Difficult to Program or Calibrate” (3 reports).

14 other recalls cover devices cleared under the same FDA submission (P980022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Difficult to Program or Calibrate3
  • Device Displays Incorrect Message3
  • No Display/Image3
  • Failure to Cycle2
  • Detachment of Device or Device Component1
  • Device Inoperable1
  • High Test Results1
  • Material Protrusion/Extrusion1

Patient problems

  • No Consequences Or Impact To Patient8
  • Hyperglycemia2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 30, 2019Obstruction of Flow; Obstruction of Flow

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD NO DELIVERY ALARM. THE BLOOD GLUCOSE WAS UNKNOWN. IT WAS REPORTED THAT THE INSULIN PUMP LOST SETTING. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9256456 ↗
  • MalfunctionOct 14, 2019No Display/Image; Device Difficult to Program or Calibrate; No Display/Image; Device Difficult to Program or Calibrate

    CUSTOMER REPORTED VIA PHONE CALL THAT KEYPAD DID NOT WORK ANYMORE AND INSULIN PUMP WENT OFF. CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN. CUSTOMER STATED THAT THE BELT CLIP CAME OFF FROM THE INSULIN PUMP. INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9183833 ↗
  • MalfunctionOct 11, 2019No Display/Image; Device Difficult to Program or Calibrate; No Display/Image; Device Difficult to Program or Calibrate

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP WAS NOT WORKING AND NONE OF THE BUTTONS WERE RESPONDING. THE CUSTOMER¿S BLOOD GLUCOSE WAS 90 MG/DL AT THE TIME OF INCIDENT. THE CUSTOMER WAS ADVICE THE INSULIN PUMP WILL NEED TO BE REPLACED. ADVICE TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO BACKUP PLAN PER HEALTH CARE PROFESSIONAL INSTRUCTIONS. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9178741 ↗
  • MalfunctionJul 3, 2019No Display/Image; Device Difficult to Program or Calibrate; No Display/Image; Device Difficult to Program or Calibrate

    THE CUSTOMER REPORTED VIA PHONE CALL THAT INSULIN PUMP HAD BLANK DISPLAY, BUTTONS WORK INTERMITTENTLY. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF INCIDENT. TROUBLESHOOTING WAS PERFORMED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8756705 ↗
  • MalfunctionMay 14, 2019Failure to Cycle; Detachment of Device or Device Component; Failure to Cycle; Detachment of Device or Device Component

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 300 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER ALSO REPORTED THAT THE INSULIN WAS SQUIRTING DURING MANUAL PRIME PROCESS. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS STICKING OUT. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WIL BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 8607077 ↗
  • MalfunctionFeb 7, 2019Failure to Cycle; Failure to Cycle

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. CUSTOMER REPORTS THEY WERE UNABLE TO CLEAR THE ALARM. CUSTOMER REPORTS PUMP DID REQUIRE REWIND. CUSTOMER NOT ABLE TO COMPLETE REWIND. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8316031 ↗
  • MalfunctionFeb 2, 2018Device Displays Incorrect Message

    THE CUSTOMER REPORTED VIA PHONE CALL THE INSULIN PUMP ALARMED MOTOR ERROR. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 170MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER STATED THAT THE INSULIN PUMP . THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS NORMAL AND THAT THE INSULIN PUMP WAS NOT EXPOSED TO HIGH MAGNETIC FIELDS AND WAS ABLE TO COMPLETE THE REWIND PROCESS. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE INSULIN PUMP IS BEING REPLACED AND IS EXPECTED TO RETURN FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7236255 ↗
  • MalfunctionDec 22, 2017High Test Results; Device Displays Incorrect Message

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 168 MG/DL. TROUBLESHOOT WAS PERFORMED FOR COMPROMISED FORCE SENSOR SYSTEM AND THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS LOOSE. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7144824 ↗
  • MalfunctionSep 25, 2017Device Inoperable; Device Displays Incorrect Message

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEIR CURRENT INSULIN PUMP HAD AN UNRESPONSIVE KEYPAD, PROMPTING THEM TO REVERT BACK TO A BACK-UP INSULIN PUMP, WHICH ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING FOR THE ALARM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE CUSTOMER REPORTED THAT THE INSULIN WAS SQUIRTING DURING MANUAL PRIME PROCESS AND THAT THEY WERE UNABLE TO EXIT THE PREPARING TO PRIME LOOP. THE DRIVE SUPPORT CAP APPEARED PROTRUDED. THE CUSTOMER WAS ASSISTED WITH BUTTON ERROR/KEYPAD ANOMALY TROUBLESHOOTING. THE CUSTOMER STATED THAT THERE WAS NO SIGNIFICANT EVENT LEADING TO THE KEYPAD ISSUES AND THAT THE CLOCK ON THE INSULIN PUMP ADVANCED WHEN OBSERVED FOR ONE MINUTE. PER BOTH TROUBLESHOOTING, THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6892339 ↗
  • InjuryJul 14, 2017Material Protrusion/Extrusion

    CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED WITH A COMPROMISED FORCE SENSOR SYSTEM DURING THE PRIMING PROCESS. THE PATIENT'S BLOOD GLUCOSE AT THE TIME OF INCIDENT WAS 240 MG/DL. THE PATIENT'S BLOOD GLUCOSE HAD ALSO GONE HIGHER THAN 400 MG/DL. THE CUSTOMER TREATED VIA LONG ACTING INSULIN. TROUBLESHOOTING WAS INITIATED FOR THE REWIND AND PRIME ISSUE. THE INSULIN PUMP WAS NOT BUMPED OR DROPPED PRIOR TO THE ALARM. THE DRIVE SUPPORT CAP WAS STICKING OUT. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROFESSIONAL'S INSTRUCTION. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 6714400 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980022. One submission often covers many models, so check the product and lot details in each record.

  • Z-1749-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
  • Z-1750-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
  • Z-1256-2025 · Jan 31, 2025 · Open, Classified
    Paradigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
  • Z-1591-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523
  • Z-1592-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523K
  • Z-1588-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722
  • Z-1587-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522
  • Z-1589-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522K
  • Z-1594-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723
  • Z-1593-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723K
  • Z-1590-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722K
  • Z-2442-2015 · Jul 22, 2015 · Terminated
    Guardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with d
  • Z-2179-2014 · Jul 21, 2014 · Terminated
    Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combinatio
  • Z-2544-2014 · Mar 13, 2014 · Terminated
    Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K

Other Sensor, glucose, invasive devices