PARADIGM® REAL-time MMT-722NAL (UDI 00613994402035)Recalls & reported problems

Medtronic Minimed, Inc. · PUMP MMT-722NAL PRDGM INS V2.4 CL EN LK · Class 3

All models of PARADIGM REAL-time (95)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a Medtronic MiniMed Paradigm insulin pump. The 43 reports mostly describe the pump showing error messages, sticking or failing to complete priming and rewind, parts loosening or detaching, and cracked or leaking components, along with shortened battery life. Seven reports mention injury, mostly high blood sugar, and no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
43
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 00613994402035), received between Jun 21, 2017 and Jul 28, 2020 — classified as 7 injuries, 36 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (13 reports).

14 other recalls cover devices cleared under the same FDA submission (P980022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Displays Incorrect Message13
  • Failure to Cycle8
  • Material Protrusion/Extrusion8
  • Detachment of Device or Device Component6
  • Fluid/Blood Leak4
  • Break3
  • Device Difficult to Program or Calibrate3
  • No Display/Image3

Patient problems

  • No Consequences Or Impact To Patient32
  • Hyperglycemia11
  • Blood Loss1
  • Hypoglycemia1
  • Nausea1

Recent adverse event reports

The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 28, 2020Mechanical Problem; Priming Problem; Mechanical Problem; Priming Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER HAD ISSUE DURING THE PRIME OR REWOUND OF INSULIN PUMP. CUSTOMER STATED THEY ARE UNABLE TO EXIT THE PREPARING TO PRIME LOOP. CUSTOMER STATED THEY DID NOT RECEIVE 2ND SERIES OF BEEPS NOR DID NUMBERS APPEAR ON THE SCREEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURN FOR ANALYSIS

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10336422 ↗
  • MalfunctionJul 28, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT DRIVE SUPPORT CAP WAS PROTRUDED AND LOOSE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10331415 ↗
  • MalfunctionApr 29, 2020Obstruction of Flow; Battery Problem; Power Problem; Obstruction of Flow; Battery Problem; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD REJECTED NEW BATTERIES, BATTERY LIFE WAS DIMINISHED FROM FIRST DAY, HAD UNEXPLAINED RANDOM NO DELIVERY ALARM AND IT WAS SLOWER AND LOUDER WHEN BATTERY WAS LOW. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10012574 ↗
  • MalfunctionApr 10, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD BUTTON ERROR. CUSTOMER STATED BUTTONS WERE NOT HELD DOWN FOR 3 MINUTES OR LONGER. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9950648 ↗
  • MalfunctionFeb 27, 2020Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP RECEIVED MOTOR ERRORS AND BATTERY LIFE HAD DIMINISHED SLIGHTLY. THE CUSTOMER ALSO REPORTED THAT THE INSULIN PUMP WAS NOT BEEN EXPOSED TO HIGH MAGNETIC FIELD. THE CUSTOMER WAS ABLE TO COMPLETE INSULIN PUMP REWIND. THE CUSTOMER REPORTED THAT THE DRIVE SUPPORT CAP WAS NORMAL. THE CUSTOMER DECLINED TROUBLESHOOTING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9767507 ↗
  • MalfunctionJan 26, 2020Failure to Cycle; Failure to Cycle

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD COMPROMISED FORCE SENSOR SYSTEM. CUSTOMER STATED THAT DRIVE SUPPORT CAP WAS NORMAL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9630562 ↗
  • MalfunctionDec 27, 2019Failure to Cycle; Detachment of Device or Device Component; Failure to Cycle; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER BLOOD GLUCOSE LEVEL WAS 295 MG/DL. IT WAS REPORTED THAT DRIVE SUPPORT CAP WAS PROTRUDED. THE INSULIN PUMP WAS NOT BUMPED OR DROPPED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO BACK UP PLAN. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 9522604 ↗
  • MalfunctionSep 17, 2019Premature Discharge of Battery; Premature Discharge of Battery

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP RECEIVED UNEXPECTED RESTART AND A COLD RESET WAS PERFORMED. THE CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN. TROUBLESHOOTING WAS DECLINED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS. -

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9074633 ↗
  • MalfunctionAug 19, 2019Failure to Cycle; Detachment of Device or Device Component; Failure to Cycle; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS LOOSE. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN PER HEALTH CARE PROFESSIONAL INSTRUCTIONS. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8905217 ↗
  • MalfunctionJul 11, 2019Failure to Cycle; Detachment of Device or Device Component; Failure to Cycle; Detachment of Device or Device Component

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD A COMPROMISED FORCE SENSOR SYSTEM ALARM AND THE DRIVE SUPPORT CAP WAS STICKING OUT. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 105 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER STATED THAT THE INSULIN PUMP WAS NEITHER BUMPED NOR DROPPED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8779483 ↗
  • MalfunctionJun 25, 2019Failure to Cycle; Failure to Cycle

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 245 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER ALSO STATE THAT THEY WERE ABLE TO CLEAR THE ALARM AND ABLE TO REWIND THE INSULIN PUMP. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8730041 ↗
  • MalfunctionJun 10, 2019Priming Problem; Priming Problem

    THE CUSTOMER REPORTED VIA CALL THAT THEY WERE STUCK REWIND PROCESS. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. CUSTOMER STATES THEY ARE UNABLE TO EXIT THE PREPARING TO PRIME LOOP. CUSTOMER STATES THEY DID NOT RECEIVE 2ND SERIES OF BEEPS NOR DID NUMBERS APPEAR ON THE SCREEN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8681563 ↗
  • MalfunctionJun 4, 2019Device Difficult to Program or Calibrate; Power Problem; Device Difficult to Program or Calibrate; Power Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEIR INSULIN PUMP HAD UP AND DOWN ARROW KEY NOT WORKING. THE CUSTOMER¿S BLOOD GLUCOSE WAS 284 MG/DL AT THE TIME OF INCIDENT. CUSTOMER REPORTS THEY GETTING THE MESSAGE OF BATTERY OUT LIMIT. CUSTOMER REPORTS THEY WERE ABLE TO CLEAR THE ALARM. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8665433 ↗
  • MalfunctionMar 15, 2019Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    IT WAS REPORTED THAT THE INSULIN PUMP HAD KEYPAD ANOMALY. THE BLOOD GLUCOSE LEVEL OF THE CUSTOMER WAS UNKNOWN. THE CUSTOMER STATED THAT KEYPAD WAS UNRESPONSIVE. THE PRODUCT WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8423561 ↗
  • MalfunctionFeb 26, 2019Obstruction of Flow; Patient Device Interaction Problem; Obstruction of Flow; Patient Device Interaction Problem

    CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD SEVERAL ISSUES. CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN AT THE TIME OF INCIDENT. CUSTOMER REPORTED THAT THE INSULIN PUMP HAD A NO DELIVERY ALARM, SCRATCHES ON THE BODY OF THE INSULIN PUMP, CRACKS ON THE RESERVOIR COMPARTMENT. CUSTOMER DECLINED TROUBLESHOOTING. INSULIN PUMP IS NOT EXPECTED TO RETURN FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8369049 ↗
  • MalfunctionJan 22, 2019Failure to Cycle; Detachment of Device or Device Component; Failure to Cycle; Detachment of Device or Device Component

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 248 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER ALSO REPORTED THAT THE INSULIN WAS SQUIRTING DURING MANUAL PRIME PROCESS. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS STICKING OUT. CUSTOMER STATES DURING SEATING THE FORCE SENSOR DID NOT DETECT THE RESERVOIR. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8265988 ↗
  • InjurySep 11, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS OVER 300MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER ALSO REPORTED THAT THE INSULIN WAS SQUIRTING DURING MANUAL PRIME PROCESS. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS LOOSE. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia; Hypoglycemia

    MAUDE report 7867419 ↗
  • MalfunctionSep 6, 2018Power Problem; Power Problem

    THE CUSTOMER REPORTED VIA PHONE THAT THE INSULIN PUMP REJECTED BATTERIES. CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN AT THE TIME OF INCIDENT. CUSTOMER STATED THAT THE BUTTONS HAD TO PRESS TWICE. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7854091 ↗
  • MalfunctionSep 6, 2018Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED VIA PHONE CALL THE INSULIN PUMP ALARMED MOTOR ERROR. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS FLUSH. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7849922 ↗
  • InjuryAug 27, 2018Patient Device Interaction Problem; Patient Device Interaction Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE LEVEL AND THEY VISITED TO EMERGENCY ROOM ON (B)(6) 2018 WITH BLOOD GLUCOSE LEVEL OF 470 MG/DL. CUSTOMER'S CURRENT BLOOD GLUCOSE LEVEL WAS 340 MG/DL. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS OVER 600 MG/DL. THE CUSTOMER WAS TREATED WITH INTRAVENOUS FLUIDS, INSULIN DRIP, INSULIN PUMP. THE DRIVE SUPPORT CAP APPEARS NORMAL. CUSTOMER HAD BEEN USING INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED HIGH BLOOD GLUCOSE EVENT. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 7820480 ↗
  • MalfunctionAug 23, 2018Failure to Cycle; Failure to Cycle

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 100 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS PROTRUDED/ LOOSE/STICKING OUT. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7811730 ↗
  • MalfunctionJul 12, 2018Failure to Cycle; Failure to Cycle

    THE CUSTOMER REPORTED VIA SOCIAL MEDIA THAT THE INSULIN PUMP HAD A COMPROMISED FORCE SENSOR SYSTEM ALARM AND INDICATED SEVERE DAMAGE IN CASING. BLOOD GLUCOSE LEVEL AT THE TIME OF THE INCIDENT WAS NOT PROVIDED. THE INSULIN PUMP WAS NOT REPLACED OR RETURNED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7680822 ↗
  • MalfunctionJun 28, 2018Device Displays Incorrect Message

    IT WAS REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL AT THE TIME OF THE INCIDENT WAS UNKNOWN. THE CUSTOMER WAS DISCONNECTED DURING PRIME. THE DRIVE SUPPORT CAP WAS NOT DAMAGED. IT WAS ADVISED TO DISCONTINUE THE USE OF THE DEVICE AND TO REVERT TO THE BACK-UP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7645218 ↗
  • InjuryJun 20, 2018Device Displays Incorrect Message

    THE CUSTOMER REPORTED VIA PHONE CALL THE INSULIN PUMP ALARMED MOTOR ERROR AND NO DELIVERY WHILE TRYING TO DELIVER A BOLUS. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 500 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS NORMAL AND THAT THE INSULIN PUMP WAS NOT EXPOSED TO HIGH MAGNETIC FIELDS AND WAS ABLE TO COMPLETE THE REWIND PROCESS. DISPLACEMENT TEST WAS PERFORMED AND PASSED. CUSTOMER HAD A HIGH BLOOD GLUCOSE OF 500 MG/DL AND TREATED WITH INSULIN PUMP. CUSTOMER DECLINED HIGH BLOOD GLUCOSE TROUBLESHOOTING. THE CUSTOMER WAS ADVISED TO MONITOR AND CALL BACK IF ISSUE RECURS. THE INSULIN PUMP IS NOT EXPECTED TO RETURN FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 7618665 ↗
  • MalfunctionJun 14, 2018Sticking

    A CUSTOMER REPORTED VIA PHONE CALL INDICATING THAT THE DRIVE SUPPORT CAP ON THEIR INSULIN PUMP WAS STICKING OUT. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 109 MG/DL AT THE TIME OF THE INCIDENT. DURING TROUBLESHOOTING, THE CUSTOMER STATED THAT THE DAMAGE OCCURRED OVER TIME. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO THEIR MEDICALLY PRESCRIBED BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7602645 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980022. One submission often covers many models, so check the product and lot details in each record.

  • Z-1750-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
  • Z-1749-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
  • Z-1256-2025 · Jan 31, 2025 · Open, Classified
    Paradigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
  • Z-1593-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723K
  • Z-1592-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523K
  • Z-1590-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722K
  • Z-1587-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522
  • Z-1591-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523
  • Z-1588-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722
  • Z-1594-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723
  • Z-1589-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522K
  • Z-2442-2015 · Jul 22, 2015 · Terminated
    Guardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with d
  • Z-2179-2014 · Jul 21, 2014 · Terminated
    Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combinatio
  • Z-2544-2014 · Mar 13, 2014 · Terminated
    Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K

Other Sensor, glucose, invasive devices

All Sensor, glucose, invasive devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page