PARADIGM® REAL-time MMT-522LNAPRecalls & reported problems

Medtronic Minimed, Inc. · PUMP MMT-522LNAP PRDGM INS V2.4 PL EN LK · Class 3

All models of PARADIGM REAL-time (95)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic MiniMed Paradigm insulin pump delivers insulin continuously through a catheter. Reports mostly describe error messages and alarms, buttons not responding, battery and power trouble, cracked parts around the reservoir, and pumping that stops or delivers no insulin. There are 37 reports, with no deaths and 9 injuries; some describe high blood sugar after insulin delivery was interrupted. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 00613994407702), received between Jun 8, 2017 and Jun 4, 2026 — classified as 9 injuries, 28 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (7 reports).

14 other recalls cover devices cleared under the same FDA submission (P980022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Displays Incorrect Message7
  • Mechanical Problem6
  • Device Difficult to Program or Calibrate5
  • Power Problem5
  • Patient Device Interaction Problem4
  • Obstruction of Flow3
  • Adverse Event Without Identified Device or Use Problem2
  • Device Operates Differently Than Expected2

Patient problems

  • No Consequences Or Impact To Patient25
  • Hyperglycemia10
  • Dyspnea1
  • Hypoglycemia1
  • Nausea1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2026Key or Button Unresponsive/not Working

    IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED RECEIVED THE BUTTON ERROR ALARM IN THE INSULIN PUMP. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-522LNAP. TROUBLESHOOTING WAS NOT PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE USE OF THE INSULIN PUMP. MMT-522LNAP WAS REQUESTED AND THE DEVICE WAS RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25390733 ↗
  • MalfunctionSep 1, 2020Mechanical Problem; Obstruction of Flow; Date/Time-Related Software Problem; Power Problem; Patient Device Interaction Problem; Mechanical Problem; Obstruction of Flow; Date/Time-Related Software Problem; Power Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD FREQUENT ALARMS, HAD UNEXPLAINED RANDOM NO DELIVERY ALARMS, HAD REJECTED NEW BATTERIES, THE CLOCK WAS SLOW, REWIND TOOK LONGER AND THE PIECES AROUND THE RESERVOIR AREA HAD CRACKED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10480836 ↗
  • MalfunctionJul 20, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD KEYPAD ANOMALY. THE BUTTONS WERE NOT RESPONDING. THE INSULIN PUMP HAD GOT WET. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10294331 ↗
  • MalfunctionJul 15, 2020Failure to Cycle; Failure to Cycle

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD COMPROMISED FORCE SENSOR SYSTEM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10274745 ↗
  • MalfunctionJul 8, 2020Mechanical Problem; Obstruction of Flow; Mechanical Problem; Obstruction of Flow

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM. CUSTOMER REPORTED THAT THE INSULIN PUMP RECEIVED NO DELIVERY ALARMS. CUSTOMER DECLINED TO TROUBLESHOOTING WITH 24 HOURS HELPLINE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10244106 ↗
  • MalfunctionJul 1, 2020Mechanical Problem; Environmental Compatibility Problem; Mechanical Problem; Environmental Compatibility Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR. CUSTOMER REPORTED THEY WERE ABLE TO CLEAR THE ALARM. CUSTOMER STATED THEY WERE ABLE TO REWIND THE INSULIN PUMP. CUSTOMER STATED DRIVE SUPPORT CAP WAS NORMAL. CUSTOMER STATED INSULIN PUMP EXPOSED TO MAGNETIC FIELD. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10217108 ↗
  • MalfunctionMar 5, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD KEYPAD ANOMALY. THE CUSTOMER STATED INSULIN PUMP'S ACT AND SPACE BUTTONS WERE NOT RESPONDING NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9791986 ↗
  • InjuryJan 23, 2020Obstruction of Flow; Obstruction of Flow

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WENT TO EMERGENCY ROOM DUE TO HIGH BLOOD GLUCOSE ON (B)(6) 2020 WITH BLOOD GLUCOSE OF 427 MG/DL. CUSTOMER TREATED WITH INSULIN SHOT 4.5 UNITS AND CURRENT BLOOD GLUCOSE LEVEL WENT DOWN TO 353 MG/DL FROM 427 MG/DL. THE CUSTOMER WAS TREATED WITH INTRAVENOUS TREATMENT WITH INSULIN IN HOSPITAL. BASED ON CUSTOMER REPORT CUSTOMER DID NOT ALLEGE THE INSULIN PUMP WAS UNDER DELIVERING. THE CUSTOMER WAS USING THE INSULIN PUMP WITHIN 48 HOURS OF HIGH BLOOD GLUCOSE EVENT. THE CUSTOMER ALSO REPORTED THAT THE INSULIN PUMP HAD NO DELIVERY ALARMS. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 9620699 ↗
  • MalfunctionAug 30, 2019Failure to Cycle; Priming Problem; Failure to Cycle; Priming Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT SHE WAS FILLING HER TUBING WHEN THE INSULIN PUMP ALARMED WITH COMPROMISED FORCE SENSOR SYSTEM ALARM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 254 MG/DL AT THE TIME OF THE INCIDENT. CUSTOMER STATED THAT THE DRIVE SUPPORT CAP APPEARED NORMAL. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO BACK UP PLAN PER HEALTH CARE PROFESSIONAL. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8948605 ↗
  • MalfunctionJul 11, 2019Insufficient Flow or Under Infusion; Patient Device Interaction Problem; Insufficient Flow or Under Infusion; Patient Device Interaction Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE CUSTOMER ALLEGING POSSIBLE INSULIN PUMP UNDER DELIVERY. THE CUSTOMER¿S HIGH BLOOD GLUCOSE WAS 300 MG/DL AT THE TIME OF INCIDENT. CUSTOMER HAD BEEN USING INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED HIGH BLOOD GLUCOSE EVENT. CUSTOMER STATED THAT THE DRIVE SUPPORT CAP APPEARS NORMAL. CUSTOMER TREATED WITH MANUAL INJECTION. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 8779945 ↗
  • MalfunctionJul 5, 2019Device Difficult to Program or Calibrate; Visual Prompts will not Clear; Device Difficult to Program or Calibrate; Visual Prompts will not Clear

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP WAS STUCK AND THE SCREEN WAS FROZEN AS WELL THE BUTTONS WERE NOT RESPONDING. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 128 MG/DL AT TIME OF INCIDENT. THE CUSTOMER STATED THAT THEY WERE UNABLE TO CLEAR THE ALARM. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. TROUBLESHOOTING WAS PERFORMED. THE INSULIN PUMP WILL BE RETURNED FOR THE ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8763484 ↗
  • MalfunctionJul 5, 2019Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD BUTTON ERROR. THE CUSTOMER¿S BLOOD GLUCOSE WAS 102 MG/DL AT THE TIME OF INCIDENT. THE CUSTOMER WAS ADVICE THE INSULIN PUMP WILL NEED TO BE REPLACED. ADVICE TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO BACKUP PLAN PER HEALTH CARE PROFESSIONAL INSTRUCTIONS. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8763252 ↗
  • MalfunctionApr 17, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD A UNEXPECTED RESTART. A COLD RESET WAS PERFORMED ERROR. THE CUSTOMER¿S BLOOD GLUCOSE WAS 90 MG/DL. TROUBLESHOOTING STEPS WERE PROVIDED FOR PUMP ERROR. CUSTOMER STATES THAT NO PUMP ERROR RECURRING DURING NORMAL PUMP USE. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8521767 ↗
  • InjuryMar 28, 2019No Display/Image; No Apparent Adverse Event; No Display/Image; No Apparent Adverse Event

    CUSTOMER REPORTED VIA PHONE CALL THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 570 MG/DL AT THE TIME OF INCIDENT. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. THE CUSTOMER WAS TREATED WITH MANUAL INJECTIONS FOR HIGH BLOOD GLUCOSE LEVEL. CUSTOMER STATED THAT THE INSULIN PUMP HAD DIED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 8459113 ↗
  • MalfunctionMar 22, 2019Mechanical Problem; Device Difficult to Program or Calibrate; Battery Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Battery Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY HAD MULTIPLE ISSUES WITH INSULIN PUMP. THE CUSTOMER BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. CUSTOMER STATED THAT THE INSULIN PUMP HAD KEYPAD RESPONSIVENESS, SHORTENED BATTERY LIFE AND INSULIN PUMP NOISY DURING REWINDS. CUSTOMER STATED THAT THE TIME CLOCK WAS ADVANCES. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8443643 ↗
  • MalfunctionFeb 1, 2019Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED VIA PHONE THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM. CUSTOMER¿S BLOOD GLUCOSE WAS 157 MG/DL AT THE TIME OF INCIDENT. INSULIN PUMP HAD A MOTOR POSITION ENCODER ERROR ALARM. DRIVE SUPPORT CAP WAS NORMAL. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8295770 ↗
  • MalfunctionDec 26, 2018Display or Visual Feedback Problem; Mechanical Problem; Power Problem; Patient Device Interaction Problem; Display or Visual Feedback Problem; Mechanical Problem; Power Problem; Patient Device Interaction Problem

    CUSTOMER REPORTED VIA PHONE CALL THAT INSULIN PUMP HAD MOTOR ERROR ALARM. CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN AT THE TIME OF INCIDENT. TROUBLESHOOTING WAS NOT PERFORMED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8195435 ↗
  • InjuryDec 20, 2018No Apparent Adverse Event; Patient Device Interaction Problem; No Apparent Adverse Event; Patient Device Interaction Problem

    CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE. THE CUSTOMER BLOOD GLUCOSE LEVEL WAS 463 MG/DL AT THE TIME OF INCIDENT. CUSTOMER STATES THE RESERVOIR WAS ABLE TO LOCK IN PLACE. CUSTOMER STATES SHE DID NOT HAVE ANY ALARMS BUT THE PUMP WAS MAKING NOISES ON THE PUMP. THE CUSTOMER DID NOT PROVIDE DETAILS REGARDING SYMPTOMS AND TREATMENT OF HIGH BLOOD GLUCOSE. CUSTOMER DECLINED TO TROUBLESHOOTING FOR THEIR HIGH BLOOD GLUCOSE LEVELS. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 8183610 ↗
  • MalfunctionOct 12, 2018Power Problem; Power Problem

    CUSTOMER REPORTED VIA PHONE CALL THAT INSULIN PUMP HAD FAILED BATTERY TEST ALARM. CUSTOMER¿S BLOOD GLUCOSE WAS 284 MG/DL AT THE TIME OF INCIDENT. PRODUCT WILL BE RETURNED FOR THE ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7959793 ↗
  • MalfunctionSep 21, 2018Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. THE PATIENT¿S BLOOD GLUCOSE LEVEL WAS 124MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER ALSO REPORTED THAT THE INSULIN WAS SQUIRTING DURING MANUAL PRIME PROCESS. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS PROTRUDED. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7897004 ↗
  • MalfunctionSep 5, 2018Power Problem; Power Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP GOT A NO POWER ALERT. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS NOT PROVIDED AT THE TIME OF THE INCIDENT. THE CUSTOMER ALSO REPORTED BATTERY ISSUES. TROUBLESHOOTING WAS NOT PROVIDED. ALERT WAS NOT CLEARED. THE INSULIN PUMP WILL NOT RETURN FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7848947 ↗
  • MalfunctionJul 31, 2018Power Problem; Power Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP GOT A FAILED BATTERY ALARM. THE BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE PRODUCT WILL NOT BE RETURNED FOR THE ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7734094 ↗
  • InjuryJun 6, 2018Device Displays Incorrect Message

    THE CUSTOMER WAS REPORTED VIA PHONE CALL THAT, THE CUSTOMER WAS HOSPITALIZED ON (B)(6) 2018 DUE TO DEHYDRATION WITH HIGH BLOOD GLUCOSE. CUSTOMER¿S BLOOD GLUCOSE AT TIME OF INCIDENT WAS 240 MG/DL AND CURRENT BLOOD GLUCOSE WAS 108 MG/DL. CUSTOMER REPORTED THE FOLLOWING SYMPTOMS RELATED TO THEIR HIGH BLOOD GLUCOSE WERE USUALLY DIFFICULTY BREATHING AND NAUSEA. CUSTOMER TREATED HIGH BLOOD GLUCOSE WITH THE INSULIN PUMP. CUSTOMER WENT TO THE EMERGENCY ROOM AS HE WAS DEHYDRATED BECAUSE HEART RATE HAD GONE UP HIGH. THOUGHT SHE WAS DEHYDRATED AND CONCERNED WITH HER HEART WAS 143 OVER 76. BLOOD GLUCOSE VALUE WHEN ADMITTED TO HOSPITAL WAS 193 MG/DL. WHEN ADMITTED AND AFTER TAKING INSULIN 2 HOURS LATER IT WAS 140 MG/DL. CUSTOMER WAS WEARING THE PUMP AT TIME OF HOSPITALIZATION. BASED ON CUSTOMER REPORT CUSTOMER DOES NOT ALLEGE PUMP WAS UNDER DELIVERING. THE DRIVE SUPPORT CAP APPEARS NORMAL. CUSTOMER STATES THAT SHE WANTS THE PUMP TO ALARM HER WHEN THE RESERVOIR IS LOW AND TO ALARM SO SHE CAN HEAR IT. AFTER REVIEWING THE PUMP SETTING HER AUDIO IS SET TO BEEP LONG AND HER LOW RES SETTINGS ARE SET TO ALARM AT 20 UNITS. CUSTOMER NOTICED THAT THE PUMP WAS NOT ALARMING WHEN INSULIN WAS LOW AND WENT TO

    Patient problems: Dyspnea; Hyperglycemia; Nausea

    MAUDE report 7572540 ↗
  • MalfunctionJun 5, 2018Device Displays Incorrect Message

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD UNEXPECTED RESTART ALARM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER REPORTED THAT THEY HAD EXPERIENCED THE UNEXPECTED ALARM. THE TROUBLESHOOTING WAS PERFORMED FOR THE INSULIN PUMP ERROR ALARM. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7572255 ↗
  • MalfunctionFeb 24, 2018Device Operates Differently Than Expected

    CUSTOMER REPORTED VIA PHONE CALL THAT THEY RECEIVED FAILED BATTERY TEST. CUSTOMER'S BLOOD GLUCOSE VALUE WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. CUSTOMER WILL BE RETURNED DEVICE FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7294155 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980022. One submission often covers many models, so check the product and lot details in each record.

  • Z-1750-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
  • Z-1749-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
  • Z-1256-2025 · Jan 31, 2025 · Open, Classified
    Paradigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
  • Z-1588-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722
  • Z-1592-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523K
  • Z-1593-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723K
  • Z-1590-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722K
  • Z-1587-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522
  • Z-1591-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523
  • Z-1594-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723
  • Z-1589-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522K
  • Z-2442-2015 · Jul 22, 2015 · Terminated
    Guardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with d
  • Z-2179-2014 · Jul 21, 2014 · Terminated
    Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combinatio
  • Z-2544-2014 · Mar 13, 2014 · Terminated
    Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K

Other Sensor, glucose, invasive devices