PARADIGM® REAL-time MMT-722LNAS (UDI 00613994409553)Recalls & reported problems

Medtronic Minimed, Inc. · PUMP MMT-722LNAS PRDGM V2.4 SK EN LK LN · Class 3

All models of PARADIGM REAL-time (95)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Medtronic MiniMed Paradigm REAL-time insulin pump has 368 reports, including 1 death and 107 injuries. Reports mostly describe screen and button problems, motor and rewind errors, and cracks, while others mention blood sugar running high or low. One open recall, begun in February 2026, concerns insulin being delivered in too large or too small amounts when the pump is raised or lowered relative to the infusion site.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
368
Deaths reported
1
Injuries reported
107
Recalls naming it
1

Summary

The FDA's MAUDE database contains 368 adverse event reports that list this exact device (DI 00613994409553), received between May 5, 2017 and May 19, 2025 — classified as 1 death, 107 injuries, 260 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (76 reports).

1 FDA recall record name this device's identifier, 1 still open.

13 other recalls cover devices cleared under the same FDA submission (P980022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 13, 2026Class IIOpen, Classified

    All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

    Z-1749-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Displays Incorrect Message76
  • Device Difficult to Program or Calibrate60
  • Mechanical Problem46
  • Adverse Event Without Identified Device or Use Problem41
  • Patient Device Interaction Problem37
  • Obstruction of Flow30
  • No Display/Image24
  • Power Problem21

Patient problems

  • No Consequences Or Impact To Patient228
  • Hyperglycemia107
  • Hypoglycemia37
  • Elevated Ketones/ Diabetic Ketoacidosis8
  • No Clinical Signs, Symptoms or Conditions3
  • Coma2
  • Death2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 25 most recent of 368 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED BUTTON ERROR. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-722LNAS. TROUBLESHOOTING WAS NOT PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. CUSTOMER DISCONTINUED USE OF THE INSULIN PUMP. MMT-722LNAS WAS REQUESTED AND CUSTOMER RESPONDED THE DEVICE WAS RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22044965 ↗
  • MalfunctionJul 18, 2022Power Problem; Power Problem

    MEDTRONIC RECEIVED INFORMATION THAT LOW BATTERY ALERT LESS THAN 1 WEEK OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15040629 ↗
  • MalfunctionApr 12, 2022Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM. CUSTOMER HAD RECEIVED MOTION SENSOR TEST FAILURE ERROR. CUSTOMER STATED THEY WERE ABLE TO REWIND THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL CONTINUE TO USE THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14080097 ↗
  • MalfunctionNov 24, 2020Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD MOTOR ERROR ALARM. THE DRIVE SUPPORT CAP WAS NORMAL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. CUSTOMER WAS ABLE TO COMPLETE INSULIN PUMP REWIND. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10896883 ↗
  • MalfunctionJul 22, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD KEYPAD ANOMALY. THE CUSTOMER WAS REPORTED THAT THE ALL BUTTONS WERE NOT RESPONDING, KEYPAD WAS ALSO ERRATIC. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10306761 ↗
  • MalfunctionJul 22, 2020No Display/Image; Mechanical Problem; Patient Device Interaction Problem; No Display/Image; Mechanical Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT SCREEN WAS NOT VISIBLE TO SEE WHAT WAS ON THE SCREEN. CUSTOMER STATED THAT IT TAKES LONGER TO REWIND. CUSTOMER STATED THAT THERE WAS CRACK ON TOP OF THE UP BUTTON AND WAVY CRACK ON THE TOP OF THE SCREEN AMOUNT OF INSULIN. CUSTOMER STATED THAT RETAINER RING WAS NOT DAMAGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10306657 ↗
  • MalfunctionJul 22, 2020Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RESERVOIR LIP RING WAS DAMAGED AND ALSO INSULIN PUMP BACK POPPED OFF. CUSTOMER STATED RESERVOIR WAS ABLE TO LOCK IN PLACE WHEN INSERTED INTO INSULIN PUMP. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10306346 ↗
  • MalfunctionJul 21, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD BUTTONS ARE NOT WORKING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10301391 ↗
  • MalfunctionJul 14, 2020Priming Problem; Priming Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS NOT PRIMING. THE CUSTOMER STATED THEY HAD LOW BLOOD GLUCOSE LEVEL AND TREATED WITH FOOD. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia; Hypoglycemia

    MAUDE report 10269928 ↗
  • InjuryJul 14, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CUSTOMER'S DAUGHTER IN LAW REPORTED VIA PHONE CALL THAT THE CUSTOMER WAS EXPERIENCING HIGH BLOOD GLUCOSE. BLOOD GLUCOSE LEVEL AT THE TIME OF THE INCIDENT WAS UNKNOWN. CUSTOMER WAS IN HOSPITAL WHICH WAS NOT RELATED TO HER DIABETES. CUSTOMER WAS HAVING UNEXPECTED HIGH BLOOD GLUCOSE OF 300MG/DL TO 400MG/DL. THE INSULIN PUMP WILL BE REPLACED, AND CUSTOMER AGREED TO RETURN THE PRODUCT FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 10269713 ↗
  • MalfunctionJul 14, 2020Mechanical Problem; Obstruction of Flow; Mechanical Problem; Obstruction of Flow

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR AND NO DELIVERY ALARMS. CUSTOMER STATED DRIVE SUPPORT CAP WAS NORMAL. CUSTOMER DECLINED TROUBLESHOOTING FOR MOTOR ERROR. CUSTOMER STATED THEY CLEARED ALARM AND INSULIN PUMP MAKING FUNNY NOISE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10269376 ↗
  • MalfunctionJul 6, 2020Premature Discharge of Battery; Device Difficult to Program or Calibrate; Premature Discharge of Battery; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE BUTTONS OF INSULIN PUMP DID NOT RESPONDING WHILE PRESSING. THE CUSTOMER STATED THAT INSULIN PUMP HAD UNEXPECTED RESTART ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10232599 ↗
  • MalfunctionJul 6, 2020Detachment of Device or Device Component; Patient Device Interaction Problem; Detachment of Device or Device Component; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT DRIVE SUPPORT CAP WAS PROTRUDING. CUSTOMER STATED THAT DRIVE SUPPORT CAP WAS DAMAGED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10231468 ↗
  • InjuryJul 1, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THE CUSTOMER WAS HOSPITALIZED DUE TO HYPOGLYCEMIA ON (B)(6) 2020. THE CUSTOMER BLOOD GLUCOSE LEVEL WAS 22 MG/DL AT THE TIME OF INCIDENCE. THE CUSTOMER WAS UNCONSCIOUS AND TAKEN TO HOSPITAL. CUSTOMER HAS BEEN USING INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED LOW BLOOD GLUCOSE EVENT. CUSTOMER STATED DRIVE SUPPORT CAP WAS PROTRUDED. THE DEVICE WILL NOT RETURNED FOR ANALYSIS.

    Patient problems: Hypoglycemia; Loss of consciousness

    MAUDE report 10222424 ↗
  • MalfunctionJul 1, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A BUTTON ERROR ALARM AND THE CUSTOMER WAS UNABLE TO CLEAR THE ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10217403 ↗
  • MalfunctionJul 1, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BUTTON ERROR MESSAGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10216964 ↗
  • MalfunctionJun 26, 2020Device Difficult to Program or Calibrate; Moisture or Humidity Problem; Device Difficult to Program or Calibrate; Moisture or Humidity Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS EXPOSED TO MOISTURE AND IT HAD KEYPAD ANOMALY. CUSTOMER STATED THAT THE KEYPAD WAS NOT RESPONSIVE AFTER THE INSULIN PUMP GOT WET. CUSTOMER STATED THE SCROLL BAR WAS CONTINUOUS MOVING WITH NO INPUT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10199299 ↗
  • MalfunctionJun 26, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP BUTTONS WERE NOT WORKING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10199141 ↗
  • MalfunctionJun 5, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD KEYPAD ANOMALY AND ACT BUTTON DID NOT WORK. CUSTOMER WAS PRESSED THE BUTTON MULTIPLE TIMES AND ISSUE WAS INTERMITTENT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10122820 ↗
  • MalfunctionMay 23, 2020Device Difficult to Program or Calibrate; Battery Problem; Priming Problem; Device Difficult to Program or Calibrate; Battery Problem; Priming Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE BUTTONS ON THE INSULIN PUMP WERE HARD TO PUSH AND HAD TO PUSH FEW TIMES TO GET WORK. CUSTOMER STATED THAT THE INSULIN WAS SQUIRTING ON PRIMING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10084509 ↗
  • InjuryMay 22, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WERE HOSPITALIZED DUE TO HYPERGLYCEMIA LEADING TO DIABETIC KETOACIDOSIS ON (B)(6) 2020 WITH BLOOD GLUCOSE OF 787 MG/DL. CUSTOMER HAD BEEN USING INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED HIGH BLOOD GLUCOSE EVENT. THE CUSTOMER REPORTED OTHER EVENTS SUCH AS VOMITING AND NAUSEA. THE OTHER BLOOD GLUCOSE LEVEL WAS 747 MG/DL. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis; Hyperglycemia; Nausea; Vomiting

    MAUDE report 10081465 ↗
  • InjuryMay 20, 2020Excess Flow or Over-Infusion; Device Difficult to Program or Calibrate; Power Problem; Excess Flow or Over-Infusion; Device Difficult to Program or Calibrate; Power Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED LOW BLOOD GLUCOSE. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 35 MG/DL, 36 MG/DL. CUSTOMER REPORTED THAT BUTTONS WERE NOT RESPONDING AND ACT AND ESC BUTTONS WERE HARD TO PRESS AT RANDOM TIMES AND EVERY ONCE IN A WHILE IT GAVE HER TOO MUCH INSULIN AND ALSO INSULIN PUMP REJECTED NEW BATTERIES. THE INSULIN PUMP AND RESERVOIR WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hypoglycemia

    MAUDE report 10073689 ↗
  • MalfunctionMay 18, 2020Premature Discharge of Battery; Battery Problem; Power Problem; Patient Device Interaction Problem; Premature Discharge of Battery; Battery Problem; Power Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MULTIPLE PUMP ERROR ALARMS. CUSTOMER REPORTED THAT THEY WERE ABLE TO CLEAR THE ALARM AND ABLE TO COMPLETE REWIND. ALARM OCCURRED SHORTLY AFTER INSERTING A NEW BATTERY. CUSTOMER ALSO REPORTED THAT THE INSULIN PUMP HAD BUTTONS ISSUE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10064352 ↗
  • MalfunctionMay 15, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE ACT BUTTON WAS NOT STICKING AND DOES NOT WORK SOMETIMES. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10060654 ↗
  • MalfunctionMay 13, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD BUTTON ERROR ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. THE CUSTOMER STATED THAT THE ACT BUTTON WAS NOT ALWAYS RESPONSIVE. THE SCREEN WAS NEVER FROZEN. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10050352 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980022. One submission often covers many models, so check the product and lot details in each record.

  • Z-1750-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
  • Z-1256-2025 · Jan 31, 2025 · Open, Classified
    Paradigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
  • Z-1594-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723
  • Z-1587-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522
  • Z-1593-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723K
  • Z-1588-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722
  • Z-1592-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523K
  • Z-1590-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722K
  • Z-1591-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523
  • Z-1589-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522K
  • Z-2442-2015 · Jul 22, 2015 · Terminated
    Guardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with d
  • Z-2179-2014 · Jul 21, 2014 · Terminated
    Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combinatio
  • Z-2544-2014 · Mar 13, 2014 · Terminated
    Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K

Other Sensor, glucose, invasive devices

All Sensor, glucose, invasive devices →