Medtronic, Inc. · LEAD 693552 US EN SI MCRD · Class 3 · Implant

All models of Medtronic Sprint Quattro Lead (67) · Sprint Quattro Secure S DXAC/DSP (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00613994446084), received between Mar 16, 2017 and Aug 7, 2025 — classified as 5 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture4
  • High impedance4
  • Inappropriate/Inadequate Shock/Stimulation2
  • Signal Artifact/Noise2
  • Adverse Event Without Identified Device or Use Problem1
  • Impedance Problem1
  • Low impedance1
  • Over-Sensing1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2
  • Shock from Patient Lead(s)2
  • Bacteremia1
  • Unspecified Infection1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO INFECTION. THE PATIENT EXPERIENCED BACTEREMIA. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bacteremia; Unspecified Infection

    MAUDE report 22738099 ↗
  • InjurySep 8, 2021Fracture; Fracture

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A POSSIBLE FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12439411 ↗
  • MalfunctionAug 9, 2021High impedance; Impedance Problem; High impedance; Impedance Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD ALERTED FOR A HIGH IMPEDANCE SPIKE ON THE RV COIL. THE LEAD ALERT WAS RESET. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12295950 ↗
  • InjuryAug 19, 2019Fracture; High impedance; Fracture; High impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED TWO HIGH IMPEDANCE MEASUREMENTS. A LEAD FRACTURE WAS SUSPECTED, AND THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8906087 ↗
  • InjuryOct 31, 2018Signal Artifact/Noise; Fracture; High impedance; Inappropriate/Inadequate Shock/Stimulation; Signal Artifact/Noise; Fracture; High impedance; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INAPPROPRIATE THERAPY DUE TO NOISE AND A CONFIRMED FRACTURE OF THE RIGHT VENTRICULAR (RV) LEAD. IT WAS NOTED THAT THE RV LEAD HAD HIGH AND UNDEFINED IMPEDANCE MEASUREMENTS. IT WAS ALSO NOTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD HIGH IMPEDANCE MEASUREMENTS. THE RV LEAD WAS EXPLANTED AND REPLACED. THE LV LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 8025391 ↗
  • MalfunctionApr 25, 2018Low impedance; Low impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD ALERTED FOR LOW PACING IMPEDANCE. THE IMPEDANCE ALERT WAS PROGRAMMED OFF AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7461569 ↗
  • InjuryMar 16, 2017Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THE PATIENT RECEIVED AN INAPPROPRIATE SHOCK. IT WAS ALSO REPORTED THERE WAS A DEVICE SENSING AND DETECTION PROBLEM AND THE LEAD NOISE ALGORITHM AND T WAVE OVER-SENSING ALGORITHM DID NOT PREVENT THE SHOCK. ADDITIONALLY, THE LEAD DISPLAYED HIGH IMPEDANCE, OVER-SENSING, NOISE AND WAS FRACTURED. IT WAS STATED THE NOISE FROM THE LEAD ¿CAUSED THE SHOCK.¿ RE-PROGRAMMING TO TURN THE DEVICE ¿OFF¿ WAS DONE. A LIFE VEST WAS PLACED ON THE PATIENT AND FOLLOWING, THE DEVICE AND LEAD WERE REMOVED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)