Sprint Quattro Secure S™ DXAC/DSP 693575 (UDI 00613994446114)Recalls & reported problems
Medtronic, Inc. · LEAD 693575 US EN SI MCRD · Class 3 · Implant
All models of Medtronic Sprint Quattro Lead (67) · Sprint Quattro Secure S DXAC/DSP (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00613994446114), received between Aug 9, 2017 and Jul 10, 2023 — classified as 7 injuries, 2 malfunctions.
The most frequently reported specific device problem is “High impedance” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- High impedance5
- Fracture3
- Over-Sensing3
- Capturing Problem2
- High Capture Threshold2
- Impedance Problem2
- Signal Artifact/Noise2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Known Impact Or Consequence To Patient4
- No Clinical Signs, Symptoms or Conditions2
- Dyspnea1
- Endocarditis1
- No Consequences Or Impact To Patient1
- Obstruction/Occlusion1
- Shock from Patient Lead(s)1
- Thrombosis/Thrombus1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryJul 10, 2023Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; Under-Sensing; Impedance Problem; High Capture Threshold; Unstable Capture Threshold; Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; Under-Sensing; Impedance Problem; High Capture Threshold; Unstable Capture Threshold
IT WAS FURTHER REPORTED THAT PRIOR TO REPROGRAMMING NOISE WAS ABLE TO BE REPRODUCED WITH ARM MOVEMENTS.
Patient problems: Dyspnea; No Clinical Signs, Symptoms or Conditions
MAUDE report 17287841 ↗InjuryNov 16, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD A POCKET INFECTION WITH ENDOCARDITIS. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT- D) SYSTEM WAS REMOVED. THE PATIENT WAS NOTED TO HAVE AN OCCLUDED/STENOSED SUPERIOR VENA CAVA (SVC). A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY AND ENDOVASCULAR MECHANICAL THROMBECTOMY WERE PERFORMED ALONG WITH STENT PLACEMENT, THE PATIENT WAS TREATED WITH ANTIBIOTICS AND A LEADLESS PACEMAKER WAS IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Endocarditis; Obstruction/Occlusion; Thrombosis/Thrombus; Unspecified Infection
MAUDE report 15808142 ↗InjuryOct 12, 2021Fracture; Capturing Problem; Fracture; Capturing Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD A THRESHOLDS RISE AFTER A TRICUSPID VALVE REPLACEMENT PROCEDURE. THE LEAD WAS SUSPECTED TO HAVE BEEN FRACTURED. A NEW PACE/SENSE LEAD WAS IMPLANTED. THE HIGH VOLTAGE PORTION OF THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12615214 ↗InjuryJun 7, 2021Signal Artifact/Noise; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Low impedance; Capturing Problem; Impedance Problem; Signal Artifact/Noise; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Low impedance; Capturing Problem; Impedance Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INAPPROPRIATE THERAPY. THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) FOR HIGH RATE NON-SUSTAINED EPISODES AND A HIGH NUMBER OF SHORT V-V INTERVALS. IN ADDITION, THE RV LEAD TRIGGERED ALERTS FOR OVERSENSING OF NOISE, HIGH PACING IMPEDANCE, AND LOW PACING IMPEDANCE. THE RV LEAD ALSO HAD AN ACUTE DECREASE IN PACING IMPEDANCE AND THRESHOLDS WERE ELEVATED. ALL VENTRICULAR FIBRILLATION (VF)/ VENTRICULAR TACHYCARDIA (VT) DETECTIONS WERE PROGRAMMED OFF. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 11947944 ↗InjuryOct 19, 2020Break; Connection Problem; High Capture Threshold; Break; Connection Problem; High Capture Threshold
IT WAS REPORTED THAT DURING A ROUTINE GENERATOR CHANGE HIGH THRESHOLDS WERE NOTED ON THE RIGHT VENTRICULAR (RV) LEAD. THE INSULATION AROUND THE RV PACE/SENSE ELECTRODE WAS "OUT OF PLACE" AND WAS UNABLE TO BE INSERTED INTO NEW DEVICE HEADER. THE PACE/SENSE PORTION OF THE LEAD WAS CAPPED AND A PREVIOUSLY CAPPED LEAD WAS UNCAPPED AND USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10697593 ↗InjuryAug 29, 2019High impedance; High impedance
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED AN ALERT FOR HIGH DEFIBRILLATION IMPEDANCE MEASUREMENT. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8946616 ↗MalfunctionOct 22, 2018High impedance; High impedance
IT WAS REPORTED THAT THE PATIENT¿S RIGHT VENTRICULAR (RV) LEAD TRIGGERED AN ALERT FOR HIGH RV COIL IMPEDANCE. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7990011 ↗InjuryMay 11, 2018Fracture; High impedance; Over-Sensing; Fracture; High impedance; Over-Sensing
IT WAS REPORTED THAT THE PATIENT WAS HEARING AUDIBLE TONES. AN INTERROGATION OF THE DEVICE WAS PERFORMED AND IT WAS FOUND THAT A LEAD INTEGRITY ALERT WAS TRIGGERED ON THE RIGHT VENTRICULAR (RV) LEAD DUE TO SIC (SENSING INTEGRITY COUNTER) AND HRNS (HIGH RATE NON-SUSTAINED) EPISODES. THERE WERE ALSO TWO ALERTS TRIGGERED FOR RV LEAD IMPEDANCE OUT OF RANGE FOR HIGH/UNDEFINED PACING IMPEDANCE. IT WAS ALSO NOTED THAT POTENTIAL OVERSENSING WAS OBSERVED ON STORED EGMS (ELECTROGRAMS). A POSSIBLE LEAD FRACTURE WAS SUSPECTED. THE RV LEAD WAS EXPLANTED AND NOT REPLACED AS THE PATIENT RECEIVED A COMPETITOR SUB Q IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7508616 ↗MalfunctionAug 9, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED THAT THE PATIENTS RIGHT VENTRICULAR (RV) LEAD HAD A "PERFORMANCE ISSUE." FOLLOW-UP WAS ATTEMPTED, BUT NO FURTHER INFORMATION WAS OBTAINED. THE LEAD HAS BEEN EXTRACTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6782704 ↗
Other Permanent defibrillator electrodes devices
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes333 deaths30,581 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M62Medtronic, Inc. · Permanent defibrillator electrodes221 deaths14,405 reports
- Capsurefix® Novus 5076-52Medtronic, Inc. · Permanent defibrillator electrodes111 deaths9,860 reports
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes80 deaths8,280 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M55Medtronic, Inc. · Permanent defibrillator electrodes116 deaths7,750 reports
- Durata™ 7122Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes151 deaths6,202 reports
- Durata™ 7122Q-65St. Jude Medical, Inc. · Permanent defibrillator electrodes86 deaths4,773 reports
- Capsurefix® Novus 5076-58Medtronic, Inc. · Permanent defibrillator electrodes47 deaths4,587 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page