Sprint Quattro Secure® 6947M62 (UDI 00643169082267)Recalls & reported problems
Medtronic, Inc. · LEAD 6947M62 DF4 ACTIVE DC US EN · Class 3 · Implant
All models of Medtronic Sprint Quattro Lead (67) · Sprint Quattro Secure (10)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Sprint Quattro Secure is a defibrillator lead, a wire that connects an implanted defibrillator to the heart. Many of its 1,154 reports describe the lead sensing wrongly, with noise and oversensing, high electrical impedance and suspected fractures. Others mention infections, shocks from the lead and, in some cases, the system being removed. The reports include 18 deaths and 699 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,157
- Deaths reported
- 18
- Injuries reported
- 702
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,157 adverse event reports that list this exact device (DI 00643169082267), received between Dec 9, 2013 and Sep 30, 2026 — classified as 18 deaths, 702 injuries, 437 malfunctions.
The most frequently reported specific device problem is “Over-Sensing” (440 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Over-Sensing440
- High impedance291
- Fracture220
- Signal Artifact/Noise211
- Impedance Problem196
- Adverse Event Without Identified Device or Use Problem175
- High Capture Threshold129
- Under-Sensing111
Patient problems
- No Clinical Signs, Symptoms or Conditions496
- No Known Impact Or Consequence To Patient311
- Unspecified Infection146
- Shock from Patient Lead(s)84
- No Consequences Or Impact To Patient49
- Endocarditis13
- Erosion12
- Bradycardia10
Recent adverse event reports
The 25 most recent of 1,157 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Fracture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954786 ↗MalfunctionSep 30, 2026Low impedance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951828 ↗InjurySep 30, 2026Appropriate Device Problem Term/Code Not Available
It was reported that the Right Ventricular (RV) lead was capped and replaced for unknown reasons. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946484 ↗InjurySep 29, 2026Signal Artifact/Noise; High impedance; Over-Sensing
It was reported that the patient presented to the emergency room with palpitations. It was noted that the Right Ventricular (RV) lead exhibited high out of range pacing impedance. A Lead Integrity Alert (LIA) triggered due to High Rate Non-Sustained episodes and Sensing Integrity Counter (SIC). It was noted that stored episodes noted RV lead oversensing of non-physiologic noise. The RV lead was capped and replaced. No further patient complications have been reported as a result of this event.
Patient problems: Palpitations
MAUDE report 50941150 ↗InjurySep 24, 2026Capturing Problem
It was reported that the Right Ventricular (RV) lead was capped due to Rising RV Thresholds and remains in the patient. A new RV lead was implanted. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50884268 ↗MalfunctionSep 23, 2026Fracture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873583 ↗InjurySep 23, 2026High impedance; Impedance Problem; Unstable Impedance
It was reported that the Right Ventricular (RV) lead was capped, and remains in the patient due to High RV Coil Impedance, Spike RV Coil Impedance, Rising RV Coil Impedance, Varying RV Coil Impedance, High SVC Coil Impedance, Spike SVC Coil Impedance and Rising SVC Coil Impedance. It was noted that a lead warning triggered for the SVC Coil and RV Coil impedance issues. It was also reported that a warning triggered for RV tip to RV coil high/undefined impedance A new RV lead was implanted. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50871856 ↗InjurySep 22, 2026Break; Failure to Capture; Over-Sensing
It was reported that the Right Ventricular (RV) lead exhibited loss of capture and oversensing. It was noted that the RV lead was damaged during a procedure. The RV lead was capped and replaced. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50861460 ↗MalfunctionAug 28, 2026Capturing Problem; Unstable Capture Threshold
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50508249 ↗InjuryAug 27, 2026Signal Artifact/Noise; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Under-Sensing
It was reported that the Right Ventricular (RV) lead exhibited undersensing, noise, oversensing resulting in false detection of arry thmia and shock being delivered. A Lead Integrity Alert (LIA) occurred for Sensing Integrity Counter (SIC) and High Rate Non-Sustained episodes. It was reported that the patient was in Coronary Artery Bypass Graft procedure during the time of the episodes and it was unclear if a magnet was placed. The lead remains in use. No further patient complications have been reported as a result of this event.
Patient problems: Shock from Patient Lead(s)
MAUDE report 50478640 ↗InjuryAug 20, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the Implantable Cardioverter Defibrillator (ICD) system was explanted due to an infection. No further patient c omplications have been reported as a result of this event.
Patient problems: Unspecified Infection
MAUDE report 50350105 ↗MalfunctionAug 19, 2026Over-Sensing
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50323494 ↗InjuryAug 4, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the Implantable Cardioverter Defibrillator (ICD) system was explanted due to an infection and erosion. A new ICD system was implanted four days later. No further patient complications have been reported as a result of this event.
Patient problems: Erosion; Unspecified Infection
MAUDE report 50132860 ↗InjuryAug 3, 2026Decreased Sensitivity; Capturing Problem; High Capture Threshold
It was reported that the Right Ventricular (RV) lead exhibited rising thresholds and diminished sensing. The RV lead was capped, and remains in the patient. A new RV lead was implanted. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50119784 ↗MalfunctionJul 27, 2026Signal Artifact/Noise; Over-Sensing; Impedance Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50038606 ↗MalfunctionJul 10, 2026Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING RESULTING IN INAPPROPRIATE PACING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25782482 ↗InjuryJul 10, 2026Signal Artifact/Noise; Over-Sensing; Unstable Impedance
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) FOR SENSING INTEGRITY COUNTER (SIC) AND PACING IMPEDANCE VARIATION. IT WAS ALSO REPORTED THAT THE LEAD EXHIBITED NOISE AND OVERSENSING. ADDITIONALLY, THE LEAD HAD A SUSPECTED FRACTURE ON THE LOW VOLTAGE TIP. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25781196 ↗DeathJul 8, 2026Defibrillation/Stimulation Problem
IT WAS REPORTED THAT THE PATIENT WITH A SINGLE-CHAMBER IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) IS DECEASED FOLLOWING CARDIAC ARREST WHERE AN EPISODE OF VENTRICULAR FIBRILLATION (VF) WAS NOT SUCCESSFULLY TERMINATED. REVIEW OF THE EPISODE AND THERAPY DATA INDICATED SIX HIGH-VOLTAGE DEFIBRILLATION SHOCKS WERE ATTEMPTED TO BE DELIVERED, HOWEVER SHORT CIRCUIT PROTECTION (SCP) WAS TRIGGERED DURING THE FIRST PHASE OF HIGH-VOLTAGE THERAPY DELIVERY FOR ALL SIX SHOCKS. THE SCP EVENTS RESULTED IN LESS THAN THE PROGRAMMED HIGH-VOLTAGE ENERGY OR NO HIGH-VOLTAGE ENERGY BEING DELIVERED WHICH FAILED TO TERMINATE THE RHYTHM. REVIEW OF HISTORICAL DEVICE DATA SHOWED NO EVIDENCE ON LEAD TRENDS SUGGESTIVE OF A RIGHT VENTRICULAR (RV) LEAD ISSUE THAT WOULD HAVE CAUSED THE SCP EVENTS. BASED ON THE AVAILABLE DATA, IT WAS NOT POSSIBLE TO DETERMINE WHICH PART OF THE ICD SYSTEM TRIGGERED SCP DURING HIGH-VOLTAGE THERAPY DELIVERY.
Patient problems: Cardiac Arrest; Ventricular Fibrillation
MAUDE report 25751974 ↗MalfunctionJun 26, 2026Decreased Sensitivity
IT WAS REPORTED THAT WHILE THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS PROGRAMMED TO MAGNETIC RESONANCE IMAGING (MRI) MODE WITH ASYNCHRONOUS PACING AND VENTRICULAR DETECTIONS DISABLED, THE PATIENT WENT INTO VENTRICULAR FIBRILLATION (VF). BEING IN MRI MODE RESULTED IN DELAY OF THERAPY AND INAPPROPRIATE PACING. THE DEVICE WAS PROGRAMMED OUT OF MRI MODE AND THE EPISODE WAS DETECTED AND SUCCESSFULLY TREATED WITH HIGH VOLTAGE THERAPY. THE ICD REMAINS IN USE. IT WAS ALSO REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) RESULTING IN UNNECESSARY PACING AT TIMES. THE RA LEAD REMAINS IN USE. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED DECREASING R-WAVE MEASUREMENTS. THE RV LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25633398 ↗InjuryJun 22, 2026Over-Sensing; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE PATIENT RECEIVED POSSIBLY INAPPROPRIATE HIGH VOLTAGE THERAPY AND PACING DUE TO RIGHT VENTRICULAR (RV) LEAD T-WAVE OVERSENSING (TWOS) THAT RESULTED IN EPISODE MISCLASSIFICATION. THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING RESULTING IN INAPPROPRIATE ATRIAL PACING. THE RV AND RA LEADS REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 25580231 ↗InjuryJun 17, 2026Fracture; High impedance; Over-Sensing
IT WAS REPORTED THAT THE HIGH VOLTAGE PORTION OF THE RIGHT VENTRICULAR (RV) COIL AND SUPERIOR VENA CAVA (SVC) COIL WERE SUSPECTED TO BE FRACTURED. THE RV LEAD EXHIBITED T-WAVE OVERSENSING (TWOS). IT WAS ALSO NOTED THAT THE RV EXHIBITED HIGH OUT OF RANGE (OOR) PACING, DEFIBRILLATION, AND SVC IMPEDANCE MEASUREMENTS. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25541275 ↗MalfunctionJun 8, 2026Fracture; High impedance
IT WAS REPORTED BY THE PATIENT THAT THEIR LEADS WERE FRACTURED. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25434223 ↗MalfunctionJun 4, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED T-WAVE OVERSENSING (TWOS). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25398739 ↗MalfunctionMay 28, 2026Signal Artifact/Noise; Over-Sensing; Under-Sensing; Low impedance; Decreased Sensitivity; Capturing Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD EXHIBITED OUT OF RANGE (OOR) LOW PACING IMPEDANCE. THE LEAD ALSO TRIGGERED LEAD INTEGRITY ALERT (LIA) FOR SENSING INTEGRITY COUNTER (SIC) AND HIGH RATE NON-SUSTAINED EPISODES AND ALSO EXHIBITED NON- PHYSIOLOGIC INTERVALS OR NOISE, DIMINISHED SENSING AND OVERSENSING THAT RESULTED IN MISCLASSIFICATION OF EPISODE TYPE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25319346 ↗MalfunctionMay 20, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED T-WAVE OVERSENSING (TWOS) POST VENTRICULAR PACING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25231835 ↗
FDA clearance and classification
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
The FDA lists this type of device as life-sustaining or life-supporting.
FDA definition of this device type: Permanent defibrillator electrodes are used with compatible implantable cardioverter defibrillators (ICD) and cardiac rhythm therapy-defibrillators (CRT-D) to sense the cardiac rhythm and to deliver high energy therapy to the heart when tachycardia is detected. Some transvenous models can also be used for pacing the right ventricle.
Common questions about the Sprint Quattro Secure® 6947M62
Has the Sprint Quattro Secure® 6947M62 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00643169082267), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Sprint Quattro Secure® 6947M62?
The FDA's MAUDE database lists 1,157 adverse event reports for this device, received between Dec 9, 2013 and Sep 30, 2026. In those reports, the most common specific device problems are “Over-Sensing” (440), “High impedance” (291) and “Fracture” (220); the most common patient problems are “Unspecified Infection” (146), “Shock from Patient Lead(s)” (84) and “Endocarditis” (13). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Sprint Quattro Secure® 6947M62?
18 reports are classified as deaths, 702 as injuries and 437 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Sprint Quattro Secure® 6947M62 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the Sprint Quattro Secure® 6947M62, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand Sprint Quattro Secure®, model 6947M62). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Permanent defibrillator electrodes devices
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes332 deaths30,477 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M62Medtronic, Inc. · Permanent defibrillator electrodes220 deaths14,369 reports
- Capsurefix® Novus 5076-52Medtronic, Inc. · Permanent defibrillator electrodes111 deaths9,852 reports
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes80 deaths8,275 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M55Medtronic, Inc. · Permanent defibrillator electrodes116 deaths7,724 reports
- Durata™ 7122Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes151 deaths6,194 reports
- Durata™ 7122Q-65St. Jude Medical, Inc. · Permanent defibrillator electrodes86 deaths4,760 reports
- Capsurefix® Novus 5076-58Medtronic, Inc. · Permanent defibrillator electrodes47 deaths4,583 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page