NA 5392 (UDI 00643169230859)Recalls & reported problems
Medtronic, Inc. · EPG 5392 MKT USA · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic 5392 external pulse generator is a pacemaker unit used outside the body. The 929 reports mostly describe connection problems, mechanical faults, blank displays, calibration issues and parts breaking, and many report no patient involvement. The reports include 2 deaths and 32 injuries, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 930
- Deaths reported
- 2
- Injuries reported
- 32
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 930 adverse event reports that list this exact device (DI 00643169230859), received between Dec 9, 2016 and Sep 23, 2026 — classified as 2 deaths, 32 injuries, 896 malfunctions.
The most frequently reported specific device problem is “Connection Problem” (370 reports).
1 other recall cover devices cleared under the same FDA submission (K132924, K150246); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Connection Problem370
- Mechanical Problem160
- No Display/Image119
- Calibration Problem91
- Break58
- Failure to Power Up43
- Display or Visual Feedback Problem42
- Component Missing35
Patient problems
- No Patient Involvement508
- No Clinical Signs, Symptoms or Conditions286
- No Known Impact Or Consequence To Patient127
- No Consequences Or Impact To Patient12
- Cardiac Arrest7
- Asystole3
- Arrhythmia2
- Bradycardia2
Recent adverse event reports
The 25 most recent of 930 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Output Problem
It was reported that the External Pulse Generator (EPG) pacing cable data was not functioning properly. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50870329 ↗MalfunctionAug 31, 2026Calibration Problem; Connection Problem
It was reported that one of the knobs of the External Pulse Generator (EPG) was physically damaged. The EPG has been returned for evaluation and subsequently tested out of specification during manufacturer's analysis. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50517822 ↗MalfunctionAug 12, 2026Calibration Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50223467 ↗MalfunctionJun 26, 2026Battery Problem; Unexpected Shutdown
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG)'S DID NOT STAY ON WHEN CHANGING BATTERY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25632608 ↗MalfunctionMay 25, 2026Use of Device Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) WAS UNABLE TO WORK PROPERLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25268480 ↗MalfunctionMay 18, 2026Output Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) FAILED PREVENTIVE MAINTENANCE TEST. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25198220 ↗MalfunctionMay 14, 2026Output above Specifications
IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE OF EXTERNAL PULSE GENERATOR (EPG) THE TOLERANCES FOR THE A TRIAL AND VENTRICULAR OUTPUT WERE MEASURING HIGH. IT WAS NOTED THAT THE BATTERY WAS DEPLETED AND REPLACED THE BATTERY WHICH RESOLVED THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25179443 ↗MalfunctionMar 30, 2026Output Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) VENTRICULAR OUTPUT WAS SET TO 20 MA, THE RANGE SHOULD BE 18-22 MA BUT IT FLUCTUATED BETWEEN 19MA AND 20MA AND THEN WENT TO 23 MA DURING PREVENTATIVE MAINTENANCE. IT WAS ALSO REPORTED THAT THERE WERE A FEW OTHER MEASUREMENTS WERE ALSO OUT OF RANGE AS WELL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24710038 ↗MalfunctionMar 12, 2026Mechanical Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) BATTERY DOOR WAS DAMAGED. IT WAS NOTED THAT THE BATTERY DOOR WAS UNABLE TO HOLD THE BATTERY SECURELY, AND THE EPG WAS UNABLE TO FUNCTION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24577829 ↗MalfunctionFeb 24, 2026Connection Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) HAD A BROKEN ATRIAL PIN CONNECTOR PORT, WHICH DID NOT HOLD THE CABLE SECURELY IN PLACE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24424965 ↗MalfunctionFeb 2, 2026Unexpected Shutdown
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) SHUT DOWN AFTER BEING POWERED ON AND DURING AN ATTEMPT TO UNLOCK IT. TROUBLESHOOTING STEPS WERE TAKEN TO REPLACE THE BATTERIES TO NO AVAIL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24231987 ↗MalfunctionJan 28, 2026No Display/Image
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) TOP SCREEN WAS UNABLE TO DISPLAY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24198356 ↗MalfunctionDec 9, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR¿S (EPG) TOP DISPLAY WAS DAMAGED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23754624 ↗MalfunctionNov 30, 2025Calibration Problem; Calibration Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR¿S (EPG) ATRIAL OUTPUT KNOB WAS NON-FUNCTIONAL. IT WAS NOTED THAT ALTHOUGH THE KNOB COULD BE TURNED, IT DID NOT CHANGE THE ATRIAL OUTPUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23677449 ↗MalfunctionNov 21, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR'S (EPG) DISPLAY WAS STARTING TO DISCOLOR. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23615067 ↗MalfunctionNov 6, 2025Connection Problem; Connection Problem
IT WAS FURTHER REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23484167 ↗MalfunctionNov 3, 2025Connection Problem; Connection Problem
THE EXTERNAL PULSE GENERATOR (EPG), WHICH WAS RETURNED AS NO LONGER REQUIRED, SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING THE MA NUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23444743 ↗MalfunctionOct 15, 2025Failure to Capture; Failure to Capture
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) WAS UNABLE TO CAPTURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23298697 ↗DeathOct 3, 2025Output Problem; Output Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT MITRAL VALVE AND TRICUSPID VALVE REPAIR SURGERY ON (B)(6) 2025. FOLLOWING THE PROCEDURE, THE PATIENT WAS CONNECTED TO AN EXTERNAL PULSE GENERATOR (EPG) TO PROVIDE PACING SUPPORT. IT WAS NOTED THAT THE EPG¿S SETTINGS WERE REPORTEDLY ESTABLISHED AND MONITORED. ON (B)(6) 2025, THE EPG MISFIRED WHILE IT WAS CONNECTED, RESULTING IN AN R-ON-T PHENOMENON AT A VULNERABLE POINT IN ITS RHYTHM WHICH LED TO A CARDIAC ARREST. IT WAS NOTED THAT THE PATIENT EXPERIENCED SEVERE PAIN, FEAR, AND PHYSICAL DISTRESS FOR MANY HOURS PRIOR TO DEATH. IT WAS REPORTED THAT THE PATIENT SUBSEQUENTLY DIED.
Patient problems: Angina; Arrhythmia; Asystole; Bradycardia; Cardiac Arrest; Discomfort; Dyspnea; Kidney injury; Muscle Weakness/Atrophy; Pulmonary Edema; Swelling/ Edema; Unspecified Infection
MAUDE report 23207092 ↗MalfunctionAug 13, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) HAD ATRIAL UNDER SENSING ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22791541 ↗MalfunctionAug 13, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) HAD ATRIAL UNDER SENSING ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22791536 ↗MalfunctionAug 13, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) HAD ATRIAL UNDER SENSING ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22791533 ↗MalfunctionAug 13, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) HAD ATRIAL UNDER SENSING ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22791527 ↗MalfunctionJul 16, 2025No Display/Image; Output Problem; No Display/Image; Output Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) EXPERIENCED INTERMITTENT LOSS OF READING OR CONNECTION DURING USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22518775 ↗MalfunctionJul 8, 2025Display or Visual Feedback Problem; Mechanical Problem; Display or Visual Feedback Problem; Mechanical Problem
IT WAS FURTHER REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22434549 ↗
FDA clearance and classification
Cleared through Special 510(k) K150246 on Feb 18, 2015 (decision: Substantially Equivalent). Submitted by Medtronic, Inc. as “Medtronic Model 5392 External Pulse Generator (EPG)”. The FDA review took 16 days. FDA record ↗
Cleared through 510(k) K132924 on Oct 31, 2013 (decision: Substantially Equivalent). Submitted by Medtronic, Inc. as “Medtronic External Pulse Generator (Epg)”. The FDA review took 43 days. FDA record ↗
The FDA regulates product code DTE (pulse-generator, pacemaker, external) under 21 CFR 870.3600 as a Class II device. 21 CFR 870.3600 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K132924, K150246. One submission often covers many models, so check the product and lot details in each record.
- Z-1356-2015 · Mar 17, 2015 · TerminatedMedtronic Dual Chamber Temporary External Pacemaker, Model 5392. Product Usage: The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual ch
Common questions about the NA 5392
Has the NA 5392 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00643169230859), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the NA 5392?
The FDA's MAUDE database lists 930 adverse event reports for this device, received between Dec 9, 2016 and Sep 23, 2026. In those reports, the most common specific device problems are “Connection Problem” (370), “Mechanical Problem” (160) and “No Display/Image” (119); the most common patient problems are “Cardiac Arrest” (7), “Asystole” (3) and “Arrhythmia” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 5392?
2 reports are classified as deaths, 32 as injuries and 896 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 5392 FDA approved?
It was cleared through 510(k) K132924 on Oct 31, 2013; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the NA 5392, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand NA, model 5392). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other PULSE-GENERATOR, PACEMAKER, EXTERNAL devices
- NA 5392Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL1 death1,934 reports
- NA 5392Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL2 deaths653 reports
- NA 5392Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL502 reports
- NA 53401Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL133 reports
- Solo Pace Control System SOLOEPG1LSolo Pace Inc. · Pulse-Generator, Pacemaker, External5 reports
- SoloPace EPG SOLOEPG1Solo Pace Inc. · Pulse-Generator, Pacemaker, External5 reports
- PRO PADZ 8900-4006Zoll Medical Corporation · PULSE-GENERATOR, PACEMAKER, EXTERNAL3 reports
- 5391Oscor Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL1 death2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page