5391Recalls & reported problems
Oscor Inc. · Single Chamber External Pacemaker · Class 2
- Adverse event reports
- 2
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169382978), received between Nov 16, 2017 and Sep 20, 2023 — classified as 1 death, 1 malfunction.
The most frequently reported specific device problem is “Device Slipped” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Device Slipped1
- Mechanical Problem1
Patient problems
- Bradycardia1
- Cardiac Arrest1
- Cardiogenic Shock1
- Death1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 20, 2023Mechanical Problem; Mechanical Problem
THE EXTERNAL PULSE GENERATOR (EPG) WHICH WAS RETURNED FOR PREVENTIVE MAINTENANCE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MAN UFACTURER¿S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17779728 ↗DeathNov 16, 2017Device Slipped
IT WAS REPORTED THE PATIENT WENT INTO ASYSTOLE WHEN THE PACING CABLES ON THE EXTERNAL PULSE GENERATOR (EPG) "POPPED OUT" AFTER THEY WERE SECURED INTO THE DEVICE. IT WAS DETERMINED THE PRONGS WERE NOT COMPLETELY TIGHTENED AND THE CABLES SLID OUT. TROUBLESHOOTING SHOWED THE PRONGS CAME LOOSE. FOLLOW UP REVEALED THE TEMPORARY LEAD WAS PLACED IN THE INTENSIVE CARE/CARDIAC CARE UNIT WHEN THE PATIENT WENT INTO SHOCK. THE PATIENT WAS TRANSFERRED TO THE CATHARIZATION LAB FOR PERMANENT PLACEMENT AND WHEN THE PATIENT WAS MOVED THE LEAD "POPPED OUT." THE PATIENT WENT INTO ASYSTOLE AND THERE WAS ELECTRO-MECHANICAL DISSOCIATION. THE LEAD WAS THEN TIGHTENED DOWN, CARDIO-PULMONARY RESUSCITATION WAS STARTED AND THE PATIENT EXPIRED 45 MINUTES TO ONE HOUR LATER. THE STATUS OF THE EPG IS NOT KNOWN AT THIS TIME. ADDITIONAL INFORMATION WAS SUBSEQUENTLY RECEIVED FROM THE PHYSICIAN THAT THE PRIMARY CAUSE OF DEATH IS BELIEVED TO BE LACTIC ACIDOSIS AND SECONDARY BRADYCARDIA. FOLLOW-UP REQUESTS FOR FURTHER INFORMATION WERE UNSUCCESSFUL.
Patient problems: Bradycardia; Cardiac Arrest; Cardiogenic Shock; Death
MAUDE report 7036844 ↗
Other PULSE-GENERATOR, PACEMAKER, EXTERNAL devices
- NA 5392Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL1 death1,934 reports
- NA 5392Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL2 deaths930 reports
- NA 5392Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL2 deaths654 reports
- NA 5392Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL504 reports
- NA 53401Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL133 reports
- SoloPace EPG SOLOEPG1Solo Pace Inc. · Pulse-Generator, Pacemaker, External5 reports
- Solo Pace Control System SOLOEPG1LSolo Pace Inc. · Pulse-Generator, Pacemaker, External5 reports
- PRO PADZ 8900-4006Zoll Medical Corporation · PULSE-GENERATOR, PACEMAKER, EXTERNAL3 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.