Skip to content

5391Recalls & reported problems

Oscor Inc. · Single Chamber External Pacemaker · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169382978), received between Nov 16, 2017 and Sep 20, 2023 — classified as 1 death, 1 malfunction.

The most frequently reported specific device problem is “Device Slipped” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Device Slipped1
  • Mechanical Problem1

Patient problems

  • Bradycardia1
  • Cardiac Arrest1
  • Cardiogenic Shock1
  • Death1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 20, 2023Mechanical Problem; Mechanical Problem

    THE EXTERNAL PULSE GENERATOR (EPG) WHICH WAS RETURNED FOR PREVENTIVE MAINTENANCE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MAN UFACTURER¿S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17779728 ↗
  • DeathNov 16, 2017Device Slipped

    IT WAS REPORTED THE PATIENT WENT INTO ASYSTOLE WHEN THE PACING CABLES ON THE EXTERNAL PULSE GENERATOR (EPG) "POPPED OUT" AFTER THEY WERE SECURED INTO THE DEVICE. IT WAS DETERMINED THE PRONGS WERE NOT COMPLETELY TIGHTENED AND THE CABLES SLID OUT. TROUBLESHOOTING SHOWED THE PRONGS CAME LOOSE. FOLLOW UP REVEALED THE TEMPORARY LEAD WAS PLACED IN THE INTENSIVE CARE/CARDIAC CARE UNIT WHEN THE PATIENT WENT INTO SHOCK. THE PATIENT WAS TRANSFERRED TO THE CATHARIZATION LAB FOR PERMANENT PLACEMENT AND WHEN THE PATIENT WAS MOVED THE LEAD "POPPED OUT." THE PATIENT WENT INTO ASYSTOLE AND THERE WAS ELECTRO-MECHANICAL DISSOCIATION. THE LEAD WAS THEN TIGHTENED DOWN, CARDIO-PULMONARY RESUSCITATION WAS STARTED AND THE PATIENT EXPIRED 45 MINUTES TO ONE HOUR LATER. THE STATUS OF THE EPG IS NOT KNOWN AT THIS TIME. ADDITIONAL INFORMATION WAS SUBSEQUENTLY RECEIVED FROM THE PHYSICIAN THAT THE PRIMARY CAUSE OF DEATH IS BELIEVED TO BE LACTIC ACIDOSIS AND SECONDARY BRADYCARDIA. FOLLOW-UP REQUESTS FOR FURTHER INFORMATION WERE UNSUCCESSFUL.

    Patient problems: Bradycardia; Cardiac Arrest; Cardiogenic Shock; Death

    MAUDE report 7036844 ↗

FDA clearance and classification

Cleared through Special 510(k) K033130 on Oct 31, 2003 (decision: Substantially Equivalent). Submitted by Osypka Medical, Inc. as “Pace 101/Pace 101h, Model 3077 Ssi Temporary Pulse Generator”. The FDA review took 31 days. FDA record ↗

The FDA regulates product code DTE (pulse-generator, pacemaker, external) under 21 CFR 870.3600 as a Class II device. 21 CFR 870.3600 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the 5391

Has the 5391 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00643169382978), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the 5391?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Nov 16, 2017 and Sep 20, 2023. In those reports, the most common specific device problems are “Device Slipped” (1) and “Mechanical Problem” (1); the most common patient problems are “Bradycardia” (1), “Cardiac Arrest” (1) and “Cardiogenic Shock” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the 5391?

1 report is classified as a death, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the 5391 FDA approved?

It was cleared through Special 510(k) K033130 on Oct 31, 2003. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the 5391, and is it still sold?

The FDA's device database lists Oscor Inc. as the company responsible for it (brand 5391, model 5391). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other PULSE-GENERATOR, PACEMAKER, EXTERNAL devices

All PULSE-GENERATOR, PACEMAKER, EXTERNAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page