Skip to content

NA 5392 (UDI 00643169508767)Recalls & reported problems

Medtronic, Inc. · EPG 5392 MKT USA UPD SCREEN · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic external pulse generator is an external pacemaker unit. Many reports describe connection problems, damaged or blank screens, broken parts and battery compartment issues, and most state that no patient was involved. The records include 1,933 reports, 59 injuries and 1 death, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,934
Deaths reported
1
Injuries reported
59
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,934 adverse event reports that list this exact device (DI 00643169508767), received between Dec 9, 2016 and Sep 11, 2026 — classified as 1 death, 59 injuries, 1,874 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (852 reports).

1 other recall cover devices cleared under the same FDA submission (K132924, K150246); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 9 reportsJan 2024: 4 reportsFeb 2024: 5 reportsMar 2024: 6 reportsApr 2024: 3 reportsMay 2024: 6 reportsJun 2024: 3 reportsJul 2024: 1 reportsAug 2024: 4 reportsSep 2024: 1 reportsOct 2024: 3 reportsNov 2024: 7 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 6 reportsMar 2025: 4 reportsApr 2025: 5 reportsMay 2025: 6 reportsJun 2025: 4 reportsJul 2025: 6 reportsAug 2025: 4 reportsSep 2025: 3 reportsOct 2025: 1 reportsNov 2025: 8 reportsDec 2025: 2 reportsJan 2026: 2 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 8 reportsMay 2026: 1 reportsJun 2026: 2 reportsJul 2026: 2 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem852
  • Mechanical Problem260
  • No Display/Image194
  • Calibration Problem179
  • Display or Visual Feedback Problem132
  • Break114
  • Component Missing79
  • Output Problem71

Patient problems

  • No Patient Involvement1,008
  • No Clinical Signs, Symptoms or Conditions716
  • No Known Impact Or Consequence To Patient189
  • No Consequences Or Impact To Patient17
  • Cardiac Arrest10
  • Ventricular Fibrillation5
  • Bradycardia3
  • Arrhythmia2

Recent adverse event reports

The 25 most recent of 1,934 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2026Device Sensing Problem

    It was reported that the External Pulse Generator (EPG) was unable to read properly. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50730141 ↗
  • MalfunctionAug 28, 2026No Display/Image

    It was reported that the External Pulse Generator (EPG)'s screen was damaged and it was noted there was deep gash on the screen. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50501369 ↗
  • MalfunctionJul 1, 2026No Display/Image; Connection Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG)'S SCREEN AND ATRIAL CONNECTOR PORT HAD AN ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25686075 ↗
  • MalfunctionJul 1, 2026No Display/Image

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG)'S SCREEN HAD AN ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25686047 ↗
  • MalfunctionJun 8, 2026Mechanical Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) REQUIRED A BATTERY DRAWER REPLACEMENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25423360 ↗
  • MalfunctionJun 5, 2026Mechanical Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) BATTERY COMPARTMENT SPRING WAS BROKEN. NO PATIENT COMPLICATIONS HAVE BEEN RE PORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25404261 ↗
  • MalfunctionMay 25, 2026Use of Device Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) WAS UNABLE TO WORK PROPERLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25268471 ↗
  • MalfunctionApr 29, 2026No Display/Image

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG)'S DISPLAY BLEED ON TOP RIGHT CORNER OF SCREEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25024523 ↗
  • MalfunctionApr 27, 2026No Display/Image

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) HAD DAMAGE ON THE INSIDE OF THE DISPLAY SCREEN. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24995389 ↗
  • MalfunctionApr 23, 2026No Display/Image

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) EXHIBITED A DISPLAY ANOMALY AT THE TOP RIGHT CORNER. THERE WAS NO PATIENT IN VOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24965428 ↗
  • MalfunctionApr 21, 2026Output Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR'S (EPG) PACING CURRENT WAS NOT WITHIN TOLERANCE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24935876 ↗
  • MalfunctionApr 21, 2026Output Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR'S (EPG) PACING CURRENT WAS NOT WITHIN TOLERANCE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24935875 ↗
  • MalfunctionApr 20, 2026Calibration Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG)'S KNOB WERE BROKEN. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24922645 ↗
  • MalfunctionApr 1, 2026Connection Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG)'S VENTRICULAR PORT WAS BROKEN. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24751516 ↗
  • InjuryApr 1, 2026No Device Output

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) WAS NOT PACING WHEN CONNECTED TO THE PATIENT, AS INDICATED ON THE MONITOR, WHILE THE EPG REMAINED POWERED ON. TROUBLESHOOTING STEPS WERE TAKEN TO REPLACE THE EPG, WHICH RESOLVED THE ISSUE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24750725 ↗
  • MalfunctionMar 12, 2026Unexpected Shutdown

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) WAS RANDOMLY TURNING OFF. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24577632 ↗
  • MalfunctionMar 6, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) DISPLAYED AN ERROR CODE DURING PREVENTATIVE MAINTENANCE TESTING. TROUBLESHOOTING STEPS WERE TAKEN INCLUDING POWERED DOWN THE UNIT TO NO AVAIL. THE EPG HAS BEEN RETURNED FOR EVALUATION AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24528418 ↗
  • MalfunctionFeb 4, 2026No Display/Image

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR'S (EPG) TOP DISPLAY WAS NOT FUNCTIONAL WHEN POWERED ON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24256238 ↗
  • MalfunctionJan 29, 2026Output Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) MEASURED OUTSIDE THE EXPECTED TOLERANCE FOR THE REFRACTORY PERIOD DURING THE PREVENTIVE MAINTENANCE CHECK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24208809 ↗
  • MalfunctionJan 26, 2026Connection Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) HAD BROKEN PLASTIC SURROUNDING OF ATRIAL AND VENTRICULAR CONNECTOR PORTS. IT WAS NOTED THAT THE LEADS WERE NOT STABLE IN THE CONNECTION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24177252 ↗
  • MalfunctionDec 23, 2025Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT THE KNOB ON THE EXTERNAL PULSE GENERATOR (EPG) WAS DETACHED, AND AFTER REATTACHMENT, THE KNOB REMAINED LOOSE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23878790 ↗
  • MalfunctionDec 19, 2025Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR¿S (EPG) ATRIAL AND VENTRICULAR CONNECTORS WERE BROKEN. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23847442 ↗
  • MalfunctionNov 27, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) HAD A BROKEN DISPLAY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23669200 ↗
  • MalfunctionNov 27, 2025Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR'S (EPG) VENTRICULAR CONNECTOR PORT WAS CRACKED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23663972 ↗
  • MalfunctionNov 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) FAILED TO REMAIN ON FOR THIRTY SECONDS WITHOUT THE BATTERY, DURING PREVENTATIVE MAINTENANCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23653439 ↗

FDA clearance and classification

Cleared through Special 510(k) K150246 on Feb 18, 2015 (decision: Substantially Equivalent). Submitted by Medtronic, Inc. as “Medtronic Model 5392 External Pulse Generator (EPG)”. The FDA review took 16 days. FDA record ↗

Cleared through 510(k) K132924 on Oct 31, 2013 (decision: Substantially Equivalent). Submitted by Medtronic, Inc. as “Medtronic External Pulse Generator (Epg)”. The FDA review took 43 days. FDA record ↗

The FDA regulates product code DTE (pulse-generator, pacemaker, external) under 21 CFR 870.3600 as a Class II device. 21 CFR 870.3600 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K132924, K150246. One submission often covers many models, so check the product and lot details in each record.

  • Z-1356-2015 · Mar 17, 2015 · Terminated
    Medtronic Dual Chamber Temporary External Pacemaker, Model 5392. Product Usage: The temporary pacemaker is used with a cardiac pacing lead system for temporary single or dual ch

Common questions about the NA 5392

Has the NA 5392 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00643169508767), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the NA 5392?

The FDA's MAUDE database lists 1,934 adverse event reports for this device, received between Dec 9, 2016 and Sep 11, 2026. In those reports, the most common specific device problems are “Connection Problem” (852), “Mechanical Problem” (260) and “No Display/Image” (194); the most common patient problems are “Cardiac Arrest” (10), “Ventricular Fibrillation” (5) and “Bradycardia” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 5392?

1 report is classified as a death, 59 as injuries and 1,874 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 5392 FDA approved?

It was cleared through 510(k) K132924 on Oct 31, 2013; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the NA 5392, and is it still sold?

The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand NA, model 5392). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other PULSE-GENERATOR, PACEMAKER, EXTERNAL devices

All PULSE-GENERATOR, PACEMAKER, EXTERNAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page