NA 53401Recalls & reported problems

Medtronic, Inc. · EPG 53401 MKT US · Class 2

Quick summary

This Medtronic external pacemaker pulse generator has 133 reports, many describing output readings out of range, intermittent operation and a blank display. One report describes the unit switching off during use, with a patient's heart rate dropping sharply. The reports mention 8 injuries and no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
133
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 133 adverse event reports that list this exact device (DI 00643169723160), received between Dec 7, 2017 and May 3, 2026 — classified as 8 injuries, 125 malfunctions.

The most frequently reported specific device problem is “Output Problem” (26 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 4 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 3 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem26
  • Electrical /Electronic Property Problem15
  • Computer Software Problem12
  • Mechanical Problem10
  • Calibration Problem8
  • Failure to Power Up8
  • Connection Problem7
  • No Display/Image7

Patient problems

  • No Clinical Signs, Symptoms or Conditions76
  • No Patient Involvement49
  • No Known Impact Or Consequence To Patient7
  • No Consequences Or Impact To Patient2
  • Asystole1
  • Bradycardia1
  • Heart Block1

Recent adverse event reports

The 25 most recent of 133 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 3, 2026Device Sensing Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) VENTRICULAR SENSITIVITY AT 10 MV WAS OUT OF TOLERANCE DURING THE CHECKOUT TEST. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25062993 ↗
  • MalfunctionApr 24, 2026Output above Specifications

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) OUTPUT MEASURED HIGHER THAN EXPECTED, AS THE ANALYZER DID NOT REGISTER BELOW 2.7 MILLIAMPERE WHEN SET TO 1 MILLIAMPERE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24979275 ↗
  • MalfunctionApr 13, 2026Output Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) WORKED INTERMITTENTLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24854881 ↗
  • MalfunctionApr 1, 2026Output Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) EXHIBITED INCONSISTENT OUTPUT DURING SERVICING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24751573 ↗
  • MalfunctionFeb 11, 2026Calibration Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) VENTRICULAR KNOB DID NOT ALWAYS CHANGE PARAMETERS WHEN TURNED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24319604 ↗
  • MalfunctionDec 17, 2025Pacing Problem; Device Sensing Problem; Pacing Problem; Device Sensing Problem

    IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE THE EXTERNAL PULSE GENERATOR (EPG) FAILED THE PULSE WIDTH TEST AT A RATE OF 30 BEATS PER MINUTE (BPM), WAS UNABLE TO PROVIDE A READING AT 30 BPM, AND ALSO FAILED THE SENSITIVITY TEST AT THE LOWEST SETTING, AS NO READING COULD BE OBTAINED ON THE TESTER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23825374 ↗
  • MalfunctionDec 4, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR¿S (EPG) DISPLAY WAS COMPLETELY BLANK AFTER IT WAS TURNED ON. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23714717 ↗
  • InjuryNov 10, 2025Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) SPONTANEOUSLY TURNED OFF WHILE IN USE WITH A PATIENT. IT WAS OBSERVED THAT THE PATIENT'S HEAR RATE WAS DROPPED TO 30 BEATS PER MINUTE (BPM). THE EPG WAS TURNED ON AND THEN SWAPPED WITH ALTERNATIVE EPG. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bradycardia

    MAUDE report 23510915 ↗
  • MalfunctionOct 21, 2025Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR¿S (EPG) OUTPUT MILLIAMPERE (MA) WAS UNABLE TO CHANGE EVEN DESPITE THE MA KNOB BEING ADJUSTED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23348810 ↗
  • MalfunctionSep 12, 2025Display or Visual Feedback Problem; Pacing Intermittently; Display or Visual Feedback Problem; Pacing Intermittently

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) HAD INTERMITTENT OUTPUT DURING TESTING. IT WAS NOTED THAT EPG DISPLAYED AN ERROR CODE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23036217 ↗
  • MalfunctionSep 3, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE EXTERNAL PULSE GENERATOR (EPG) WAS RETURNED FOR TEST AND CALIBRATION AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING THE MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22949189 ↗
  • MalfunctionApr 25, 2025Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) SHUT DOWN DURING THE SURGICAL PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21906631 ↗
  • MalfunctionMar 10, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR'S (EPG) DISPLAY WAS BLANK WHEN IT WAS TURNED ON. TROUBLESHOOTING STEPS WERE TAKEN TO INCLUDE PERFORMING A POWER CYCLE AND CHANGING TO NEW BATTERIES, HOWEVER THE ISSUE PERSISTED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21561146 ↗
  • MalfunctionMar 4, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR'S (EPG) BATTERY DOOR WAS CRACKED AND DAMAGED, CAUSING THE BATTERY CONTACT TO FALL OUT. IT WAS ALSO NOTED THAT THE EPG'S OUTER COVER WAS MISSING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21513256 ↗
  • MalfunctionFeb 4, 2025Output above Specifications; Output above Specifications

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) FAILED TESTING AT 1 MILLIAMP (MA) AS IT PRODUCED AN OUTPUT OF 2.5 MILLIAMPS (MA) AND REQUIRED CALIBRATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21304646 ↗
  • MalfunctionNov 11, 2024Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) POWERED OFF UNEXPECTEDLY A FEW MINUTES AFTER POWERING ON. THE EPG WAS CHECKED SEVERAL TIMES HOWEVER THE ISSUE PERSISTED AFTER CHANGING THE EPG BATTERIES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20651303 ↗
  • MalfunctionOct 10, 2024Calibration Problem; Calibration Problem

    THE EXTERNAL PULSE GENERATOR (EPG) WAS RETURNED FOR PREVENTIVE MAINTENANCE AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUF ACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20415171 ↗
  • MalfunctionOct 7, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE EXTERNAL PULSE GENERATOR (EPG) WHICH WAS RETURNED FOR PREVENTIVE MAINTENANCE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20381801 ↗
  • MalfunctionAug 22, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) WHICH WAS RETURNED FOR PREVENTIVE MAINTENANCE SUBSEQUENTLY TESTED OUT OF SPE CIFICATION DURING MANUFACTURER¿S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20032243 ↗
  • MalfunctionJun 4, 2024Use of Device Problem; Device Sensing Problem; Use of Device Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) HAD A SENSING ISSUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19457557 ↗
  • MalfunctionMar 28, 2024Output Problem; Output Problem

    IT WAS REPORTED THAT EXTERNAL PULSE GENERATOR'S OUTPUTS WERE NOT CONSISTENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18993414 ↗
  • MalfunctionMar 27, 2024No Device Output; No Device Output

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) WAS UNABLE TO PACE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18986864 ↗
  • MalfunctionMar 26, 2024Output Problem; Output Problem

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) WAS UNABLE TO MAINTAIN THE PROGRAMMED RATE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18977204 ↗
  • MalfunctionMar 7, 2024Electrical /Electronic Property Problem; Failure to Power Up; Electrical /Electronic Property Problem; Failure to Power Up

    IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) WAS UNABLE TO TURN ON. THE EPG WAS RETURNED FOR EVALUATION AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18855375 ↗
  • MalfunctionFeb 23, 2024No Device Output; No Device Output

    THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18765568 ↗