NA 5392 (UDI 00763000162443)Recalls & reported problems
Medtronic, Inc. · EPG 5392 USA FW/IFU UPDATE · Class 2
Quick summary
This Medtronic external pacemaker is a device that paces the heart from outside the body, and its 497 reports often describe no display, connection problems, poor heartbeat sensing, or the unit failing to power up or turning off. Patient problems listed include cardiac arrest and ventricular fibrillation, an irregular heart rhythm. There are no deaths and 31 injuries reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 504
- Deaths reported
- 0
- Injuries reported
- 32
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 504 adverse event reports that list this exact device (DI 00763000162443), received between Feb 27, 2019 and Sep 29, 2026 — classified as 32 injuries, 472 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (80 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- No Display/Image80
- Connection Problem61
- Device Sensing Problem60
- Output Problem54
- Mechanical Problem53
- Failure to Power Up49
- Calibration Problem45
- Unexpected Shutdown40
Patient problems
- No Clinical Signs, Symptoms or Conditions432
- No Patient Involvement51
- No Known Impact Or Consequence To Patient9
- Arrhythmia6
- Cardiac Arrest5
- Ventricular Fibrillation4
- Asystole3
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 25 most recent of 504 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Mechanical Problem
It was reported that the External Pulse Generator (EPG)'s battery door was broken. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927983 ↗MalfunctionSep 25, 2026Mechanical Problem
It was further noted that there was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50898452 ↗MalfunctionSep 25, 2026Failure to Capture; Device Sensing Problem
It was reported that the External Pulse Generator (EPG) and the pacing cables experienced loss of capture and sensing while in use with pacer-dependent patients during surgeries. It was noted that the cables and the EPG were used successfully in other rooms, so the issue might have been related to electromagnetic interference from new vein-harvesting equipment being used in the operating room. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50895634 ↗MalfunctionSep 25, 2026No Display/Image
It was reported that the External Pulse Generator (EPG) flickered on and off when turned on. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50894272 ↗MalfunctionSep 25, 2026Calibration Problem
It was reported that the External Pulse Generator (EPG)'s Ventricular knob was not adjusting voltage when turned. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50893814 ↗MalfunctionSep 17, 2026Mechanical Problem
It was reported that the External Pulse Generator (EPG)'s battery drawer was broken and was not able to hold the battery in place. It was noted that the EPG's functionality was limited. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50807322 ↗MalfunctionSep 8, 2026Break; Failure to Power Up
It was reported that the External Pulse Generator (EPG) was unable to turn on and Intravenous (IV) pole hanger was broken. The EPG has been returned for evaluation and subsequently tested out of specification during manufacturer's analysis. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50664820 ↗MalfunctionSep 8, 2026Output Problem
It was reported that the External Pulse Generator (EPG) was unable to read 0 -10 milliamperes during preventive checks for the output tests. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50661780 ↗MalfunctionSep 8, 2026Output Problem
It was reported that the External Pulse Generator (EPG) was unable to read 0 -10 milliamperes during preventive checks for the output tests. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50661779 ↗MalfunctionAug 31, 2026No Display/Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50519575 ↗MalfunctionAug 28, 2026Device Sensing Problem
It was reported that the External Pulse Generator (EPG) incorrectly sensed the T wave and delivered pacing output. It was noted that the sense amplitude function was not working correctly.No patient complications have been reported as a result of this event.
Patient problems: Arrhythmia; Cardiac Arrest; No Clinical Signs, Symptoms or Conditions; Ventricular Fibrillation
MAUDE report 50501939 ↗MalfunctionAug 24, 2026Calibration Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50410732 ↗InjuryAug 24, 2026Output Problem
It was reported that the External Pulse Generator (EPG) paced on the T wave while connected to the patient and that the EPG was set to ventricular demand mode (VVI) at 60 beats per minute (bpm), which was reported to have caused Torsades de Pointes. It was noted that the EPG was pacing at a rate higher than the patient's intrinsic heart rate. No further patient complications have been reported as a result of this event.
Patient problems: Arrhythmia
MAUDE report 50402777 ↗MalfunctionAug 18, 2026Mechanical Problem
It was reported that the External Pulse Generator (EPG)'s battery drawer was unable to hold the batteries in place. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50300531 ↗MalfunctionAug 16, 2026Unexpected Shutdown
It was reported that the External Pulse Generator (EPG) turned off while connected to the patient. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50268254 ↗MalfunctionAug 12, 2026Connection Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50223962 ↗MalfunctionAug 12, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50219000 ↗MalfunctionAug 11, 2026Mechanical Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50206669 ↗MalfunctionJul 8, 2026Connection Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25753099 ↗MalfunctionJun 24, 2026Device Sensing Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) FAILED THE THE SENSITIVITY TEST. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25604122 ↗MalfunctionJun 24, 2026Device Sensing Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) FAILED THE THE SENSITIVITY TEST. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25604121 ↗MalfunctionJun 11, 2026Device Sensing Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) VENTRICULAR SENSITIVITY TESTED OUT OF RANGE FOR BOTH THE 10 MILLIVOLT (MV) AND 20 MILLIVOLT (MV) SETTINGS. IT WAS NOTED THAT THE TEST RESULT AT 2 MILLIVOLT (MV) WERE IN RANGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25468848 ↗MalfunctionJun 10, 2026Crack
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG)'S SCREEN WAS CRACKED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25453015 ↗MalfunctionMay 25, 2026No Display/Image; Mechanical Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG)'S SCREEN WAS BROKEN AND THE GASKET CAME OUT OF CASE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25268591 ↗MalfunctionMay 22, 2026Unexpected Shutdown
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) RANDOMLY POWERED OFF WITH NEW BATTERIES BEING INSTALLED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25253909 ↗
Other PULSE-GENERATOR, PACEMAKER, EXTERNAL devices
- NA 5392Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL1 death1,934 reports
- NA 5392Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL2 deaths930 reports
- NA 5392Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL2 deaths654 reports
- NA 53401Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL133 reports
- SoloPace EPG SOLOEPG1Solo Pace Inc. · Pulse-Generator, Pacemaker, External5 reports
- Solo Pace Control System SOLOEPG1LSolo Pace Inc. · Pulse-Generator, Pacemaker, External5 reports
- PRO PADZ 8900-4006Zoll Medical Corporation · PULSE-GENERATOR, PACEMAKER, EXTERNAL3 reports
- 5391Oscor Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL1 death2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.