NA 5392 (UDI 00763000162443)Recalls & reported problems
Medtronic, Inc. · EPG 5392 USA FW/IFU UPDATE · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic external pacemaker is a device that paces the heart from outside the body, and its 497 reports often describe no display, connection problems, poor heartbeat sensing, or the unit failing to power up or turning off. Patient problems listed include cardiac arrest and ventricular fibrillation, an irregular heart rhythm. There are no deaths and 31 injuries reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 502
- Deaths reported
- 0
- Injuries reported
- 31
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 502 adverse event reports that list this exact device (DI 00763000162443), received between Feb 27, 2019 and Sep 29, 2026 — classified as 31 injuries, 471 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (80 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- No Display/Image80
- Connection Problem61
- Device Sensing Problem60
- Output Problem53
- Mechanical Problem52
- Failure to Power Up49
- Calibration Problem45
- Unexpected Shutdown40
Patient problems
- No Clinical Signs, Symptoms or Conditions431
- No Patient Involvement51
- No Known Impact Or Consequence To Patient9
- Arrhythmia5
- Cardiac Arrest5
- Ventricular Fibrillation4
- Asystole3
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 25 most recent of 502 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Mechanical Problem
It was reported that the External Pulse Generator (EPG)'s battery door was broken. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927983 ↗MalfunctionSep 25, 2026Mechanical Problem
It was reported that the External Pulse Generator (EPG)'s battery drawer connector was missing. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50898452 ↗MalfunctionSep 25, 2026Failure to Capture; Device Sensing Problem
It was reported that the External Pulse Generator (EPG) and the pacing cables experienced loss of capture and sensing while in use with pacer-dependent patients during surgeries. It was noted that the cables and the EPG were used successfully in other rooms, so the issue might have been related to electromagnetic interference from new vein-harvesting equipment being used in the operating room. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50895634 ↗MalfunctionSep 25, 2026No Display/Image
It was reported that the External Pulse Generator (EPG) flickered on and off when turned on. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50894272 ↗MalfunctionSep 25, 2026Calibration Problem
It was reported that the External Pulse Generator (EPG)'s Ventricular knob was not adjusting voltage when turned. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50893814 ↗MalfunctionSep 17, 2026Mechanical Problem
It was reported that the External Pulse Generator (EPG)'s battery drawer was broken and was not able to hold the battery in place. It was noted that the EPG's functionality was limited. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50807322 ↗MalfunctionSep 8, 2026Break; Failure to Power Up
It was reported that the External Pulse Generator (EPG) was unable to turn on and Intravenous (IV) pole hanger was broken. The EPG has been returned for evaluation and subsequently tested out of specification during manufacturer's analysis. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50664820 ↗MalfunctionSep 8, 2026Output Problem
It was reported that the External Pulse Generator (EPG) was unable to read 0 -10 milliamperes during preventive checks for the output tests. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50661780 ↗MalfunctionSep 8, 2026Output Problem
It was reported that the External Pulse Generator (EPG) was unable to read 0 -10 milliamperes during preventive checks for the output tests. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50661779 ↗MalfunctionAug 31, 2026No Display/Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50519575 ↗MalfunctionAug 28, 2026Device Sensing Problem
It was reported that the External Pulse Generator (EPG) incorrectly sensed the T wave and delivered pacing output. It was noted that the sense amplitude function was not working correctly.No patient complications have been reported as a result of this event.
Patient problems: Arrhythmia; Cardiac Arrest; No Clinical Signs, Symptoms or Conditions; Ventricular Fibrillation
MAUDE report 50501939 ↗MalfunctionAug 24, 2026Calibration Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50410732 ↗MalfunctionAug 16, 2026Unexpected Shutdown
It was reported that the External Pulse Generator (EPG) turned off while connected to the patient. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50268254 ↗MalfunctionAug 12, 2026Connection Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50223962 ↗MalfunctionAug 12, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50219000 ↗MalfunctionAug 11, 2026Mechanical Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50206669 ↗MalfunctionJul 8, 2026Connection Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) HAD BROKEN VENTRICULAR CONNECTOR. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25753099 ↗MalfunctionJun 24, 2026Device Sensing Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) FAILED THE THE SENSITIVITY TEST. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25604122 ↗MalfunctionJun 24, 2026Device Sensing Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) FAILED THE THE SENSITIVITY TEST. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25604121 ↗MalfunctionJun 11, 2026Device Sensing Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) VENTRICULAR SENSITIVITY TESTED OUT OF RANGE FOR BOTH THE 10 MILLIVOLT (MV) AND 20 MILLIVOLT (MV) SETTINGS. IT WAS NOTED THAT THE TEST RESULT AT 2 MILLIVOLT (MV) WERE IN RANGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25468848 ↗MalfunctionJun 10, 2026Crack
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG)'S SCREEN WAS CRACKED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25453015 ↗MalfunctionMay 25, 2026No Display/Image; Mechanical Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG)'S SCREEN WAS BROKEN AND THE GASKET CAME OUT OF CASE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25268591 ↗MalfunctionMay 22, 2026Unexpected Shutdown
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) RANDOMLY POWERED OFF WITH NEW BATTERIES BEING INSTALLED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25253909 ↗MalfunctionMay 22, 2026Crack
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) HAD TWO CRACKS ON THE LOWER SCREEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25251188 ↗MalfunctionMay 19, 2026No Device Output; Battery Problem
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG)'S BATTERIES WERE DEPLETED IN LESS THAN ONE DAY AND THE EPG SUBSEQUENTLY STOPPED PACING. THE INFORMATION REGARDING THE PROGRAMMED VALUES WAS NOT AVAILABLE. IT WAS NOTED LATER THE TESTED VALUES WERE FOUND WITHIN NOMINAL LIMITS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25213912 ↗
FDA clearance and classification
Cleared through Special 510(k) K201011 on May 13, 2020 (decision: Substantially Equivalent). Submitted by Medtronic, Inc. as “Medtronic Model 5392 External Pulse Generator (EPG)”. The FDA review took 26 days. FDA record ↗
Cleared through Special 510(k) K190825 on Apr 30, 2019 (decision: Substantially Equivalent). Submitted by Medtronic, Inc. as “Medtronic Model 5392 External Pulse Generator (EPG)”. The FDA review took 29 days. FDA record ↗
Cleared through Special 510(k) K181973 on Aug 21, 2018 (decision: Substantially Equivalent). Submitted by Medtronic, Inc. as “Medtronic Model 5392 External Pulse Generator (EPG)”. The FDA review took 28 days. FDA record ↗
The FDA regulates product code DTE (pulse-generator, pacemaker, external) under 21 CFR 870.3600 as a Class II device. 21 CFR 870.3600 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 5392
Has the NA 5392 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00763000162443), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 5392?
The FDA's MAUDE database lists 502 adverse event reports for this device, received between Feb 27, 2019 and Sep 29, 2026. In those reports, the most common specific device problems are “No Display/Image” (80), “Connection Problem” (61) and “Device Sensing Problem” (60); the most common patient problems are “Arrhythmia” (5), “Cardiac Arrest” (5) and “Ventricular Fibrillation” (4). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 5392?
None of the reports are classified as deaths, 31 as injuries and 471 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 5392 FDA approved?
It was cleared through Special 510(k) K181973 on Aug 21, 2018; its record lists 3 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the NA 5392, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand NA, model 5392). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other PULSE-GENERATOR, PACEMAKER, EXTERNAL devices
- NA 5392Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL1 death1,934 reports
- NA 5392Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL2 deaths930 reports
- NA 5392Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL2 deaths653 reports
- NA 53401Medtronic, Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL133 reports
- Solo Pace Control System SOLOEPG1LSolo Pace Inc. · Pulse-Generator, Pacemaker, External5 reports
- SoloPace EPG SOLOEPG1Solo Pace Inc. · Pulse-Generator, Pacemaker, External5 reports
- PRO PADZ 8900-4006Zoll Medical Corporation · PULSE-GENERATOR, PACEMAKER, EXTERNAL3 reports
- 5391Oscor Inc. · PULSE-GENERATOR, PACEMAKER, EXTERNAL1 death2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record