Sprint Quattro Secure S® 6935M72 (UDI 00643169309784)Recalls & reported problems
Medtronic, Inc. · LEAD 6935M72 USA DF4 EN NG ACI · Class 3 · Implant
All models of Medtronic Sprint Quattro Lead (67) · Sprint Quattro Secure S (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00643169309784), received between Apr 6, 2017 and Sep 9, 2025 — classified as 10 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (8 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Nov 8, 2021Class IITerminated
The back of some Medical Device Identification Cards provided to patients were printed with an erroneous statement stating the following: "This patient has a complete MR conditional system implanted."
Z-0390-2022 · Root cause: Error in labeling · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture8
- Over-Sensing8
- Signal Artifact/Noise6
- High impedance5
- Inappropriate/Inadequate Shock/Stimulation2
- Under-Sensing2
- Appropriate Device Problem Term/Code Not Available1
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Known Impact Or Consequence To Patient3
- Shock from Patient Lead(s)3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
InjurySep 9, 2025Signal Artifact/Noise; Intermittent Capture; Fracture; High impedance; Over-Sensing; Under-Sensing; Signal Artifact/Noise; Intermittent Capture; Fracture; High impedance; Over-Sensing; Under-Sensing
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) DEVICE DELIVERED INAPPROPRIATE THERAPIES TO THE PATIENT AND THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENT(S) AT TIMES. IT WAS FURTHER REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED LEAD INTEGRITY ALERTS (LIAS) DUE TO HIGH RATE-NON-SUSTAINED EPISODES, OVERSENSING, NON-PHYSIOLOGIC/SENSING INTEGRITY COUNTER (SIC) WITH ONE THOUSAND TWO HUNDRED NINETY-NINE SHORT V-V INTERVALS, NOISE, UNDEFINED HIGH PACING IMPEDANCE, AND UNDERSENSING SAVED ON ELECTROGRAMS AND IN ADDITION, APPEARED TO RESULT IN INTERMITTENT CAPTURE DUE TO A LEAD FRACTURE. ON THE DAY OF EXTRACTION, IMAGING AND INTERROGATION OF THE DEVICE WERE PERFORMED AND INAPPROPRIATE SENSING IN BIPOLAR AND INTEGRATED BIPOLAR MODE COULD BE OBSERVED AND WITH ARM MOVEMENT SHOWED AN INCREASE OF OVERSENSING. THE CRT-D, RA LEAD, AND RV LEAD WERE SUBSEQUENTLY REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 23006906 ↗InjuryDec 30, 2024Fracture; Fracture
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) EXHIBITED A FRACTURE IN THE COIL OF THE HIGH VOLTAGE PORTION. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21034217 ↗MalfunctionJul 5, 2023Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING WHICH WAS SUSPECTED TO BE MYOPOTENTIAL OVERSENSING. THE LEAD REMAINS IN USE, AND THE CLINICIAN WILL CONTINUE TO MONITOR THE LEAD AT THIS TIME. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17258460 ↗InjuryApr 3, 2023Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS PRESENTING WITH T-WAVE OVERSENSING (TWOS) POST-VENTRICULAR PACE. REPROGRAMMING WAS PERFORMED UNTIL THE OVERSENSING WAS COVERED UP, AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16663691 ↗InjuryAug 24, 2022Fracture; Appropriate Device Problem Term/Code Not Available; Fracture; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD FAILED AND THAT IT HAD FRACTURED. IT WAS INDICATED THAT THE LEAD HAD BEEN L ASER-EXTRACTED AND THAT THERE WAS NO FURTHER INFORMATION AVAILABLE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15284100 ↗MalfunctionMay 20, 2022Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD APPEARED TO EXHIBIT OCCASIONAL UNDERSENSING ON STORED EPISODES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14453243 ↗InjuryFeb 4, 2022Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; High Capture Threshold; Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; High Capture Threshold
IT WAS REPORTED THAT A LEAD INTEGRITY ALERT (LIA) TRIGGERED DUE TO HIGH RATE NON-SUSTAINED EPISODES AND SENSING INTEGRITY COUNTER (SIC). IN ADDITION, A LEAD NOISE WARNING TRIGGERED, AND OVERSENSING, HIGH PACING IMPEDANCE, AND HIGH THRESHOLDS WERE LATER OBSERVED. THE RIGHT VENTRICULAR (RV) LEAD WAS SUSPECTED FOR FRACTURE. THE LEAD WAS CUT, CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13454049 ↗InjuryDec 2, 2021Signal Artifact/Noise; Fracture; Over-Sensing; Signal Artifact/Noise; Fracture; Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE AND HIGH SENSING INTEGRITY COUNTER (SIC). THE LEAD WAS SUSPECTED TO BE FRACTURED AND WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12916384 ↗InjuryJun 8, 2020Signal Artifact/Noise; High impedance; Over-Sensing; Signal Artifact/Noise; High impedance; Over-Sensing
IT WAS FURTHER REPORTED THAT THE LEAD EXHIBITED NOISE AND ADDITIONAL OVERSENSING WAS RECORDED BY THE DEVICE AS NON-SUSTAINED VENTRICULAR TACHYCARDIA (VT) EPISODES. THE LEAD WAS EXPLANTED AND REPLACED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10128019 ↗InjuryFeb 5, 2020Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE PATIENT RECEIVED AN INAPPROPRIATE SHOCK DUE TO OVERSENSING OF NOISE, AND THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCES AND A POSSIBLE FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. DURING THE REPLACEMENT PROCEDURE, THE NEW RV LEAD WAS INITIALLY PLACED AND EXHIBITED A LOSS OF CAPTURE AT HIGH LEVELS AND HIGH IMPEDANCES. IT WAS SUSPECTED THAT THE LEAD HAD FRACTURED AS BLOOD WAS OBSERVED IN THE LUMEN, WHICH LED TO THE SEVERE PLACEMENT DIFFICULTY. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 9673564 ↗InjuryAug 26, 2019Fracture; High impedance; Inappropriate/Inadequate Shock/Stimulation; Impedance Problem; Fracture; High impedance; Inappropriate/Inadequate Shock/Stimulation; Impedance Problem
IT WAS FURTHER REPORTED THE RIGHT VENTRICULAR (RV) LEAD WAS EXPLANTED AND REPLACED.
Patient problems: No Known Impact Or Consequence To Patient; Shock from Patient Lead(s)
MAUDE report 8933141 ↗InjuryApr 10, 2018Signal Artifact/Noise; Break; Fracture; Over-Sensing; Signal Artifact/Noise; Break; Fracture; Over-Sensing
IT WAS REPORTED THAT A LEAD INTEGRITY ALERT (LIA) TRIGGERED FOR THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD EXHIBITED NOISE, ELEVATED SENSING INTEGRITY COUNTER (SIC), OVERSENSING, AND WAS POSSIBLY FRACTURED. THE LEAD WAS TURNED OFF AND LATER EXPLANTED AND REPLACED. DURING THE LEAD REPLACEMENT PROCEDURE, THE PHYSICIAN NOTICED BLOOD AND FLUID UNDERNEATH THE INSULATION OF THE LEAD . THE PHYSICIAN WAS UNABLE TO LOCATE WHERE THE INSULATION BREACH OCCURRED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7414912 ↗MalfunctionApr 6, 2017Positioning Failure; Positioning Failure
IT WAS REPORTED THAT THE DURING IMPLANT OF THE RIGHT VENTRICULAR (RV) LEAD, THERE WAS DIFFICULTY PLACING THE LEAD DUE TO DIFFICULT PATIENT ANATOMY. AS A RESULT THE LEAD UNDERWENT A SIGNIFICANT AMOUNT OF PHYSICAL MANIPULATION AND WHEN THE LEAD WAS FINALLY POSITIONED IN AN IDEAL LOCATION, THE HELIX WOULD NOT EXTEND. THE LEAD WAS NOT USED AND THE CASE WAS ABANDONED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6467078 ↗
Other Permanent defibrillator electrodes devices
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes333 deaths30,581 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M62Medtronic, Inc. · Permanent defibrillator electrodes221 deaths14,405 reports
- Capsurefix® Novus 5076-52Medtronic, Inc. · Permanent defibrillator electrodes111 deaths9,860 reports
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes80 deaths8,280 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M55Medtronic, Inc. · Permanent defibrillator electrodes116 deaths7,750 reports
- Durata™ 7122Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes151 deaths6,202 reports
- Durata™ 7122Q-65St. Jude Medical, Inc. · Permanent defibrillator electrodes86 deaths4,773 reports
- Capsurefix® Novus 5076-58Medtronic, Inc. · Permanent defibrillator electrodes47 deaths4,587 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page