Sprint Quattro® 6946M62Recalls & reported problems
Medtronic, Inc. · LEAD 6946M62 USA EN SPRINT DF4 PASSIVE · Class 3 · Implant
All models of Medtronic Sprint Quattro Lead (67)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Sprint Quattro 6946M62 is an implanted lead wire that connects a defibrillator or pacemaker to the heart. The 44 reports mostly describe the lead picking up signals it shouldn't (over- or under-sensing), needing more energy to pace the heart, and lead integrity or noise alerts; some patients reported palpitations, shocks from the lead, or infection that led to the device being removed. These reports include 1 death and 17 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 44
- Deaths reported
- 1
- Injuries reported
- 17
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 44 adverse event reports that list this exact device (DI 00643169335578), received between Dec 9, 2016 and Feb 16, 2026 — classified as 1 death, 17 injuries, 26 malfunctions.
The most frequently reported specific device problem is “Over-Sensing” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Over-Sensing12
- Decreased Sensitivity7
- Adverse Event Without Identified Device or Use Problem5
- High Capture Threshold5
- High impedance5
- Under-Sensing5
- Failure to Capture4
- Appropriate Device Problem Term/Code Not Available3
Patient problems
- No Clinical Signs, Symptoms or Conditions14
- No Known Impact Or Consequence To Patient14
- Unspecified Infection5
- Shock from Patient Lead(s)3
- Dyspnea2
- Arrhythmia1
- Death1
- Dysphasia1
Recent adverse event reports
The 25 most recent of 44 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 16, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING ON STORED EPISODES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24345784 ↗InjuryJan 27, 2026Signal Artifact/Noise; Failure to Capture; High impedance; Over-Sensing
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY ROOM (ER) WITH PALPITATIONS. UPON REVIEW, IT WAS NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) DUE TO INCREASED SENSING INTEGRITY COUNTER (SIC) AND HIGH RATE NON-SUSTAINED EPISODES. IN ADDITION, THE RV LEAD EXHIBITED OVERSENSING AND NOISE. THE RV LEAD ALSO TRIGGERED AN ALERT FOR UNDEFINED HIGH PACING IMPEDANCES. LOSS OF CAPTURE WAS ALSO NOTED ON THE RV LEAD AND THE RV AMPLITUDE WAS SET TOO HIGH. THE HIGH VOLTAGE THERAPY WAS REPROGRAMMED; VENTRICULAR FIBRILLATION (VF) DETECTION WAS PROGRAMMED OFF. THE RV LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Arrhythmia
MAUDE report 24189515 ↗MalfunctionAug 18, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE DEVICE WAS EMITTING AN ALERT TONE MULTIPLE TIMES A DAY. THE CADENCE OF THE ALERT WAS SUGGESTIVE OF A RIGHT VENTRICULAR (RV) LEAD INTEGRITY ALERT. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22833458 ↗MalfunctionMay 17, 2024Failure to Capture; Failure to Capture
IT WAS REPORTED THAT THE PATIENT HAD TACHYCARDIA WITH A HEART RATE OF 140 BPM. THE NURSE REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS NOT CAPTURING. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Tachycardia
MAUDE report 19345044 ↗MalfunctionJan 24, 2024Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A HIGH NUMBER OF SENSING INTEGRITY COUNTERS (SIC). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18572173 ↗MalfunctionNov 9, 2023Failure to Capture; Capturing Problem; High Capture Threshold; Unstable Capture Threshold; Failure to Capture; Capturing Problem; High Capture Threshold; Unstable Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH, VARYING AND INCREASING THRESHOLDS. IT WAS ALSO NOTED THAT THE RV LEAD CAPTURE COULD NOT BE CONFIRMED. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18103745 ↗InjuryAug 29, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO AN INFECTION. APPROXIMATELY 2 WEEKS LATER, A REPLACEMENT SYSTEM WAS IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 17644150 ↗InjuryJun 30, 2023Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD OVERSENSING CAUSING INAPPROPATE SHOCKS. REPROGRAMMING OCCURRED AND THE LEAD WAS LATER CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 17238177 ↗InjuryApr 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO AN INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 16700468 ↗MalfunctionFeb 15, 2023Decreased Sensitivity; Decreased Sensitivity
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAS DIMINISHED R-WAVE SENSING. THE LEAD REMAINS IN USE. IT WAS FURTHER REPORTED THAT THE RIGHT ATRIAL (RA) LEAD HAS UNDERSENSING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED, AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16378727 ↗MalfunctionNov 1, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT HAD A "WIRE ISSUE" AND SHORTNESS OF BREATH. THE RIGHT VENTRICULAR (RV) LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Dysphasia; Dyspnea
MAUDE report 15708040 ↗MalfunctionAug 29, 2022Under-Sensing; Under-Sensing
IT WAS REPORTED THAT UNDERSENSING WAS NOTED ON THE ATRIAL AND VENTRICULAR CHANNELS. THE RIGHT ATRIAL (RA) LEAD AND RIGHT VENTRICULAR (RV) LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15312085 ↗MalfunctionAug 18, 2022Device Sensing Problem; Unstable Capture Threshold; Device Sensing Problem; Unstable Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNSTABLE CAPTURE AND SENSING TRENDS. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15249376 ↗MalfunctionAug 11, 2022Decreased Sensitivity; Decreased Sensitivity
IT WAS REPORTED THAT CHRONICALLY LOW R WAVES WERE NOTED ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15213017 ↗MalfunctionJun 24, 2022Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THAT THE DEVICE HAS SHOWN ATRIAL UNDERSENSING. THE DEVICE IS STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE ATRIAL LEAD CONTINUED TO EXHIBIT UNDERSENSING WHICH RESULTED IN INAPPROPRIATE ATRIAL LEAD PACING AND INAPPROPRIATE RIGHT VENTRICULAR (RV) LEAD PACING. IT WAS ALSO NOTED THAT THE LEFT VENTRICULAR (LV) LEAD EXHIBITED A HIGH CAPTURE MARGIN AND IT WAS NOT PACING ENOUGH. THE LEADS REMAINS IN USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14802663 ↗MalfunctionMar 25, 2022Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD INTERMITTENT UNDERSENSING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13904300 ↗MalfunctionFeb 23, 2022High impedance; High Capture Threshold; High impedance; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD HIGH THRESHOLDS AND CHRONICALLY HIGH IMPEDANCE. IT WAS ALSO NOTED THAT THE RA LEAD WAS UNDERSENSING. NO ACTION WAS TAKEN AND THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13585425 ↗MalfunctionApr 23, 2021Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAS HIGH RATE NON-SUSTAINED EPISODES AND APPEARS TO BE OVERSENSING SEEN ON STORED ELECTROGRAMS (EGMS). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11717074 ↗MalfunctionMar 23, 2021Decreased Sensitivity; Decreased Sensitivity
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD HIGH R-WAVES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11552938 ↗InjuryDec 15, 2020Fracture; High impedance; Fracture; High impedance
IT WAS REPORTED THAT THE PATIENT CAME TO THE EMERGENCY ROOM AFTER HEARING AN ALERT. THE SUPERIOR VENA CAVA (SVC) COIL ON THE RIGHT VENTRICULAR (RV) LEAD HAD HIGH IMPEDANCE. REPROGRAMMING WAS PERFORMED TO TURN THE SVC COIL OFF. THREE DAYS LATER, AN ALERT WAS TRIGGERED FOR HIGH RV COIL IMPEDANCE. A POSSIBLE LEAD FRACTURE WAS SUSPECTED. THE SVC COIL WAS PROGRAMMED ON AGAIN BUT ALERTS WERE TURNED OFF. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11012081 ↗MalfunctionOct 8, 2020Decreased Sensitivity; Decreased Sensitivity
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD CHRONICALLY LOW R WAVES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10649899 ↗MalfunctionJul 8, 2020Over-Sensing; Over-Sensing
IT WAS FURTHER REPORTED THAT THE PATIENT'S HISTORY OF HEART FAILURE EXACERBATION, ACUTE DECOMPENSATED HEART FAILURE (ADHF), AND AN INABILITY TO TAKE PRESCRIBED MEDICATIONS WAS THE REASON THE PATIENT HAS EXPERIENCED DYSPNEA. THERE WERE SEVERAL MEDICATION ADJUSTMENTS.
Patient problems: Dyspnea
MAUDE report 10245948 ↗InjuryJun 18, 2020Signal Artifact/Noise; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Signal Artifact/Noise; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE PATIENT WAS INAPPROPRIATELY SHOCKED BY THE RIGHT VENTRICULAR (RV) LEAD FOR OVERSENSING OF NOISE, ARTIFACTS, AND T-WAVE. A FOLLOW UP WITH THE PATIENT¿S HEALTHCARE PROVIDER YIELDED THAT THE PATIENT WAS SEEN AND THE LEAD WAS ASSESS. THE LEAD CONTINUES TO BE MONITORED AND REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 10166285 ↗InjuryMay 7, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS REMOVED DUE TO POCKET INFECTION. THE PATIENT RECEIVED A LEADLESS IMPLANTABLE PULSE GENERATOR (IPG) A FEW DAYS LATER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 10031905 ↗MalfunctionFeb 5, 2020Decreased Sensitivity; Decreased Sensitivity
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED CHRONICALLY LOW R-WAVES. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9671319 ↗
Other Permanent defibrillator electrodes devices
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes333 deaths30,581 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M62Medtronic, Inc. · Permanent defibrillator electrodes221 deaths14,405 reports
- Capsurefix® Novus 5076-52Medtronic, Inc. · Permanent defibrillator electrodes111 deaths9,860 reports
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes80 deaths8,280 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M55Medtronic, Inc. · Permanent defibrillator electrodes116 deaths7,750 reports
- Durata™ 7122Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes151 deaths6,202 reports
- Durata™ 7122Q-65St. Jude Medical, Inc. · Permanent defibrillator electrodes86 deaths4,773 reports
- Capsurefix® Novus 5076-58Medtronic, Inc. · Permanent defibrillator electrodes47 deaths4,587 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page