Sprint Quattro Secure S® 6935M49 (UDI 00643169550889)Recalls & reported problems
Medtronic, Inc. · LEAD 6935M49 US DF4 NG ACI · Class 3 · Implant
All models of Medtronic Sprint Quattro Lead (67) · Sprint Quattro Secure S (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169550889), received between Apr 6, 2017 and Nov 13, 2023 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Over-Sensing” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Over-Sensing3
- Fracture2
- Defective Device1
- High impedance1
- Impedance Problem1
- Inappropriate/Inadequate Shock/Stimulation1
- Low impedance1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Clinical Signs, Symptoms or Conditions1
- Shock from Patient Lead(s)1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryNov 13, 2023Fracture; Over-Sensing; Impedance Problem; Fracture; Over-Sensing; Impedance Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS SUSPECTED TO BE FRACTURED AT THE LOW VOLTAGE PORTION OF THE LEAD AND TRIGGERED A LEAD INTEGRITY ALERT (LIA) FOR HIGH RATE NON-SUSTAINED EPISODES AND SENSING INTEGRITY COUNTER (SIC). ADDITIONALLY, THE LEAD DISPLAYED A SPIKE IN PACING IMPEDANCE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18120951 ↗MalfunctionMar 31, 2020Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAS T-WAVE OVERSENSING (TWOS) POST VENTRICULAR PACING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient; Shock from Patient Lead(s)
MAUDE report 9907590 ↗InjuryJun 25, 2018Fracture; High impedance; Over-Sensing; Low impedance; Fracture; High impedance; Over-Sensing; Low impedance
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD INTEGRITY ALERT (LIA) HAD TRIGGERED DUE TO SENSING INTEGRITY COUNTER (SIC) AND A HIGH IMPEDANCE. IT WAS ALSO NOTED THERE WAS A PACING IMPEDANCE MEASUREMENT OF 0. A FRACTURE WAS SUSPECTED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7634559 ↗MalfunctionApr 6, 2017Defective Device; Defective Device
IT WAS REPORTED THAT A RIGHT VENTRICULAR (RV) LEAD WAS NEVER USED DUE TO ITS SHAPE. THE PHYSICIAN NOTED THAT THERE WAS A CURVE AT THE PROXIMAL END OF THE COIL AND NEAR THE CONNECTOR SLEEVE WITHOUT THE USE OF A STYLET. THE LEAD WAS NEVER USED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6470703 ↗
Other Permanent defibrillator electrodes devices
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes333 deaths30,581 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M62Medtronic, Inc. · Permanent defibrillator electrodes221 deaths14,405 reports
- Capsurefix® Novus 5076-52Medtronic, Inc. · Permanent defibrillator electrodes111 deaths9,860 reports
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes80 deaths8,280 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M55Medtronic, Inc. · Permanent defibrillator electrodes116 deaths7,750 reports
- Durata™ 7122Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes151 deaths6,202 reports
- Durata™ 7122Q-65St. Jude Medical, Inc. · Permanent defibrillator electrodes86 deaths4,773 reports
- Capsurefix® Novus 5076-58Medtronic, Inc. · Permanent defibrillator electrodes47 deaths4,587 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.