Sprint Quattro Secure S® 6935M49 (UDI 00643169550889)Recalls & reported problems

Medtronic, Inc. · LEAD 6935M49 US DF4 NG ACI · Class 3 · Implant

All models of Medtronic Sprint Quattro Lead (67) · Sprint Quattro Secure S (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169550889), received between Apr 6, 2017 and Nov 13, 2023 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Over-Sensing3
  • Fracture2
  • Defective Device1
  • High impedance1
  • Impedance Problem1
  • Inappropriate/Inadequate Shock/Stimulation1
  • Low impedance1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • No Clinical Signs, Symptoms or Conditions1
  • Shock from Patient Lead(s)1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 13, 2023Fracture; Over-Sensing; Impedance Problem; Fracture; Over-Sensing; Impedance Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS SUSPECTED TO BE FRACTURED AT THE LOW VOLTAGE PORTION OF THE LEAD AND TRIGGERED A LEAD INTEGRITY ALERT (LIA) FOR HIGH RATE NON-SUSTAINED EPISODES AND SENSING INTEGRITY COUNTER (SIC). ADDITIONALLY, THE LEAD DISPLAYED A SPIKE IN PACING IMPEDANCE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18120951 ↗
  • MalfunctionMar 31, 2020Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAS T-WAVE OVERSENSING (TWOS) POST VENTRICULAR PACING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient; Shock from Patient Lead(s)

    MAUDE report 9907590 ↗
  • InjuryJun 25, 2018Fracture; High impedance; Over-Sensing; Low impedance; Fracture; High impedance; Over-Sensing; Low impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD INTEGRITY ALERT (LIA) HAD TRIGGERED DUE TO SENSING INTEGRITY COUNTER (SIC) AND A HIGH IMPEDANCE. IT WAS ALSO NOTED THERE WAS A PACING IMPEDANCE MEASUREMENT OF 0. A FRACTURE WAS SUSPECTED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7634559 ↗
  • MalfunctionApr 6, 2017Defective Device; Defective Device

    IT WAS REPORTED THAT A RIGHT VENTRICULAR (RV) LEAD WAS NEVER USED DUE TO ITS SHAPE. THE PHYSICIAN NOTED THAT THERE WAS A CURVE AT THE PROXIMAL END OF THE COIL AND NEAR THE CONNECTOR SLEEVE WITHOUT THE USE OF A STYLET. THE LEAD WAS NEVER USED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6470703 ↗