Skip to content

Medivators Hookup (REF 28511)Recalls & reported problems

Medivators Inc. · Assembly, Hookup, OLYMPUS connects MEDIVATORS reprocessing equipment to endoscopes or related accessories. · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00677964066726), received between Mar 7, 2024 and Mar 7, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Difficult to Setup or Prepare” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K102996, K172677); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 13, 2023Open, Classified

    Identification label on affected hookups could become illegible over time due to the label's ink coming off. This may potentially prevent users from being able to effectively identify the hookup and may ultimately result in postponement of patient procedures.

    Z-0617-2024 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Device Difficult to Setup or Prepare1
  • Insufficient Device Problem Information1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 7, 2024Device Difficult to Setup or Prepare; Insufficient Device Problem Information; Device Difficult to Setup or Prepare; Insufficient Device Problem Information

    THE USER FACILITY REPORTED VIA MEDWATCH REPORT # 4400480000-2023-8027 THAT, "THE PATIENT NEEDED AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) RELATED TO EPIGASTRIC PAIN, GASTRITIS WITHOUT BLEEDING, AND UNSPECIFIED ABDOMINAL PAIN. STAFF TRIED TWICE TO LOAD THE ADVANTAGE MEDIVATOR TO PREPARE SUPPLIES, BUT THE MACHINE MALFUNCTIONED AND ALARMED ON TWO SEPARATE ATTEMPTS RESULTING IN INABILITY TO PERFORM PROCEDURE. PATIENT WAS MONITORED OVER NIGHT AND DISCHARGED TO RECEIVE PROCEDURE AT AN OUTPATIENT CLINIC IN THE AREA." NO REPORT OF INJURY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18862803 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K102996, K172677. One submission often covers many models, so check the product and lot details in each record.

  • Z-2537-2024 · Jun 13, 2024 · Open, Classified
    Hookup Model/Catalog Number: 2-8-540HAN Product Description: Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor and Advantage Plus" Pass-Thru
  • Z-2538-2024 · Jun 13, 2024 · Open, Classified
    Hookup Model/Catalog Number: 2-8-540CAS Product Description: Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor and Advantage Plus" Pass-Thru
  • Z-2536-2024 · Jun 13, 2024 · Open, Classified
    Hookup Model/Catalog Number: 2-8-540 Product Description: Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor and Advantage Plus" Pass-Thru Au

Other Accessories, cleaning, for endoscope devices

All Accessories, cleaning, for endoscope devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.