Advantage Plus Pass-Thru 1-2-610.003Recalls & reported problems
Medivators Inc. · Advantage Plus Pass-Thru · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00677964095832), received between Mar 8, 2022 and May 27, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K172677); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash4
- Adverse Event Without Identified Device or Use Problem1
- Computer Operating System Problem1
- Device Handling Problem1
- Device Reprocessing Problem1
- Filtration Problem1
- Insufficient Device Problem Information1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Bruise/Contusion2
- Eye Irritation1
- Pain1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 27, 2026Leak/Splash
THE USER FACILITY REPORTED THAT DISINFECTANT LEAKED FROM THEIR ADVANTAGE PLUS PASS-THRU SUBSEQUENTLY CAUSING TWO EMPLOYEES TO EXPERIENCE EYE IRRITATION. MEDICAL TREATMENT WAS SOUGHT AND ADMINISTERED (EYE IRRIGATION).
Patient problems: Eye Irritation
MAUDE report 25305670 ↗MalfunctionMar 18, 2026Device Reprocessing Problem
THE USER FACILITY REPORTED THAT THEIR ADVANTAGE PLUS PASS-THRU WAS NOT WORKING PROPERLY. BECAUSE THE FACILITY HAD NO OTHER AVAILABLE METHODS OF REPROCESSING DEVICES, PROCEDURES WERE POSTPONED. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24627771 ↗MalfunctionMar 18, 2026Computer Operating System Problem
THE USER FACILITY REPORTED THAT THEIR ADVANTAGE PLUS PASS-THRU WAS NOT WORKING PROPERLY. BECAUSE THE FACILITY HAD NO OTHER AVAILABLE METHODS OF REPROCESSING DEVICES, PROCEDURES WERE POSTPONED. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24627765 ↗MalfunctionNov 7, 2025Filtration Problem; Filtration Problem
THE USER FACILITY REPORTED THAT THE FILTERS OF TWO ADVANTAGE PLUS PASS-THRU UNITS REQUIRE REPLACEMENT EVERY 2-3 DAYS. AS A RESULT, MULTIPLE PATIENT PROCEDURES WERE POSTPONED AS THE USER FACILITY DID NOT HAVE ALTERNATIVE REPROCESSING METHODS AVAILABLE. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23498736 ↗MalfunctionFeb 7, 2025Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT THE FILTERS OF TWO ADVANTAGE PLUS PASS-THRU UNITS CLOGGED AND GAVE A WATERFLOW ALARM. AS A RESULT, MULTIPLE PATIENT PROCEDURES WERE POSTPONED AS THE USER FACILITY DID NOT HAVE ALTERNATIVE REPROCESSING METHODS AVAILABLE. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21337340 ↗MalfunctionDec 27, 2024Mechanical Problem; Mechanical Problem
THE USER FACILITY REPORTED THAT THEIR ADVANTAGE PLUS PASS-THRU ALARMED FOR AN INTERNAL FLUID LEAK ERROR IN THE RIGHT BASIN. WHEN THE EMPLOYEE ACKNOWLEDGED THE ALARM, THE LEFT BASIN DOOR OPENED AND "SPRAYED" DISINFECTANT INTO THE ROOM SUBSEQUENTLY CONTACTING AN EMPLOYEE'S EYES. THE EMPLOYEE SOUGHT TREATMENT AT THE FACILITY'S EMERGENCY ROOM.
Patient problems: Pain
MAUDE report 21025582 ↗MalfunctionJul 11, 2024Adverse Event Without Identified Device or Use Problem; Device Handling Problem; Adverse Event Without Identified Device or Use Problem; Device Handling Problem
THE USER FACILITY REPORTED THAT THEIR ADVANTAGE PLUS PASS-THRU DOOR CLOSED AND CONTACTED AN OPERATOR'S HAND RESULTING IN A BUMP AND SLIGHT BRUISE ON THE BACK OF THEIR HAND. NO MEDICAL TREATMENT SOUGHT OR ADMINISTERED.
Patient problems: Bruise/Contusion
MAUDE report 19726552 ↗MalfunctionMar 28, 2024Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT WATER WAS LEAKING FROM THEIR ADVANTAGE PLUS PASS THRU ENDOSCOPE REPROCESSING SYSTEM INTO THE FLOOR BELOW. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18990170 ↗MalfunctionOct 6, 2022Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT THEIR ADVANTAGE PLUS PASS-THRU DOOR OPENED WHILE IN-CYCLE, SPRAYING LIQUID ALL OVER THE ROOM. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15550914 ↗MalfunctionMar 8, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
THE USER FACILITY REPORTED VIA VIGILANCE REPORT TO BFARM THAT AS AN EMPLOYEE WAS UNLOADING THEIR ADVANTAGE PLUS PASS-THRU ENDOSCOPE REPROCESSING SYSTEM THE DOOR CONTACTED THE EMPLOYEES ARM RESULTING IN TWO BRUISES. THE EMPLOYEE RETURNED TO WORK.
Patient problems: Bruise/Contusion
MAUDE report 13698353 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K172677. One submission often covers many models, so check the product and lot details in each record.
- Z-0617-2024 · Nov 13, 2023 · Open, ClassifiedHookup Accessories used with MEDIVATORS ADVANTAGE PLUS and ADVANTAGE PLUS Pass Thru Automated Endoscope Reprocessors Model # Model Description 2-8-001 HU ZERO CHANNEL NO LT 2-8-01
Other Accessories, Cleaning, For Endoscope devices
- ENDOSCOPE REPROCESSOR OER-ProOlympus Medical Systems Corp. · ACCESSORIES, CLEANING, FOR ENDOSCOPE3 recalls1,163 reports
- ENDOSCOPE REPROCESSOR OER-EliteOlympus Medical Systems Corp. · Accessories, cleaning, for endoscope4 recalls323 reports
- SINGLE USE DISTAL COVER MAJ-2315Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope2 recalls294 reports
- CONNECTING TUBE MAJ-2110Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall213 reports
- CONNECTING TUBE MAJ-2111Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall164 reports
- CONNECTING TUBE MAJ-2113Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall159 reports
- CONNECTING TUBE MAJ-2112Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall143 reports
- ENDOSCOPE REPROCESSOR OER-MINIOlympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope2 recalls11 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page