Capsure® EPI 4965-15 (UDI 00681490116336)Recalls & reported problems
Medtronic, Inc. · LEAD 4965-15 CAPSURE EPI PACING US · Class 3 · Implant
All models of Capsure EPI (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This is a Medtronic Capsure EPI 4965-15 lead, a permanent pacing wire implanted in the heart. Most of the 68 reports describe leads breaking, or failing to pace the heart properly, along with high or low electrical resistance readings; in many cases the leads were capped and replaced. The reports list 55 injuries and no deaths, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 68
- Deaths reported
- 0
- Injuries reported
- 55
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 68 adverse event reports that list this exact device (DI 00681490116336), received between Jan 25, 2017 and Apr 23, 2026 — classified as 55 injuries, 13 malfunctions.
The most frequently reported specific device problem is “Fracture” (34 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture34
- High impedance20
- Failure to Capture16
- High Capture Threshold16
- Low impedance6
- Electrical /Electronic Property Problem5
- Impedance Problem5
- Break3
Patient problems
- No Known Impact Or Consequence To Patient37
- No Clinical Signs, Symptoms or Conditions19
- No Consequences Or Impact To Patient3
- Syncope3
- Arrhythmia2
- Device Overstimulation of Tissue2
- Bradycardia1
- Cardiomyopathy1
Recent adverse event reports
The 25 most recent of 68 reports, in the reporter's own words as published by the FDA.
InjuryApr 23, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A FRACTURE, HIGH THRESHOLDS, LOSS OF CAPTURE AND TRIGGERED AN ALERT FOR OUT OF RANGE (OOR) HIGH IMPEDANCE. THE RIGHT ATRIAL (RA) LEAD WAS NOTED TO BE UNUSABLE. THE RA LEAD WAS PROGRAMMED OFF, CAPPED AND LATER EXPLANTED AND REPLACED. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24963646 ↗InjuryJun 4, 2025Failure to Capture; Failure to Capture
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) AND RIGHT VENTRICULAR (RV) EPICARDIAL PACING LEADS EXHIBITED NO CAPTURE. THE LEADS WERE INACTIVATED AND REMAIN IN THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22144280 ↗InjuryJun 4, 2025Failure to Capture; Failure to Capture
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) AND RIGHT VENTRICULAR (RV) EPICARDIAL PACING LEADS EXHIBITED NO CAPTURE. THE LEADS WERE INACTIVATED AND REMAIN IN THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22144255 ↗InjuryMay 2, 2025Failure to Capture; Electrical /Electronic Property Problem; High impedance; Failure to Capture; Electrical /Electronic Property Problem; High impedance
IT WAS FURTHER REPORTED THAT THE HIGH OUT OF RANGE IMPEDANCE ON THE RV LEAD OCCURRED PRIOR TO PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21947971 ↗MalfunctionMay 2, 2025Electrical /Electronic Property Problem; Impedance Problem; Electrical /Electronic Property Problem; Impedance Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCES, NO CAPTURE AND A POLARITY SWITCH. A LEAD IMPEDANCE WARNING AND POLARITY SWITCH WAS ALSO OBSERVED ON THE RIGHT ATRIAL (RA) LEAD. IT WAS NOTED THAT THE RA LEAD FAILED AN ATRIAL LEAD POSITION CHECK. THE RV LEAD WAS CAPPED AND REPLACED. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21947964 ↗InjuryMar 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS FURTHER REPORTED THAT THE EPICARDIAL RIGHT ATRIAL (RA) LEAD WAS CAPPED AND REMAINS IN THE PATIENT. THE IMPLANTABLE PULSE GENERATOR (IPG) ALSO EXHIBITED NORMAL ELECTIVE REPLACEMENT INDICATOR (ERI).
Patient problems: Device Overstimulation of Tissue
MAUDE report 21571951 ↗MalfunctionJun 21, 2024Low impedance; Low impedance
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LEAD WARNING. THE IMPEDANCE WAS NOTED TO BE CHRONICALLY LOW. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19582269 ↗MalfunctionApr 24, 2024Failure to Capture; Fracture; Failure to Capture; Fracture
IT WAS REPORTED THAT THE DURING A GENERATOR CHANGE OUT WHEN ATTEMPTING TO CONNECT THE RIGHT ATRIAL (RA) LEAD TO THE REPLACEMENT IMPLANTABLE PULSE GENERATOR (IPG) NO CLICKS WERE AUDIBLE AND THE LEAD WAS EASILY PULLED OUT OF THE HEADER. THE LEAD WAS TAKEN OUT OF THE HEADER, AND THE SET SCREW WAS VISUALLY CONFIRMED TO NOT BE OBSTRUCTING THE PORT. THE LEAD WAS THEN RE-INSERTED INTO THE HEADER WHILE THE WRENCH WAS HELD IN PLACE TO VENT ANY TRAPPED AIR. AGAIN, NO WRENCH CLICKS WERE HEARD AND THE LEAD WAS NOT SECURED. THE IPG WAS REMOVED AND REPLACED. WHEN WRENCHING THE RA LEAD INTO THE REPLACEMENT IPG WRENCH CLICKS WERE HEARD IMMEDIATELY AND THE LEAD WAS SECURED. A VISIBLE FRACTURE WAS OBSERVED AT THE PIN PLUG JUNCTION TO THE MAIN LEAD BODY THAT HAD BEEN SUSPECTED DUE TO NO CAPTURE OF THE ATRIAL LEAD. THE IPG WAS ATTEMPTED NOT USED, REMOVED AND REPLACED. THE RIGHT ATRIAL (RA) LEAD REMAINS IN USE FOR SENSING PURPOSES. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19173886 ↗InjuryMar 7, 2024Fracture; High impedance; Fracture; High impedance
IT WAS REPORTED THAT THE EPICARDIAL RIGHT ATRIAL (RA) LEAD HAD A SUSPECTED FRACTURE. THE RA LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18862675 ↗InjuryNov 22, 2023High Capture Threshold; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18189696 ↗InjuryJan 6, 2023Fracture; Fracture
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A FRACTURE. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16110898 ↗InjuryNov 23, 2022Fracture; Fracture
IT WAS REPORTED THAT THE EPICARDIAL RIGHT ATRIAL (RA) LEAD AND RIGHT VENTRICULAR (RV) LEAD EXHIBITED A FRACTURE. THE RA AND RV LEADS WERE CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Heart Block; No Clinical Signs, Symptoms or Conditions; Syncope/Fainting
MAUDE report 15850434 ↗InjurySep 14, 2022Failure to Capture; Electrical /Electronic Property Problem; High impedance; Failure to Capture; Electrical /Electronic Property Problem; High impedance
IT WAS REPORTED THAT THE PATIENT EXPERIENCED DIAPHRAGMATIC STIMULATION/ABDOMINAL STIMULATION AND NERVE STIMULATION. IT WAS ALSO REPORTED THAT THE EPICARDIAL RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH/UNDEFINED IMPEDANCE IN BIPOLAR CONFIGURATION AND A POLARITY SWITCH. IT WAS ALSO REPORTED THAT THE RA LEAD EXHIBITED NO CAPTURE. THE RA LEAD WAS REPROGRAMMED AND REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Device Overstimulation of Tissue; Undesired Nerve Stimulation
MAUDE report 15416410 ↗InjuryMay 3, 2022Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT EXHIBITED PRESYNCOPE. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A CONFIRMED FRACTURE. THE RV LEAD WAS CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Presyncope
MAUDE report 14266042 ↗InjuryFeb 11, 2022Fracture; High impedance; Pacing Problem; Under-Sensing; Device Sensing Problem; Fracture; High impedance; Pacing Problem; Under-Sensing; Device Sensing Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS NOT SENSING LEADING TO LEAD FAILURE, RESULTING IN THE IMPLANTABLE PULSE GENERATOR (IPG) BEING REPROGRAMMED TO MONITOR AND SENSE ONLY MODE. IT WAS ALSO NOTED THAT THE IPG DID NOT RECORD MARKERS ON ELECTROGRAMS (EGM) CORRECTLY. THE RV LEAD AND IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13514295 ↗InjuryNov 23, 2021Failure to Capture; Electrical /Electronic Property Problem; Fracture; High impedance; Low impedance; High Capture Threshold; Failure to Capture; Electrical /Electronic Property Problem; Fracture; High impedance; Low impedance; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCE, LOW IMPEDANCE, HIGH THRESHOLDS, NO CAPTURE, POLARITY SWITCH AND A POSSIBLE FRACTURE. IT WAS ALSO REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS, NO CAPTURE, HIGH/UNDEFINED IMPEDANCE, POLARITY SWITCH AND POSSIBLE FRACTURE. BOTH HE RA AND RV LEADS WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12859253 ↗InjurySep 3, 2021High impedance; Pacing Problem; High impedance; Pacing Problem
IT WAS FURTHER REPORTED THAT THE RA LEAD AND RV LEAD WERE EXPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12420579 ↗InjuryJun 17, 2021Electrical /Electronic Property Problem; Low impedance; High Capture Threshold; Electrical /Electronic Property Problem; Low impedance; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS, ATRIAL LEAD POSITION CHECK FAIL AND LOW UNIPOLAR IMPEDANCE. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOW UNIPOLAR IMPEDANCE, HIGH THRESHOLDS, SHORT V-V INTERVALS. THE LEADS WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12016932 ↗InjuryJun 17, 2021Over-Sensing; Low impedance; High Capture Threshold; Over-Sensing; Low impedance; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS, ATRIAL LEAD POSITION CHECK FAIL AND LOW UNIPOLAR IMPEDANCE. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOW UNIPOLAR IMPEDANCE, HIGH THRESHOLDS, SHORT V-V INTERVALS. THE LEADS WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12016916 ↗InjuryJun 14, 2021Fracture; Under-Sensing; Fracture; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED A FRACTURE AND NO SENSING. THE RA LEAD WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11990006 ↗InjuryJun 1, 2021Failure to Capture; High impedance; High Capture Threshold; Failure to Capture; High impedance; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH PACING IMPEDANCE AND NO CAPTURE AT MAX OUTPUTS. THE LEAD WAS REPROGRAMMED AND THEN CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11911833 ↗MalfunctionMay 31, 2021Low impedance; Low impedance
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED A LOW UNIPOLAR LEAD IMPEDANCE WARNING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11906312 ↗InjuryFeb 25, 2021Intermittent Capture; Failure to Capture; Intermittent Capture; Failure to Capture
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT WITH SEIZURE-LIKE SYMPTOMS AND SIGNIFICANT BRADYCARDIA. THEY ADMINISTERED MEDICATION TO THE PATIENT WHICH THEY DID NOT TOLERATE VERY WELL AND WAS VOMITING. IT WAS NOTED THAT THEIR PACEMAKER WAS NOT CAPTURING AND IT WAS IDENTIFIED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOSS OF CAPTURE EVEN AT THE DEVICES MAXIMUM PROGRAMMABLE OUTPUTS. SOME INTERMITTENT CAPTURE WAS ALSO OBSERVED WHEN THE PATIENT WAS LYING ON THEIR RIGHT SIDE. THE RIGHT VENTRICULAR (RV) LEAD AND THE REMAINING SYSTEM WAS PROGRAMMED OFF AND INACTIVATED AND WILL BE REMOVED AT A LATER DATE. A NEW IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia; Convulsion/Seizure; Vomiting
MAUDE report 11377242 ↗InjuryOct 9, 2020Break; High Capture Threshold; Break; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS AND "INSULATION STRIPPED WHEN TAKEN OUT OF HEADER". THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10654738 ↗InjuryAug 26, 2020Fracture; High impedance; Impedance Problem; Fracture; High impedance; Impedance Problem
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED FRACTURE, HIGH IMPEDANCE, LEAD IMPEDANCE WARNING, NO CAPTURE AND UNDERSENSING. THE LEAD WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10452926 ↗
Other Permanent defibrillator electrodes devices
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes333 deaths30,581 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M62Medtronic, Inc. · Permanent defibrillator electrodes221 deaths14,405 reports
- Capsurefix® Novus 5076-52Medtronic, Inc. · Permanent defibrillator electrodes111 deaths9,860 reports
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes80 deaths8,280 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M55Medtronic, Inc. · Permanent defibrillator electrodes116 deaths7,750 reports
- Durata™ 7122Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes151 deaths6,202 reports
- Durata™ 7122Q-65St. Jude Medical, Inc. · Permanent defibrillator electrodes86 deaths4,773 reports
- Capsurefix® Novus 5076-58Medtronic, Inc. · Permanent defibrillator electrodes47 deaths4,587 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page