EV1000 EV1000ARecalls & reported problems
Edwards Lifesciences LLC · EV1000 PLATFORM · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00690103189650), received between Mar 4, 2022 and Apr 13, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 13, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED IN THE EV1000 SYSTEM THAT WHEN CLEARSIGHT STARTS THE SYSTEM GETS TO A GOOD PHYSIOCAL CONTINUES FOR A WHILE AND THEN COMPLETELY STOPS READING AND BEGINS TO RECALIBRATE. IT ACTED AS IF THE CUFF WAS LEAKING BUT THEY SWITCHED OUT SEVERAL CUFFS AND HAD THE SAME OCCURRENCE. THEY ALSO REPORTED THE BLOOD PRESSURES BEING VERY OFF. THE INACCURATE VALUES WERE IN COMPARISON TO A BRACHIAL CUFF. THE MAPS WERE OFF BY MORE THAN 20MMHG THUS MAKING SYS AND DIA EVEN FURTHER APART. CLEARSIGHT WAS READING CLINICALLY LOW WHILE THE CUFF WAS NORMAL. THERE WERE NO CLINICAL SIGNS FOR HYPOTENSION.THIS WAS DURING USE. THERE IS NO PATIENT INJURY. PATIENT DEMOGRAPHICS WERE REQUESTED BUT UNAVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14092334 ↗MalfunctionMar 4, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED BY THE BIOMED THAT THERE WERE FLUCTUATING AND INCORRECT VALUES WITH THE EV1000 SYSTEM. WHEN THEY EXCHANGED THE SUSPECT EV1000 SYSTEM FOR ANOTHER ONE, THEN THE ISSUE WAS RESOLVED. THERE WERE NO ERROR MESSAGES PRESENT. THERE WAS NO INAPPROPRIATE PATIENT TREATMENT REPORTED. THE PATIENT DEMOGRAPHICS HAS BEEN REQUESTED, BUT NOT PROVIDED. SEVERAL ATTEMPTS HAVE BEEN MADE TO THE FACILITY FOR ADDITIONAL DETAILS AND NO RESPONSE RECEIVED. THERE WAS NO PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13673390 ↗
FDA clearance and classification
The FDA regulates product code DXG (computer, diagnostic, pre-programmed, single-function) under 21 CFR 870.1435 as a Class II device. 21 CFR 870.1435 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the EV1000 EV1000A
Has the EV1000 EV1000A been recalled?
No FDA recall record in our data names this exact device identifier (DI 00690103189650), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the EV1000 EV1000A?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Mar 4, 2022 and Apr 13, 2022. In those reports, the most common specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EV1000 EV1000A?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the EV1000 EV1000A, and is it still sold?
The FDA's device database lists Edwards Lifesciences LLC as the company responsible for it (brand EV1000, model EV1000A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION devices
- FLOTRAC MHD8Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall5 reports
- FLOTRAC, VAMP MHD6AZEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall4 reports
- FLOTRAC, VAMP, TRUWAVE MHD6C502Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall2 reports
- FLOTRAC, VAMP MHD6AZ5Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall1 report
- SWAN-GANZ 70CC2Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION4 reports
- EDWARDS OXIMETRY CENTRAL VENOUS CATHETER X3820STEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION2 reports
- ICU Medical 58400-000Icu Medical, Inc. · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 report
- ICU Medical 58400-000RIcu Medical, Inc. · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page