Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840619099459), received between Jul 24, 2024 and Jul 24, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Battery Problem” (1 report).
4 other recalls cover devices cleared under the same FDA submission (K152006); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Battery Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 24, 2024Battery Problem; Battery Problem
THE ORIGINAL EVENT OCCURRED ON AN UNSPECIFIED DATE AND INVOLVED A COGENT WHERE IT WAS REPORTED THE BASE ISN'T SHOWING CERTAIN VALUES, AND TSH900411 FOR THE SCREEN WHICH IS NOT BOOTING UP WITH ERROR. THE DEVICE WAS RETURNED FOR INVESTIGATION AND THE COMPLAINT WAS CONFIRMED OF THE TABLET NOT SHOWING VALUES. THE DEVICE FAILED VISUAL INSPECTION WITH A SWOLLEN BATTERY. THE DEVICE WAS UNABLE TO BE CHARGED THROUGH AN EXTERNAL SOURCE AND THE DISPLAY WAS NOT SHOWN. THE PROBABLE CAUSE OF THE BATTERY NOT ABLE TO BE CHARGED AND NO DISPLAY HAD OCCURRED DUE TO A DEAD BATTERY. THERE WAS NOT PATIENT INVOLVEMENT AND NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19816640 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K152006. One submission often covers many models, so check the product and lot details in each record.
- Z-0859-2021 · Nov 9, 2020 · Terminatedicumedical Cogent" Hemodynamic Monitoring System - Product Usage: intended for patients for whom the monitoring of continuous cardiac output and calculated hemodynamic parameters i
- Z-1927-2020 · Feb 14, 2020 · TerminatedThe Cogent Hemodynamic Monitoring System. REF58400-000, GTIN 00840619099459; REF 58400-000R, GTIN 0087709094314; REF 58403, GTIN 00840619079505; REF 58401, GTIN 00840619079499 - Pr
- Z-1835-2020 · Jul 12, 2019 · TerminatedICU Medical Cogent Hemodynamic Monitoring System, Cogent Item Number 58400-000, UDI: (01)00840619079499
- Z-1775-2018 · Jan 5, 2018 · TerminatedHemodynamic monitoring system (HMS), comprised of a Patient Interface Module (Part Number 58401-000 and a User Interface Module (Part Number 58403-000). Intended for patients for
Other COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION devices
- FLOTRAC MHD8Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall5 reports
- FLOTRAC, VAMP MHD6AZEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall4 reports
- FLOTRAC, VAMP, TRUWAVE MHD6C502Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall2 reports
- FLOTRAC, VAMP MHD6AZ5Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall1 report
- SWAN-GANZ 70CC2Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION4 reports
- EDWARDS OXIMETRY CENTRAL VENOUS CATHETER X3820STEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION2 reports
- EV1000 EV1000AEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION2 reports
- FLOTRAC MHD65Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page