EDWARDS OXIMETRY CENTRAL VENOUS CATHETER X3820STRecalls & reported problems
Edwards Lifesciences LLC · EDWARDS OXIMETRY CENTRAL VENOUS CATHETER · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00690103197587), received between Jan 7, 2025 and Jan 7, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Insufficient Flow or Under Infusion” (2 reports).
6 other recalls cover devices cleared under the same FDA submission (K053609, K110167); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Flow or Under Infusion2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 7, 2025Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion
IT WAS REPORTED THAT THERE WAS A MALFUNCTION WITH THE OXIMETRY CENTRAL VENOUS CATHETER, DURING USE. THERE WAS LEAKAGE FROM THE OPTICAL MODULE CONNECTOR OF THE CATHETER. THIS RESULTED IN MEDICATION LOSS OF POTASSIUM. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21096653 ↗MalfunctionJan 7, 2025Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion
IT WAS REPORTED THAT THERE WAS A MALFUNCTION WITH THE OXIMETRY CENTRAL VENOUS CATHETER, DURING USE. THE CATHETER BODY BECAME KINKED NEAR THE DISTAL END OF THE BACKFORM. AS A RESULT, THE CATECHOLAMINE MEDICATION WAS NOT INFUSED CONTINUOUSLY INTO THE PATIENT. THE MEDICATION BECAME STUCK AT THE KINKED PORTION OF THE CATHETER. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21096650 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K053609, K110167. One submission often covers many models, so check the product and lot details in each record.
- Z-2283-2021 · Jul 9, 2021 · TerminatedEdwards PediaSat Oximetry Catheter Set, REF: XT3515SJ, Exterior Diameter 5.5F, Usable Length 15 cm, Number of Lumens: 3, Rx Only, Sterile EO, UDI: (01)00690103197501 *NOT DISTR
- Z-2282-2021 · Jul 9, 2021 · TerminatedEdwards PediaSat Oximetry Catheter Set, REF: XT358KTP, Exterior Diameter 5.5F, Usable Length 8 cm, Number of Lumens: 3, Recommended Guidewire Size: GW: 0.025" (0.64 mm), Rx Only,
- Z-2280-2021 · Jul 9, 2021 · TerminatedEdwards PediaSat Oximetry Catheter Kit, REF: XT358SJ, Usable Length 8 cm, Number of Lumens 3, Exterior Diameter 5.5F, Rx Only, Sterile EO, UDI: (01)00690103197495 *NOT DISTRIBUT
- Z-2279-2021 · Jul 9, 2021 · TerminatedEdwards PediaSat Oximetry Catheter Kit, REF: XT3515SP, Exterior Diameter 5.5F (1.83 mm), Usable Length 15 cm, Number of Lumens 3, Recommended Guidewire Size 0.025" (0.64 mm), Rx On
- Z-2281-2021 · Jul 9, 2021 · TerminatedEdwards PediaSat Oximetry Catheter Set, REF: XT358SP, Exterior Diameter 5.5F (1.83 mm), Usable Length 9 cm, Number of Lumens: 3, Recommended Guidewire Size: GW: 0.025" (0.64 mm),
- Z-2278-2021 · Jul 9, 2021 · TerminatedEdwards PediaSat Oximetry Catheter Kit, REF: XT3515KTP, Exterior Diameter 5.5F (1.83 mm), Usable Length 15 cm, Number of Lumens 3, Recommended Guidewire Size 0.025" (0.64 mm), Rx O
Other COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION devices
- FLOTRAC MHD8Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall5 reports
- FLOTRAC, VAMP MHD6AZEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall4 reports
- FLOTRAC, VAMP, TRUWAVE MHD6C502Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall2 reports
- FLOTRAC, VAMP MHD6AZ5Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall1 report
- SWAN-GANZ 70CC2Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION4 reports
- EV1000 EV1000AEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION2 reports
- ICU Medical 58400-000RIcu Medical, Inc. · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 report
- PEDIASAT XT245SPEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.