FLOTRAC, VAMP MHD6AZ5Recalls & reported problems
Edwards Lifesciences LLC · FLOTRAC SENSOR W/ VAMP ADULT (60"/152CM) · Class 2
All models of FLOTRAC, VAMP (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00690103194746), received between Sep 18, 2024 and Sep 18, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K131588, K131892, K152980); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Oct 4, 2023Class IIOpen, Classified
Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer.
Z-0412-2024 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 18, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED DURING USE THAT MHD6AZ5 FLOTRAC SENSOR WITH VAMP ADULT WAS USED ON A PATIENT AND THE DIASTOLIC WAS IN THE 20S ON THE A LINE BUT THE UPPER ARM CUFF DIASTOLIC WAS IN THE 70 TO 80S. THE WAVEFORM WAS PERFECT ON THE A LINE AND THE SQUARE WAVE TEST WAS GOOD. THEY RE ZEROED AND NOTHING CHANGED SO THEY DECIDED TO CHANGE OUT THE TRANSDUCER COMPLETELY AND IT MATCHED THE NIBP CUFF RIGHT AWAY. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20252181 ↗
FDA clearance and classification
Cleared through 510(k) K152980 on Jan 19, 2016 (decision: Substantially Equivalent). Submitted by Edwards Lifesciences, LLC as “Flo Trac sensor, Volume View sensor”. The FDA review took 102 days. FDA record ↗
Cleared through 510(k) K131588 on May 22, 2014 (decision: Substantially Equivalent). Submitted by Edwards Lifesciences, LLC as “Vigileo Arterial Pressure Cardiac Output/Oximetry Monitor; Vigileo Apco/Oximetry Monitor”. The FDA review took 356 days. FDA record ↗
Cleared through 510(k) K131892 on May 22, 2014 (decision: Substantially Equivalent). Submitted by Edwards Lifesciences, LLC as “Ev1000 Clinical Platform”. The FDA review took 331 days. FDA record ↗
The FDA regulates product code DXG (computer, diagnostic, pre-programmed, single-function) under 21 CFR 870.1435 as a Class II device. 21 CFR 870.1435 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K131588, K131892, K152980. One submission often covers many models, so check the product and lot details in each record.
- Z-0409-2024 · Oct 4, 2023 · Open, ClassifiedAcumen IQ Sensor, REFs: AIQS6, AIQS65, AIQS6R, AIQS6R5, AIQS8, AIQS85, AIQS8R5, Rx only, Sterile EO. For use in intravascular pressure monitoring.
- Z-0411-2024 · Oct 4, 2023 · Open, ClassifiedFloTrac Sensor, REF: MHD8 & MHD85,MHD6R5, MHD8R, MHD8R5, MHD8S, STERILEEO, Rx only. For use in intravascular pressure monitoring.
- Z-0410-2024 · Oct 4, 2023 · Open, ClassifiedAcumen IQ Sensor with VAMP System, REFs: AIQS6AZ & AIQS6AZ5, STERILEEO, Rx only. For use in intravascular pressure monitoring.
Common questions about the FLOTRAC, VAMP MHD6AZ5
Has the FLOTRAC, VAMP MHD6AZ5 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0412-2024 (Class II), began on Oct 4, 2023: Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the FLOTRAC, VAMP MHD6AZ5?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Sep 18, 2024. That report lists device problem “Incorrect, Inadequate or Imprecise Result or Readings” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the FLOTRAC, VAMP MHD6AZ5?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the FLOTRAC, VAMP MHD6AZ5 FDA approved?
It was cleared through 510(k) K131588 on May 22, 2014; its record lists 3 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the FLOTRAC, VAMP MHD6AZ5, and is it still sold?
The FDA's device database lists Edwards Lifesciences LLC as the company responsible for it (brand FLOTRAC, VAMP, model MHD6AZ5). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION devices
- FLOTRAC MHD8Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall5 reports
- FLOTRAC, VAMP MHD6AZEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall4 reports
- FLOTRAC, VAMP, TRUWAVE MHD6C502Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall2 reports
- SWAN-GANZ 70CC2Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION4 reports
- EDWARDS OXIMETRY CENTRAL VENOUS CATHETER X3820STEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION2 reports
- EV1000 EV1000AEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION2 reports
- ICU Medical 58400-000Icu Medical, Inc. · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 report
- ICU Medical 58400-000RIcu Medical, Inc. · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page