FLOTRAC, VAMP MHD6AZ5Recalls & reported problems
Edwards Lifesciences LLC · FLOTRAC SENSOR W/ VAMP ADULT (60"/152CM) · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00690103194746), received between Sep 18, 2024 and Sep 18, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K131588, K131892, K152980); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Oct 4, 2023Open, Classified
Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer.
Z-0412-2024 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 18, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED DURING USE THAT MHD6AZ5 FLOTRAC SENSOR WITH VAMP ADULT WAS USED ON A PATIENT AND THE DIASTOLIC WAS IN THE 20S ON THE A LINE BUT THE UPPER ARM CUFF DIASTOLIC WAS IN THE 70 TO 80S. THE WAVEFORM WAS PERFECT ON THE A LINE AND THE SQUARE WAVE TEST WAS GOOD. THEY RE ZEROED AND NOTHING CHANGED SO THEY DECIDED TO CHANGE OUT THE TRANSDUCER COMPLETELY AND IT MATCHED THE NIBP CUFF RIGHT AWAY. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20252181 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131588, K131892, K152980. One submission often covers many models, so check the product and lot details in each record.
- Z-0411-2024 · Oct 4, 2023 · Open, ClassifiedFloTrac Sensor, REF: MHD8 & MHD85,MHD6R5, MHD8R, MHD8R5, MHD8S, STERILEEO, Rx only. For use in intravascular pressure monitoring.
- Z-0410-2024 · Oct 4, 2023 · Open, ClassifiedAcumen IQ Sensor with VAMP System, REFs: AIQS6AZ & AIQS6AZ5, STERILEEO, Rx only. For use in intravascular pressure monitoring.
- Z-0409-2024 · Oct 4, 2023 · Open, ClassifiedAcumen IQ Sensor, REFs: AIQS6, AIQS65, AIQS6R, AIQS6R5, AIQS8, AIQS85, AIQS8R5, Rx only, Sterile EO. For use in intravascular pressure monitoring.
Other COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION devices
- FLOTRAC MHD8Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall5 reports
- FLOTRAC, VAMP MHD6AZEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall4 reports
- FLOTRAC, VAMP, TRUWAVE MHD6C502Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall2 reports
- SWAN-GANZ 70CC2Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION4 reports
- EV1000 EV1000AEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION2 reports
- EDWARDS OXIMETRY CENTRAL VENOUS CATHETER X3820STEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION2 reports
- ICU Medical 58400-000RIcu Medical, Inc. · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 report
- PEDIASAT XT245SPEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.