FLOTRAC MHD8Recalls & reported problems
Edwards Lifesciences LLC · FLOTRAC SENSOR · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00690103194609), received between Aug 8, 2024 and Aug 14, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K131588, K131892, K152980); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Oct 4, 2023Open, Classified
Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer.
Z-0411-2024 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Incorrect Measurement3
- Appropriate Device Problem Term/Code Not Available1
- Delivered as Unsterile Product1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 14, 2025Incorrect Measurement; Incorrect Measurement
AS REPORTED, AFTER BEING CORRECTLY CONNECTED, THIS FLOTRAC SENSOR CONSISTENTLY DISPLAYED A SIGNIFICANTLY LOW DIASTOLIC BLOOD PRESSURE OF 10 MMHG, DESPITE THE EXPECTED NORMAL RANGE BEING 60/80 MMHG. NO ERROR MESSAGE WAS DISPLAYED. THE PATIENT WAS NOT TREATED ACCORDING TO THE INCORRECT VALUES. THE ISSUE WAS SOLVED REPLACING THE UNIT. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22802671 ↗MalfunctionJul 2, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
AS REPORTED, BEFORE OPENING THE PACKAGE OF THIS FLOTRAC SENSOR, THE PACKAGE WAS FOUND BROKEN AND THE STERILITY WAS AFFECTED. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22382421 ↗MalfunctionJun 4, 2025Incorrect Measurement; Incorrect Measurement
AS REPORTED, DURING USE WITH PATIENT, THE BLOOD PRESSURE VALUE OBTAINED WITH THIS FLOTRAC SENSOR WAS TOO HIGH AND INACCURATE. THE EXPECTED VALUE ACCORDING TO PATIENT STATUS WAS AROUND 100 MMHG; HOWEVER, THE VALUE PROVIDED BY THE DEVICE WAS UNABLE TO BE OBTAINED. THE PATIENT WAS NOT TREATED ACCORDING TO THE INACCURATE VALUE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION, SINCE IT WAS DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22138683 ↗MalfunctionSep 4, 2024Incorrect Measurement; Incorrect Measurement
AS REPORTED, DURING USE IN PATIENT, THIS FLOTRAC SENSOR PROVIDED AN INACCURATE OUTPUT. THE ISSUE WAS SOLVED REPLACING THE DEVICE. NO FURTHER INFORMATION AVAILABLE REGARDING THE INACCURATE VALUES OBTAINED. THERE WAS NO ALLEGATION OF PATIENT INJURY, AND PATIENT WAS NOT TREATED ACCORDING TO THE INACCURATE VALUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20133797 ↗MalfunctionAug 8, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
AS REPORTED WITH THIS FLOTRAC SENSOR, THE END OF THE PRESSURE TUBING, THE STOPCOCK LUER WAS DETACHED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19938356 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131588, K131892, K152980. One submission often covers many models, so check the product and lot details in each record.
- Z-0409-2024 · Oct 4, 2023 · Open, ClassifiedAcumen IQ Sensor, REFs: AIQS6, AIQS65, AIQS6R, AIQS6R5, AIQS8, AIQS85, AIQS8R5, Rx only, Sterile EO. For use in intravascular pressure monitoring.
- Z-0412-2024 · Oct 4, 2023 · Open, ClassifiedFloTrac Sensor with Vamp System, REF: MHD6AZ, MHD6AZ5 , MHD6C502 STIRLEEO, Rx only. For use in intravascular pressure monitoring.
- Z-0410-2024 · Oct 4, 2023 · Open, ClassifiedAcumen IQ Sensor with VAMP System, REFs: AIQS6AZ & AIQS6AZ5, STERILEEO, Rx only. For use in intravascular pressure monitoring.
Other COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION devices
- FLOTRAC, VAMP MHD6AZEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall4 reports
- FLOTRAC, VAMP, TRUWAVE MHD6C502Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall2 reports
- FLOTRAC, VAMP MHD6AZ5Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall1 report
- SWAN-GANZ 70CC2Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION4 reports
- EV1000 EV1000AEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION2 reports
- EDWARDS OXIMETRY CENTRAL VENOUS CATHETER X3820STEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION2 reports
- ICU Medical 58400-000RIcu Medical, Inc. · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 report
- PEDIASAT XT245SPEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page