Rapicide PA High-Level Disinfectant ML020117Recalls & reported problems
Medivators Inc. · Rapicide PA High-Level Disinfectant, Intl Part A, 5 Ltr, and Part B Solutions, 5 Ltr, Assembly · Class 2
All models of Rapicide PA High-Level Disinfectant (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00724995245962), received between Jul 9, 2026 and Sep 10, 2026 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K082988); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 21, 2026Open, Classified
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
Z-0156-2027 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Leak/Splash5
- Device Handling Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Burn(s)1
- Eye Irritation1
- Upper Respiratory Tract Irritation1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 10, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50721691 ↗MalfunctionAug 28, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50498790 ↗InjuryAug 13, 2026Device Handling Problem
The user facility reported while an employee was utilizing a Rapicide PA Test Strip test after a cycle, the collection container was dropped subsequently causing Rapicide PA High-Level Disinfectant to splash into the employee¿s eye. The employee flushed their eyes with water and sought treatment at the facility¿s Emergency Department.
Patient problems: Eye Irritation
MAUDE report 50238207 ↗InjuryJul 31, 2026Leak/Splash
The user facility reported that an employee came into contact with Rapicide PA High-level Disinfectant and sustained a burn. It is unknown whether medical treatment was sought or administered.
Patient problems: Burn(s)
MAUDE report 50098269 ↗MalfunctionJul 16, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25846920 ↗MalfunctionJul 9, 2026Leak/Splash
Patient problems: Upper Respiratory Tract Irritation
MAUDE report 25770997 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082988. One submission often covers many models, so check the product and lot details in each record.
- Z-0154-2027 · Sep 21, 2026 · Open, ClassifiedRAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115
- Z-0155-2027 · Sep 21, 2026 · Open, ClassifiedRAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116
- Z-0617-2024 · Nov 13, 2023 · Open, ClassifiedHookup Accessories used with MEDIVATORS ADVANTAGE PLUS and ADVANTAGE PLUS Pass Thru Automated Endoscope Reprocessors Model # Model Description 2-8-001 HU ZERO CHANNEL NO LT 2-8-01
Other Accessories, cleaning, for endoscope devices
- ENDOSCOPE REPROCESSOR OER-ProOlympus Medical Systems Corp. · ACCESSORIES, CLEANING, FOR ENDOSCOPE3 recalls1,163 reports
- ENDOSCOPE REPROCESSOR OER-EliteOlympus Medical Systems Corp. · Accessories, cleaning, for endoscope4 recalls323 reports
- SINGLE USE DISTAL COVER MAJ-2315Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope2 recalls294 reports
- CONNECTING TUBE MAJ-2110Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall213 reports
- CONNECTING TUBE MAJ-2111Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall164 reports
- CONNECTING TUBE MAJ-2113Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall159 reports
- CONNECTING TUBE MAJ-2112Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall143 reports
- ENDOSCOPE REPROCESSOR OER-MINIOlympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope2 recalls11 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.