Skip to content

Rapicide PA High-Level Disinfectant ML020117Recalls & reported problems

Medivators Inc. · Rapicide PA High-Level Disinfectant, Intl Part A, 5 Ltr, and Part B Solutions, 5 Ltr, Assembly · Class 2

All models of Rapicide PA High-Level Disinfectant (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00724995245962), received between Jul 9, 2026 and Sep 10, 2026 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K082988); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 3 reportsAug 2026: 2 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 21, 2026Open, Classified

    If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.

    Z-0156-2027 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Leak/Splash5
  • Device Handling Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Burn(s)1
  • Eye Irritation1
  • Upper Respiratory Tract Irritation1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50721691 ↗
  • MalfunctionAug 28, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50498790 ↗
  • InjuryAug 13, 2026Device Handling Problem

    The user facility reported while an employee was utilizing a Rapicide PA Test Strip test after a cycle, the collection container was dropped subsequently causing Rapicide PA High-Level Disinfectant to splash into the employee¿s eye. The employee flushed their eyes with water and sought treatment at the facility¿s Emergency Department.

    Patient problems: Eye Irritation

    MAUDE report 50238207 ↗
  • InjuryJul 31, 2026Leak/Splash

    The user facility reported that an employee came into contact with Rapicide PA High-level Disinfectant and sustained a burn. It is unknown whether medical treatment was sought or administered.

    Patient problems: Burn(s)

    MAUDE report 50098269 ↗
  • MalfunctionJul 16, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25846920 ↗
  • MalfunctionJul 9, 2026Leak/Splash

    Patient problems: Upper Respiratory Tract Irritation

    MAUDE report 25770997 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K082988. One submission often covers many models, so check the product and lot details in each record.

  • Z-0154-2027 · Sep 21, 2026 · Open, Classified
    RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115
  • Z-0155-2027 · Sep 21, 2026 · Open, Classified
    RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116
  • Z-0617-2024 · Nov 13, 2023 · Open, Classified
    Hookup Accessories used with MEDIVATORS ADVANTAGE PLUS and ADVANTAGE PLUS Pass Thru Automated Endoscope Reprocessors Model # Model Description 2-8-001 HU ZERO CHANNEL NO LT 2-8-01

Other Accessories, cleaning, for endoscope devices

All Accessories, cleaning, for endoscope devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page