Skip to content

Welch Allyn, Inc. 901060 (REF 65NTXX-B)Recalls & reported problems

Welch Allyn, Inc. · Welch Allyn Connex 6500 Electronic Vitals Documentation System; Nellcor SpO2, SureTemp Plus Thermometry, SureBP Non-invasive Blood Pressure, Pulse Rate, MAP, BP Averaging, Nurse Call, Four USB Ports for Accessories, Internal 802.11 a/b/g Wireless Radio; 100-240 V, 50-60 Hz AC, IEC Plug Type-B · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00732094088151), received between Jul 14, 2021 and Nov 4, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fire” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 24, 2025Class IIOpen, Classified

    There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.

    Z-1558-2025 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fire3
  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Electric Shock1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 4, 2025Fire; Fire

    CUSTOMER REPORTED THAT THE CVSM DEVICE CAUGHT FIRE. IT WAS ADDITIONALLY STATED THAT STAFF WAS ABLE TO UNPLUG THE DEVICE FROM AC POWER AND PUT THE FIRE WITH A FIRE EXTINGUISHER. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23462513 ↗
  • MalfunctionJul 9, 2025Fire; Sparking; Fire; Sparking

    IT WAS REPORTED THAT WHEN A CAREGIVER ATTEMPTED TO USE THE VSM 6500 VITAL SIGNS MONITORING DEVICE, THE DEVICE STARTED SPARKING, THE CAREGIVER FELT AN ELECTRIC SHOCK UP THE ARM, AND THE DEVICE WAS THEN ENGULFED IN FLAMES. THE CAREGIVER WAS EVALUATED IN THE EMERGENCY ROOM. THE CUSTOMER REPORTED THAT NO SIGNIFICANT INJURY OCCURRED. ANY TREATMENT PROVIDED TO THE CAREGIVER WHILE UNDER OBSERVATION IN THE ER IS UNKNOWN.

    Patient problems: Electric Shock

    MAUDE report 22456553 ↗
  • MalfunctionJul 14, 2021Fire; Fire

    WELCH ALLYN RECEIVED A REPORT FROM A HILLROM TECHNICIAN STATING THE CVSM 6500 CAUGHT ON FIRE. THE DEVICE WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED.THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12169698 ↗

Other MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS) devices

All MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS) devices →