Welch Allyn, Inc. 901001Recalls & reported problems
Welch Allyn, Inc. · Welch Allyn Connex Spot Monitor Accessory Power Management Stand · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00732094210613), received between May 13, 2021 and Sep 25, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Sparking” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K142356); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 24, 2025Class IIOpen, Classified
There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
Z-1559-2025 · Root cause: Process design · Recall details · FDA record ↗
Most reported problems
Device problems
- Sparking2
- Electrical Overstress1
- Electrical Power Problem1
- Failure to Power Up1
- Fire1
- Melted1
- Overheating of Device1
- Thermal Decomposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2026Melted; Failure to Power Up; Sparking; Electrical Overstress
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50902927 ↗MalfunctionSep 24, 2026Thermal Decomposition of Device; Sparking
Baxter received a report stating that a Power Management Stand emitted sparks and burning smell. The power cable appeared to be blackened. The process step during which this occurred was unknown. There was no patient/user injury, medical intervention, symptom, or adverse event reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50877612 ↗MalfunctionFeb 4, 2025Overheating of Device; Overheating of Device
CUSTOMER REPORTED THAT A STRONG SMELL OF BURNING WAS COMING FROM THE ACCESSORY POWER MANAGEMENT (APM). THE APM WAS UNSCREWED AND IT WAS FOUND THAT AN INTERNAL COMPONENT HAS BURNED ON THE POWER SUPPLY BOARD. THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21302445 ↗MalfunctionJun 18, 2024Fire; Fire
IT WAS REPORTED THAT THE CONNEX SPOT MONITOR WAS PLUGGED IN WHEN STAFF HEARD A POP. STAFF THEN NOTED SMOKE AND FIRE COMING FROM THE DEVICE. THERE WAS NO REPORT OF SERIOUS INJURY. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19555810 ↗MalfunctionMay 13, 2021Electrical Power Problem; Electrical Power Problem
THE CUSTOMER REPORTED THAT THE POWER CORD ATTACHED TO THE APM EXPLODED WHILE THE DEVICE WAS IN USE. THE CUSTOMER ALSO STATED THAT BOTH THE APM AND THE CSM ARE FUNCTIONING AS DESIGNED. THERE WAS NO REPORT OF SERIOUS INJURY. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11819202 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K142356. One submission often covers many models, so check the product and lot details in each record.
- Z-1564-2025 · Mar 21, 2025 · Open, ClassifiedWelch Allyn Connex Vital Signs Monitor, Product Code/Part Numbers: 1) ) Product Code/Part # 67MXTP-B-ECG3A; 2) ) Product Code/Part # 67MXTX-B; 3) ) Product Code/Part # 67MXXP-B
- Z-1565-2025 · Mar 21, 2025 · Open, ClassifiedWelch Allyn Spot Vital Signs 4400 Device, Product Code/Part Numbers: 1) 44WT-2; 2) 44WT-3; 3) 44WT-4; 4) 44WT-6; 5) 44WT-B; 6) 44XT-2; 7) 44XT-4; 8) 44XT-6; 9) 44XT-B.
- Z-2581-2023 · Aug 10, 2023 · Open, ClassifiedWelch Allyn Connex Spot Monitor
Other Monitor, physiological, patient (without arrhythmia detection or alarms) devices
- Rad-97 9738Masimo Corporation · Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)1 death222 reports
- Xper Flex Cardio Physiomonitoring System Xper Flex Cardio FC2010 REV CPhilips Ds North America LLC · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls1 death42 reports
- Xper Flex Cardio Physiomonitoring System FC 2010Philips Ds North America LLC · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls2 deaths38 reports
- Welch Allyn, Inc. 901058Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls5 reports
- Welch Allyn, Inc. 901060Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls3 reports
- Welch Allyn, Inc. 901060Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls2 reports
- McKesson Cardiology Hemo 14.0Change Healthcare Canada Company · Monitor,Physiological,Patient(Without Arrhythmia Detection Or Alarms)2 recalls2 deaths2 reports
- Change Healthcare Cardiology Hemo 14.2Change Healthcare Canada Company · Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)2 recalls1 death1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page