Skip to content

Welch Allyn, Inc. 901001Recalls & reported problems

Welch Allyn, Inc. · Welch Allyn Connex Spot Monitor Accessory Power Management Stand · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00732094210613), received between May 13, 2021 and Sep 25, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Sparking” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K142356); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 24, 2025Class IIOpen, Classified

    There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.

    Z-1559-2025 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Sparking2
  • Electrical Overstress1
  • Electrical Power Problem1
  • Failure to Power Up1
  • Fire1
  • Melted1
  • Overheating of Device1
  • Thermal Decomposition of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Melted; Failure to Power Up; Sparking; Electrical Overstress

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50902927 ↗
  • MalfunctionSep 24, 2026Thermal Decomposition of Device; Sparking

    Baxter received a report stating that a Power Management Stand emitted sparks and burning smell. The power cable appeared to be blackened. The process step during which this occurred was unknown. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50877612 ↗
  • MalfunctionFeb 4, 2025Overheating of Device; Overheating of Device

    CUSTOMER REPORTED THAT A STRONG SMELL OF BURNING WAS COMING FROM THE ACCESSORY POWER MANAGEMENT (APM). THE APM WAS UNSCREWED AND IT WAS FOUND THAT AN INTERNAL COMPONENT HAS BURNED ON THE POWER SUPPLY BOARD. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21302445 ↗
  • MalfunctionJun 18, 2024Fire; Fire

    IT WAS REPORTED THAT THE CONNEX SPOT MONITOR WAS PLUGGED IN WHEN STAFF HEARD A POP. STAFF THEN NOTED SMOKE AND FIRE COMING FROM THE DEVICE. THERE WAS NO REPORT OF SERIOUS INJURY. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19555810 ↗
  • MalfunctionMay 13, 2021Electrical Power Problem; Electrical Power Problem

    THE CUSTOMER REPORTED THAT THE POWER CORD ATTACHED TO THE APM EXPLODED WHILE THE DEVICE WAS IN USE. THE CUSTOMER ALSO STATED THAT BOTH THE APM AND THE CSM ARE FUNCTIONING AS DESIGNED. THERE WAS NO REPORT OF SERIOUS INJURY. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11819202 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K142356. One submission often covers many models, so check the product and lot details in each record.

  • Z-1564-2025 · Mar 21, 2025 · Open, Classified
    Welch Allyn Connex Vital Signs Monitor, Product Code/Part Numbers: 1) ) Product Code/Part # 67MXTP-B-ECG3A; 2) ) Product Code/Part # 67MXTX-B; 3) ) Product Code/Part # 67MXXP-B
  • Z-1565-2025 · Mar 21, 2025 · Open, Classified
    Welch Allyn Spot Vital Signs 4400 Device, Product Code/Part Numbers: 1) 44WT-2; 2) 44WT-3; 3) 44WT-4; 4) 44WT-6; 5) 44WT-B; 6) 44XT-2; 7) 44XT-4; 8) 44XT-6; 9) 44XT-B.
  • Z-2581-2023 · Aug 10, 2023 · Open, Classified
    Welch Allyn Connex Spot Monitor

Other Monitor, physiological, patient (without arrhythmia detection or alarms) devices

All Monitor, physiological, patient (without arrhythmia detection or alarms) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page