Sprint Quattro Secure S MRI™ SureScan™ 6935M62 (UDI 00763000180126)Recalls & reported problems

Medtronic, Inc. · LEAD 6935M62 MRI US DF4 RCMCRD · Class 3 · Implant

All models of Medtronic Sprint Quattro Lead (67) · Sprint Quattro Secure S MRI SureScan (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Sprint Quattro Secure S defibrillator lead is a wire placed in the heart. Of its 275 reports, 141 are injury reports, 131 are malfunction reports, and 3 deaths are listed. Many describe incorrect heartbeat sensing and noise, and some describe infections or a heart wall perforation. All 275 reports are from the latest year. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
309
Deaths reported
3
Injuries reported
156
Recalls naming it
0

Summary

The FDA's MAUDE database contains 309 adverse event reports that list this exact device (DI 00763000180126), received between Nov 3, 2025 and Sep 30, 2026 — classified as 3 deaths, 156 injuries, 150 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (50 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 6 reportsDec 2025: 8 reportsJan 2026: 23 reportsFeb 2026: 34 reportsMar 2026: 29 reportsApr 2026: 44 reportsMay 2026: 46 reportsJun 2026: 29 reportsJul 2026: 13 reportsAug 2026: 31 reportsSep 2026: 46 reportsOct 2026: 0 reports20242025202646
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem79
  • High Capture Threshold50
  • Over-Sensing44
  • Under-Sensing36
  • Signal Artifact/Noise25
  • High impedance21
  • Appropriate Device Problem Term/Code Not Available18
  • Capturing Problem18

Patient problems

  • No Clinical Signs, Symptoms or Conditions202
  • Unspecified Infection35
  • Medical device site infection19
  • Purulent Discharge16
  • Erythema14
  • Pain11
  • Swelling/ Edema10
  • Bacteremia9

Recent adverse event reports

The 25 most recent of 309 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Cardiac Arrest; Cardiac Tamponade; Perforation

    MAUDE report 50952740 ↗
  • InjurySep 30, 2026Inappropriate/Inadequate Shock/Stimulation; Device Dislodged or Dislocated

    Patient problems: Shock from Patient Lead(s); Twiddlers Syndrome

    MAUDE report 50951886 ↗
  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Implantable Cardioverter Defibrillator (ICD) system was explanted due to infection with bacteremia. No further patient complications have been reported as a result of this event.

    Patient problems: Bacteremia; Unspecified Infection

    MAUDE report 50951211 ↗
  • InjurySep 29, 2026Low impedance; High Capture Threshold

    It was reported that the Right Ventricular (RV) lead triggered alerts due to high thresholds and also an alert for low, out-of-range pacing impedance. The impedance measurement was noted to be consistent with historical values. The lead was capped and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933442 ↗
  • MalfunctionSep 25, 2026Capturing Problem; High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50903739 ↗
  • InjurySep 25, 2026High Capture Threshold; Unstable Capture Threshold

    Patient problems: Pericardial Effusion

    MAUDE report 50902376 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Medical device site infection; Purulent Discharge

    MAUDE report 50890942 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Implantable Cardioverter Defibrillator (ICD) system was explanted due to infection with fever, bacteremia, and purulent discharge/pus. The organism was identified as Staphylococcus aureus. No further patient complications have been reported as a result of this event.

    Patient problems: Bacteremia; Bacterial Infection; Fever; Purulent Discharge

    MAUDE report 50885223 ↗
  • MalfunctionSep 24, 2026Over-Sensing; High Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited high capture threshold and increased Sensing Integrity Counter (SIC). The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884600 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that approximately one month post replacement procedure, the patient developed a pocket infection. It was noted that the patient developed purulent discharge/pus. The Cardiac Resynchronization Therapy Defibrillator (CRT-D) system was explanted and replaced with a Leadless Implantable Pulse Generator (IPG). No further patient complications have been reported as a result of this event.

    Patient problems: Unspecified Infection

    MAUDE report 50877005 ↗
  • MalfunctionSep 23, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872638 ↗
  • MalfunctionSep 23, 2026High Capture Threshold; Unstable Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited high capture thresholds. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50869322 ↗
  • MalfunctionSep 18, 2026High impedance; Capturing Problem; Impedance Problem; High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50829011 ↗
  • InjurySep 18, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50828369 ↗
  • MalfunctionSep 18, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50827807 ↗
  • InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50827494 ↗
  • InjurySep 18, 2026Decreased Sensitivity; Capturing Problem; High Capture Threshold; Unstable Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50819578 ↗
  • InjurySep 17, 2026High Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited high thresholds. The lead was explanted and replaced. No patient comp lications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50809544 ↗
  • MalfunctionSep 17, 2026High impedance

    It was reported that the Right Ventricular (RV) lead exhibited high Out of Range (OOR) pacing impedance several times during an implant procedure. The RV lead was removed and replaced with a new lead. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50809243 ↗
  • MalfunctionSep 17, 2026High Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited high thresholds. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50808828 ↗
  • InjurySep 17, 2026Failure to Capture; Device Dislodged or Dislocated

    It was reported that the Right Ventricular (RV) lead became dislodged and exhibited loss of capture. The RV lead was repositioned and remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50806853 ↗
  • MalfunctionSep 17, 2026Over-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited T-Wave Oversensing (TWOS) noted on a stored episode resulting in misc lassification of arrhythmia episode type. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50806193 ↗
  • InjurySep 17, 2026Over-Sensing

    It was reported that post implant the Right Ventricular (RV) lead exhibited T-Wave Oversensing (TWOS) post ventricular pacing. The RV lead also triggered a Lead Integrity Alert (LIA) due to High Rate Non-Sustained episodes and Sensing Integrity Counter (SIC). Repr ogramming was performed but TWOS was still present. Additionally, the Right Atrial (RA) lead exhibited Far Field R-Wave (FFRW) oversensing. Reprogramming was also performed for the RA lead. The RV and RA leads remain in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50805601 ↗
  • MalfunctionSep 17, 2026Over-Sensing

    It was reported that one day after Right Ventricular (RV) lead implant, the patient experienced a low heart rate in the 30¿s beats per minute (bpm). The patient was symptomatic and telemetry indicated the low rate for an extended period and T-Wave Oversensing (TWOS). It was suspected that the TWOS was clinically related. The lead remains in use. No further patient complications have been reported as a result of this event.

    Patient problems: Bradycardia

    MAUDE report 50804668 ↗
  • InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Cardiac Resynchronization Therapy Defibrillator (CRT-D) device system was explanted due to a pocket infection with purulent discharge. No further patient complications have been reported as a result of this event.

    Patient problems: Medical device site infection; Purulent Discharge

    MAUDE report 50787295 ↗

Other Permanent defibrillator electrodes devices

All Permanent defibrillator electrodes devices →