Bard® Eagle® Inflation Device 325110Recalls & reported problems

C. R. Bard, Inc. · Bard® Eagle® Inflation Device · Class 2

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00801741067242), received between Oct 10, 2019 and Oct 9, 2024 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak4
  • Material Fragmentation4
  • Inflation Problem3
  • Break1
  • Device Handling Problem1
  • Disconnection1
  • Leak/Splash1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 9, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE URETERAL STRICTURE WAS ENCOUNTERED DURING URETERAL STONE SURGERY USING A URETEROSCOPIC BALLOON DILATION CATHETER, WHICH WAS UNPACKED WITH THE PRESSURIZED PUMP FOUND TO BE BROKEN. THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT SET OF BALLOONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20406513 ↗
  • MalfunctionJun 18, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT UPON OPENING THE BALLOON DILATION CATHETER, IT WAS FOUND THAT THE PRESSURE PUMP GAUGE WAS DAMAGED AND COULD NOT BE USED NORMALLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19561316 ↗
  • MalfunctionMay 8, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT BEFORE THE URETERAL STRICTURE DILATION SURGERY, WHEN THE EAGLE INFLATION DEVICE WAS OPENED AND READY TO BE USED, IT WAS OBVIOUS THAT THE PRESSURE PUMP GAUGE WAS BROKEN, AND THE PRESSURE WAS SLOWLY PRESSURIZED WITHOUT ANY CHANGE IN THE PRESSURE VALUE AND THE SOLUTION WAS CHANGE THE SET OF BALLOONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19269132 ↗
  • MalfunctionApr 10, 2024Disconnection; Disconnection

    IT WAS REPORTED THAT THE DOCTOR WAS PERFORMING URETERAL STRICTURE DILATION SURGERY, THE RELEVANT PREPARATIONS FOR THE OPERATION HAD BEEN COMPLETED, AND THE SURGICAL CONSUMABLES WERE OPENED STEP BY STEP ACCORDING TO THE SURGICAL STEPS, AND AFTER THE U30 WAS OPENED, IT WAS FOUND THAT THE PRESSURE PUMP GAUGE FELL OFF AND COULD NOT BE USED. THE DOCTOR REOPENED A SET OF U30S TO COMPLETE THE OPERATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19080237 ↗
  • MalfunctionNov 20, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE EAGLE INFLATION DEVICE WAS NOT WORKING AT ALL AND WAS LEAKING WATER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18169806 ↗
  • MalfunctionOct 31, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT UPON UNPACKING THE PACKAGE DURING THE OPERATION, IT WAS FOUND THAT THE PUMP OF THE BALLOON DILATION CATHETER HAD BEEN BROKEN AND DAMAGED, AND THEN IT WITHDREW AND REPLACED WITH A NEW PACKAGING PRODUCT TO CONTINUE THE OPERATION, WITHOUT ADVERSE EFFECTS ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18039287 ↗
  • MalfunctionJul 5, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE PRESSURIZATION PUMP FOR THE X-FORCE U30 URETEROSCOPIC BALLOON DILATION CATHETER LEAKED WATER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17258444 ↗
  • MalfunctionSep 14, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE PRESSURE PUMP IN THE EAGLE INFLATION DEVICE LEAKED WATER WHEN IN USE BY THE CLINICIAN. IT WAS MENTIONED THAT A NEW PRESSURE PUMP WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15419689 ↗
  • MalfunctionOct 2, 2020Device Handling Problem; Inflation Problem; Device Handling Problem

    IT WAS REPORTED THAT THE INFLATION DEVICE PUMP TO INFLATE THE BALLOON DID NOT WORK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10623201 ↗
  • MalfunctionFeb 5, 2020Break; Inflation Problem; Leak/Splash; Inflation Problem

    IT WAS REPORTED THAT THE EAGLE INFLATION DEVICE WAS FOUND TO BE DAMAGED UPON OPENING THE PACKAGE. THE PUMP LEAKED DURING PRETEST.

    Patient problems: No Patient Involvement

    MAUDE report 9671927 ↗
  • MalfunctionOct 10, 2019Fluid/Blood Leak; Inflation Problem; Use of Device Problem; Use of Device Problem

    IT WAS REPORTED LIQUID WAS LEAKING FROM THE DEVICE. THE BALLOON DID NOT INFLATE.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 9174758 ↗