Bard Peripheral Vascular, Inc. · TRUGUIDE® Disposable Coaxial Biopsy Needle with Depth Stop, 15g x 13.8cm · Class 2

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00801741084966), received between Dec 22, 2023 and Feb 17, 2026 — classified as 14 malfunctions.

The most frequently reported specific device problem is “Device Markings/Labelling Problem” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 7 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Markings/Labelling Problem9
  • Break3
  • Component Misassembled2
  • Device Contamination with Chemical or Other Material2
  • Device-Device Incompatibility2

Patient problems

  • No Clinical Signs, Symptoms or Conditions14

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 17, 2026Device Markings/Labelling Problem

    ON (B)(6) 2026, A PATIENT UNDERWENT A COMPUTED TOMOGRAPHY BIOPSY PROCEDURE USING THE MISSION AND COAXIAL. DURING THE PROCEDURE, THE DEVICE COAXIAL THAT WAS PACKAGED/LABELED INCORRECTLY. FURTHER IT WAS REPORTED THAT LABELED AS A FIFTEEN G BIOPSY CANNULA BUT WAS UNABLE TO FIT 1616MS DOWN IT. COAXIAL NEEDLE WAS USED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24361873 ↗
  • MalfunctionSep 23, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDE KIDNEY BIOPSY PROCEDURE VIA NORMAL DENSITY TISSUE USING THE COAXIAL. DURING THE PROCEDURE, THE DEVICE MISMATCHED WITH THE SIZE OF BIOPSY NEEDLE DUE TO WRONG LABELING OF COAXIAL NEEDLE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23120570 ↗
  • MalfunctionSep 23, 2025Device Markings/Labelling Problem; Device-Device Incompatibility; Device Markings/Labelling Problem; Device-Device Incompatibility

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDE KIDNEY BIOPSY PROCEDURE VIA NORMAL DENSITY TISSUE USING THE COAXIAL. DURING THE PROCEDURE, THE DEVICE MISMATCHED WITH THE SIZE OF BIOPSY NEEDLE DUE TO WRONG LABELING OF COAXIAL NEEDLE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23120430 ↗
  • MalfunctionSep 23, 2025Device Markings/Labelling Problem; Device-Device Incompatibility; Device Markings/Labelling Problem; Device-Device Incompatibility

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ULTRASOUND GUIDE KIDNEY BIOPSY PROCEDURE VIA NORMAL DENSITY TISSUE USING THE COAXIAL. DURING THE PROCEDURE, THE DEVICE MISMATCHED WITH THE SIZE OF BIOPSY NEEDLE DUE TO WRONG LABELING OF COAXIAL NEEDLE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23120427 ↗
  • MalfunctionSep 18, 2025Break; Break

    ON (B)(6) 2025, A PATIENT UNDERWENT A COMPUTED TOMOGRAPHY-GUIDED LYMPH BIOPSY THROUGH NORMAL DENSITY TISSUE USING A TRUGUIDE DEVICE. DURING THE PROCEDURE, THE COAXIAL NEEDLE WAS ALLEGEDLY FRACTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WERE NO REPORTED PATIENT INJURIES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23082588 ↗
  • MalfunctionSep 18, 2025Break; Break

    ON (B)(6) 2025, A PATIENT UNDERWENT A CT GUIDED LYMPH BIOPSY PROCEDURE USING NORMAL DENSITY TISSUE WITH TRUGUIDE DEVICE. DURING THE PROCEDURE, THE COAXIAL NEEDLE WAS ALLEGEDLY FRACTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23082359 ↗
  • MalfunctionSep 18, 2025Break; Break

    ON (B)(6) 2025, A PATIENT UNDERWENT A CT GUIDED LYMPH BIOPSY PROCEDURE USING NORMAL DENSITY TISSUE WITH TRUGUIDE DEVICE. DURING THE PROCEDURE, THE COAXIAL NEEDLE WAS ALLEGEDLY FRACTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23082358 ↗
  • MalfunctionSep 2, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem

    ON (B)(6) 2025, A PATIENT UNDERWENT A COMPUTED TOMOGRAPHY-GUIDED RENAL BIOPSY PROCEDURE UTILIZING THE MAXCORE AND TRUEGUIDE SYSTEMS. DURING THE PROCEDURE, THE DEVICE WAS FOUND TO BE IN IMPROPER PACKAGING. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN ALTERNATIVE DEVICE. THERE WERE NO REPORTED INJURIES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22936213 ↗
  • MalfunctionAug 18, 2025Device Markings/Labelling Problem; Component Misassembled; Device Markings/Labelling Problem; Component Misassembled

    ON (B)(6) 2025, RECEIVED A COMPLAINT DEFINES THAT A PATIENT UNDERWENT A BIOPSY PROCEDURE USING THE TRUGUIDE COAXIAL NEEDLE. PRIOR THE PROCEDURE, THE PRODUCT WAS NOT APPROPRIATE AS IT WAS MUCH SHORTER THAN 15G. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22826815 ↗
  • MalfunctionAug 18, 2025Device Markings/Labelling Problem; Component Misassembled; Device Markings/Labelling Problem; Component Misassembled

    ON (B)(6) 2025, RECEIVED A COMPLAINT DEFINES THAT A PATIENT UNDERWENT A BIOPSY PROCEDURE USING THE TRUGUIDE COAXIAL NEEDLE. PRIOR THE PROCEDURE, THE PRODUCT WAS NOT APPROPRIATE AS IT WAS MUCH SHORTER THAN 15G. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22826813 ↗
  • MalfunctionJun 18, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem

    ON (B)(6) 2025, A PATIENT UNDERWENT A BIOPSY PROCEDURE USING THE TRUGUIDE. PRIOR TO THE PROCEDURE, IT WAS NOTED THE INCORRECT LENGTH TRUGUIDE WAS PACKAGED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22278453 ↗
  • MalfunctionJun 7, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem

    ON (B)(6) 2025, A PATIENT UNDERWENT A BIOPSY PROCEDURE USING THE TRUGUIDE. PRIOR TO THE PROCEDURE, IT WAS NOTED THE INCORRECT LENGTH TRUGUIDE WAS PACKAGED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22163982 ↗
  • MalfunctionMar 26, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT PRIOR TO A RENAL BIOPSY PROCEDURE, AN IMPURITY WAS ALLEGEDLY FOUND IN THE COAXIAL BIOPSY NEEDLE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18984407 ↗
  • MalfunctionDec 22, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT PRIOR TO A BIOPSY PROCEDURE, THE DEVICE ALLEGEDLY HAD A FOREIGN PARTICLE INSIDE THE PRIMARY CONTAINER. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18381845 ↗

Other Biopsy needle devices

All Biopsy needle devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.