TRUGUIDE C1820BRecalls & reported problems
Bard Peripheral Vascular, Inc. · TRUGUIDE® Disposable Coaxial Biopsy Needle with Depth Stop and Additional Blunt Tip Stylet, 17g x 17cm · Class 2
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741085062), received between May 30, 2018 and Sep 21, 2026 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Device Markings/Labelling Problem2
- Failure to Obtain Sample1
- Illegible Information1
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Foreign Body In Patient1
- Injury1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2026Illegible Information
On (b)(6) 2026, a patient underwent a biopsy procedure using a TruGuide coaxial needle. During the procedure, it was found that the needle size and other information on the packaging were printed in a very small font, making the label difficult to read and requiring magnification to clearly identify the needle size. It was further reported that previous packaging had larger, more readable text and that all needles in the latest shipment appeared to have a new label with a smaller font. The BD labeled packaging was reported to have a much smaller font than the Bard labeled packaging, which remained easily readable. The complainant expressed concern that the reduced font size could increase the risk of selecting the incorrect needle size. The affected devices were used on patients, as the concern was limited to the labeling. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50836525 ↗MalfunctionFeb 27, 2026Device Markings/Labelling Problem
ON (B)(6) 2025, A PATIENT UNDERWENT AN CT GUIDED BIOPSY PROCEDURE BY USING THE TRUGUIDE COAXIAL. IT WAS REPORTED THAT DURING THE PROCEDURE, THE DEVICE ALLEGEDLY FAILED TO OBTAIN SAMPLE. IT WAS FURTHER REPORTED THAT LABELING MISMATCH, THE YELLOW BARCODE LABELS INDICATED C1816A, BUT THE ACTUAL MODEL WAS C1820B. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24466395 ↗MalfunctionOct 31, 2025Failure to Obtain Sample; Device Markings/Labelling Problem; Failure to Obtain Sample; Device Markings/Labelling Problem
ON (B)(6) 2025, A PATIENT UNDERWENT AN CT GUIDED BIOPSY PROCEDURE USING THE TRUGUIDE COAXIAL. DURING PROCEDURE, THE DEVICE ALLEGEDLY FAILED TO OBTAIN SAMPLE. IT WAS FURTHER REPORTED THAT LABELING MISMATCH, THE YELLOW BARCODE LABELS INDICATED C1816A, BUT THE ACTUAL MODEL WAS C1820B.THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434369 ↗InjuryMay 22, 2020Break; Break
IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BIOPSY, THE COAXIAL ALLEGEDLY BROKE INSIDE THE PATIENT. THE DEVICE WAS REMOVED IN AN ADDITIONAL SURGERY OPERATION. THE PATIENT IS REPORTEDLY STABLE.
Patient problems: Injury
MAUDE report 10082873 ↗MalfunctionJul 2, 2018Break; Break
IT WAS REPORTED THAT POST ULTRASOUND-GUIDED LIVER BIOPSY, THE DISTAL TIP OF THE COAXIAL CANNULA ALLEGEDLY APPEARED TRUNCATED UPON REMOVAL FROM THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7656930 ↗MalfunctionMay 30, 2018Material Fragmentation
AT TERMINATION OF AN ULTRASOUND GUIDED LIVER BIOPSY PROCEDURE AND REMOVAL OF COAXIAL NEEDLE, THE DISTAL TIP OF THE OUTER COAXIAL CANNULA APPEARED TO BE TRUNCATED. THEREFORE THERE WAS CONCERN FOR A RETAINED FRAGMENT.
Patient problems: Foreign Body In Patient
MAUDE report 7552842 ↗
Other Biopsy needle devices
- MAGNUM MN1410Bard Peripheral Vascular, Inc. · Biopsy needle1 recall118 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4021Olympus Medical Systems Corp. · Biopsy needle1 recall94 reports
- EchoTip G31520Cook Ireland Ltd · Biopsy needle1 recall84 reports
- Single Use Aspiration Needle NA-201SX-4021Olympus Medical Systems Corp. · Biopsy needle1 death126 reports
- EchoTip ProCore G34785Cook Ireland Ltd · Biopsy needle97 reports
- MAGNUM MN1820Bard Peripheral Vascular, Inc. · Biopsy needle1 recall37 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4022Olympus Medical Systems Corp. · Biopsy needle1 recall29 reports
- Interject M00518250Boston Scientific Corporation · Biopsy needle70 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.