TRUGUIDE C2016ARecalls & reported problems
Bard Peripheral Vascular, Inc. · TRUGUIDE® Disposable Coaxial Biopsy Needle with Depth Stop and Additional Blunt Tip Stylet, 19g x 13.8cm · Class 2
All models of TRUGUIDE (14)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00801741085109), received between Jun 28, 2018 and Oct 9, 2025 — classified as 6 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
- Device Markings/Labelling Problem5
- Adverse Event Without Identified Device or Use Problem3
- Material Twisted/Bent2
- Detachment of Device or Device Component1
- Device-Device Incompatibility1
- Improper or Incorrect Procedure or Method1
- Migration or Expulsion of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Pneumothorax3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
InjuryOct 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A LUNG BIOPSY PROCEDURE USING THE TRUGUIDE. DURING THE PROCEDURE, PATIENT EXPERIENCING PNEUMOTHORAX. HOWEVER, THE CURRENT PATIENT STATUS UNKNOWN.
Patient problems: Pneumothorax
MAUDE report 23263145 ↗InjuryOct 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A LUNG BIOPSY PROCEDURE USING THE TRUGUIDE. DURING THE PROCEDURE, PATIENT EXPERIENCING PNEUMOTHORAX. HOWEVER, THE CURRENT PATIENT STATUS UNKNOWN.
Patient problems: Pneumothorax
MAUDE report 23263144 ↗InjuryOct 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A LUNG BIOPSY PROCEDURE USING THE TRUGUIDE. DURING THE PROCEDURE, PATIENT EXPERIENCING PNEUMOTHORAX. HOWEVER, THE CURRENT PATIENT STATUS UNKNOWN.
Patient problems: Pneumothorax
MAUDE report 23263143 ↗InjurySep 11, 2025Break; Improper or Incorrect Procedure or Method; Break; Improper or Incorrect Procedure or Method
ON (B)(6) 2025, A PATIENT UNDERWENT A COMPUTED TOMOGRAPHY GUIDE PELVIC ABSCESS BIOPSY USING THE COAXIAL BIOPSY NEEDLE. DURING THE PROCEDURE, IT WAS REPORTED THAT THE NEEDLE ALLEGEDLY BROKE AT 4 CM. THE 8CM NEEDLE PIECE REMAINED INSIDE THE PATIENT AND REQUIRED SURGICAL INTERVENTION TO REMOVE. HOWEVER, THE CURRENT PATIENT STATUS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 23023585 ↗MalfunctionAug 4, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A COMPUTED TOMOGRAPHY GUIDE LUNG BIOPSY PROCEDURE VIA SOFT DENSITY TISSUE USING THE TRUEGUIDE DEVICE. PRIOR THE PROCEDURE, THE PACKAGING AND GOODS ARE NOT MATCHING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22697280 ↗MalfunctionAug 4, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A COMPUTED TOMOGRAPHY GUIDE LUNG BIOPSY PROCEDURE VIA SOFT DENSITY TISSUE USING THE TRUEGUIDE DEVICE. PRIOR THE PROCEDURE, THE PACKAGING AND GOODS ARE NOT MATCHING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22697278 ↗MalfunctionAug 4, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A COMPUTED TOMOGRAPHY GUIDE LUNG BIOPSY PROCEDURE VIA SOFT DENSITY TISSUE USING THE TRUEGUIDE DEVICE. PRIOR THE PROCEDURE, THE PACKAGING AND GOODS ARE NOT MATCHING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22697277 ↗MalfunctionAug 4, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A COMPUTED TOMOGRAPHY GUIDE LUNG BIOPSY PROCEDURE VIA SOFT DENSITY TISSUE USING THE TRUEGUIDE DEVICE. PRIOR THE PROCEDURE, THE PACKAGING AND GOODS ARE NOT MATCHING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22697276 ↗MalfunctionAug 4, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A COMPUTED TOMOGRAPHY GUIDE LUNG BIOPSY PROCEDURE VIA SOFT DENSITY TISSUE USING THE TRUEGUIDE DEVICE. PRIOR THE PROCEDURE, THE PACKAGING AND GOODS ARE NOT MATCHING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22697275 ↗MalfunctionJul 15, 2024Break; Break
IT WAS REPORTED THAT DURING A COMPUTED TOMOGRAPHY GUIDED LUNG BIOPSY PROCEDURE THROUGH NORMAL DENSITY TISSUE, THE STYLET PART OF THE NEEDLE ALLEGEDLY BROKE OFF INSIDE THE PATIENT'S BODY. REPORTEDLY, THE BROKEN FRAGMENTS OF THE DEVICE WERE REMOVED FROM THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19747154 ↗InjuryMay 4, 2023Break; Material Twisted/Bent; Break; Material Twisted/Bent
IT IS REPORTED THAT DURING A BIOPSY PROCEDURE, COAXIAL NEEDLE WAS ALLEGEDLY KINKED. IT WAS FURTHER REPORTED THAT THE NEEDLE ALLEGEDLY FRACTURED. REPORTEDLY NEEDLE FRAGMENT WAS DISSECTED OUT WITH INCISION AND REMOVED WITH HEMOSTAT. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 16866113 ↗MalfunctionFeb 15, 2022Break; Break
IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE DISTAL END OF THE NEEDLE ALLEGEDLY BROKE INSIDE THE PATIENT. THE CURRENT PATIENT STATUS WAS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13534685 ↗MalfunctionFeb 14, 2022Break; Break
IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE DISTAL PIECE OF THE DEVICE ALLEGEDLY BROKE DURING ADVANCEMENT. THE BROKEN PIECE CAME OUT OF THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13521366 ↗MalfunctionJul 2, 2021Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
IT WAS REPORTED THAT PRIOR TO A LUNG BIOPSY, THE DEVICE ALLEGEDLY HAD INCOMPATIBILITY ISSUE AND THE NEEDLE TIP WAS ALLEGEDLY BROKEN WHILE FORCEFUL INSERTION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12108100 ↗InjuryJun 28, 2018Break; Migration or Expulsion of Device; Detachment of Device or Device Component; Material Twisted/Bent; Break; Migration or Expulsion of Device; Detachment of Device or Device Component; Material Twisted/Bent
IT WAS REPORTED THAT DURING A BIOPSY, THE COAXIAL INTRODUCER FRACTURED INSIDE THE PATIENT. IT WAS FURTHER REPORTED THAT THE FRAGMENT WAS REMOVED THROUGH THE PATIENT'S ABDOMEN VIA SURGICAL INTERVENTION. POST SURGERY, THE PATIENT'S CONDITION WAS STABLE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7648894 ↗
FDA clearance and classification
The FDA regulates product code FCG (biopsy needle) under 21 CFR 876.1075 as a Class II device. 21 CFR 876.1075 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the TRUGUIDE C2016A
Has the TRUGUIDE C2016A been recalled?
No FDA recall record in our data names this exact device identifier (DI 00801741085109), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the TRUGUIDE C2016A?
The FDA's MAUDE database lists 15 adverse event reports for this device, received between Jun 28, 2018 and Oct 9, 2025. In those reports, the most common specific device problems are “Break” (7), “Device Markings/Labelling Problem” (5) and “Material Twisted/Bent” (2); the most common patient problem is “Pneumothorax” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the TRUGUIDE C2016A?
None of the reports are classified as deaths, 6 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the TRUGUIDE C2016A, and is it still sold?
The FDA's device database lists Bard Peripheral Vascular, Inc. as the company responsible for it (brand TRUGUIDE, model C2016A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Biopsy needle devices
- MAGNUM MN1410Bard Peripheral Vascular, Inc. · Biopsy needle1 recall119 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4021Olympus Medical Systems Corp. · Biopsy needle1 recall94 reports
- EchoTip G31520Cook Ireland Ltd · Biopsy needle1 recall84 reports
- Single Use Aspiration Needle NA-201SX-4021Olympus Medical Systems Corp. · Biopsy needle1 death126 reports
- EchoTip ProCore G34785Cook Ireland Ltd · Biopsy needle97 reports
- MAGNUM MN1820Bard Peripheral Vascular, Inc. · Biopsy needle1 recall37 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4022Olympus Medical Systems Corp. · Biopsy needle1 recall28 reports
- Interject M00518250Boston Scientific Corporation · Biopsy needle70 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page