ENDOTAK RELIANCE® 0265Recalls & reported problems

Boston Scientific Corporation · Steroid-eluting Bipolar Pace/Sense and Defibrillation Lead · Class 3 · Implant

All models of ENDOTAK RELIANCE (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00802526431289), received between Jul 11, 2017 and Dec 22, 2025 — classified as 4 injuries, 4 malfunctions.

The most frequently reported specific device problem is “High impedance” (6 reports).

3 other recalls cover devices cleared under the same FDA submission (P910073); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High impedance6
  • High Capture Threshold2
  • Signal Artifact/Noise2
  • Adverse Event Without Identified Device or Use Problem1
  • Ambient Noise Problem1
  • Defective Device1
  • Fracture1
  • Over-Sensing1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient2
  • Bradycardia1
  • Unspecified Infection1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 22, 2025High impedance; High impedance

    IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED OUT OF RANGE PACING IMPEDANCE MEASUREMENTS ON THE RIGHT VENTRICULAR (RV) LEAD. THE PATIENT WAS BROUGHT INTO THE CLINIC AND ISOMETRIC TESTING WAS PERFORMED. THE RV PACE THRESHOLDS AND SENSING WERE STABLE AND NO NOISE WAS NOTED ON THE LEAD OR IN ANY STORED EVENTS. THE PLAN IS CONTINUE MONITORING. CURRENTLY, THIS PRODUCT REMAINS IN SERVICE AND NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23873469 ↗
  • MalfunctionFeb 25, 2025High impedance; High impedance

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH OUT OF RANGE IMPEDANCE MEASUREMENTS, WITH THE INCREASE BEEN GRADUAL. TECHNICAL SERVICES (TS) WAS CONSULTED AND DISCUSSED THERAPY DELIVERY AS WELL AS DEVICE PROGRAMMING. PATIENT WILL CONTINUE TO BE MONITORED. THIS RV LEAD REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21459872 ↗
  • MalfunctionNov 21, 2024High impedance; High Capture Threshold; High impedance; High Capture Threshold

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH OUT OF RANGE PACE IMPEDANCE MEASUREMENTS. THRESHOLD MEASUREMENTS WERE ALSO NOTED TO HAVE INCREASED SLIGHTLY TO 2V. TECHNICAL SERVICES (TS) WAS CONSULTED AND DISCUSSED THIS COULD BE DUE TO CHANGES IN LEAD TISSUE INTERFACE WITH THE TIP ELECTRODE DUE TO SCAR TISSUE OF CALCIFICATION. TS RECOMMENDED TO CONTUE TO MONITOR. NO ADVERSE PATIENT EFFECTS WERE REPORTED. AT THIS TIME, THIS LEAD REMAINS IN SERVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20749925 ↗
  • InjuryJan 2, 2024Signal Artifact/Noise; Over-Sensing; Pacing Problem; Defective Device; Signal Artifact/Noise; Over-Sensing; Pacing Problem; Defective Device

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR LEAD WAS EXPLANTED DUE TO EXHIBITING NOISE, OVERSENSING AND PACING INHIBITION. INSULATION DAMAGE ALONG THE LEAD BODY WAS SUSPECTED. ANOTHER LEAD WAS IMPLANTED INSTEAD. NO FURTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Bradycardia

    MAUDE report 18431217 ↗
  • InjuryNov 17, 2023Signal Artifact/Noise; High impedance; High Capture Threshold; Signal Artifact/Noise; High impedance; High Capture Threshold

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS EXPLANTED DUE TO HIGH IMPEDANCES OUT OF RANGE ON SEVERAL CHECKS, ELEVATED THRESHOLDS, AND NOISE SIGNALS. THE PHYSICIAN ELECTED TO REPLACE THE RV LEAD WITH A NEW ONE AND SURGICALLY ABANDONED THE EXISTING. A FLUOROSCOPY WAS PERFORMED AT IMPLANT, BUT THERE WAS NOTHING DETECTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18162049 ↗
  • InjuryNov 9, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS PART OF A SYSTEM REVISION DUE TO INFECTION. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED. THE RV LEAD WAS EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 15759058 ↗
  • InjuryMar 14, 2019High impedance; High impedance

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS SURGICALLY CAPPED AND REPLACED DUE TO OBSERVATIONS OF HIGH OUT OF RANGE IMPEDANCES GREATER THAN 2000 OHMS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8421935 ↗
  • MalfunctionJul 11, 2017Fracture; High impedance; Ambient Noise Problem; Fracture; High impedance; Ambient Noise Problem

    BOSTON SCIENTIFIC RECEIVED INFORMATION FROM THE WARRANTY DEPARTMENT THAT THIS LOCAL REPRESENTATIVE WAS CONTACTED BY A HOSPITAL STAFF MEMBER VIA PHONE MESSAGE. THE STAFF MEMBER WAS INQUIRING ABOUT BOSTON SCIENTIFIC'S INDIGENT PROGRAM. THE PATIENT WAS HOSPITALIZED AND HAS NO INSURANCE. THE PATIENT HAS A NON-BOSTON SCIENTIFIC DEVICE AND A BOSTON SCIENTIFIC LEAD THAT HAS MALFUNCTION. THE PHYSICIAN INTENDS TO REPLACE THE IMPLANT LEAD AND THE HOSPITAL WAS LOOKING FOR AN INDIGENT DONATION. NO PATIENT NAME OR IMPLANTED PRODUCT INFORMATION WAS PROVIDED TO THE LOCAL REPRESENTATIVE. THE SPECIFIC LEAD MALFUNCTION WAS NOT IDENTIFIED AND THE LEAD CURRENTLY REMAINS IN-SERVICE. THE LOCAL REPRESENTATIVE WAS TRANSFERRED TO THE PERSON IN CUSTOMER SERVICES WHO COORDINATES THE INDIGENT PROGRAM. BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE LOCAL REPRESENTATIVE WAS ABLE TO IDENTIFY THE PATIENT NAME AND MODEL#/SN# OF THE IMPLANTABLE RIGHT VENTRICULAR (RV) DEFIBRILLATION LEAD. ACCORDING TO THE LOCAL REPRESENTATIVE, THIS RV DEFIBRILLATION LEAD WAS EXHIBITING HIGH RV PACING IMPEDANCE AND ELECTROGRAM NOISE. THE PHYSICIAN SUSPECTS A LEAD CONDUCTOR FRACTURE. THE LOCAL REPRESENTATIVE HAS NOT BEEN NOTIFIED OF

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6701061 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P910073. One submission often covers many models, so check the product and lot details in each record.

  • Z-2326-2025 · Jul 24, 2025 · Open, Classified
    ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296
  • Z-2327-2025 · Jul 24, 2025 · Open, Classified
    RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692,
  • Z-2325-2025 · Jul 24, 2025 · Open, Classified
    ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o18

Other Permanent defibrillator electrodes devices

All Permanent defibrillator electrodes devices →