RELIANCE 4-FRONT™ 0650Recalls & reported problems
Boston Scientific Corporation · Steroid-eluting Bipolar Pace/Sense and Defibrillation Lead · Class 3 · Implant
All models of RELIANCE 4-FRONT (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00802526592843), received between Oct 12, 2021 and Aug 29, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
3 other recalls cover devices cleared under the same FDA submission (P910073); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture1
- High impedance1
- Nonstandard Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 29, 2025Fracture; Nonstandard Device; Fracture; Nonstandard Device
IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS EXPLANTED DUE TO LEAD WAS FRACTURED. A NEW LEAD WITH WAS IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. ADDITIONAL INFORMATION INDICATES THAT THE RV LEAD FRACTURE WAS CONFIRMED BY LEAD MEASUREMENTS AND AN X-RAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22920756 ↗MalfunctionOct 12, 2021High impedance; High impedance
IT WAS REPORTED THAT THIS PATIENT'S REMOTE COMMUNICATOR DISPLAYED A RED CALL DOCTOR ICON. SUBSEQUENTLY, TROUBLESHOOTING WAS PERFORMED, AND THE COMMUNICATOR WAS POWER CYCLED, AND A SUCCESSFUL PATIENT-INITIATED INTERROGATION WAS SENT. THE PATIENT WAS REFERRED TO CONTACT THEIR CLINIC REGARDING THE RED CALL DOCTOR ICON. ADDITIONAL INFORMATION WAS RECEIVED THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED A HIGH OUT OF RANGE RIGHT VENTRICULAR (RV) LEAD PACING IMPEDANCE MEASUREMENT, MEASURING GREATER THAN 2000 OHMS. SUBSEQUENTLY, THIS PATIENT WAS BROUGHT INTO CLINIC FOR FURTHER TROUBLESHOOTING. THE PHYSICIAN OPTED TO TURN TACHY THERAPIES OFF FOR THIS PATIENT AT THAT TIME. AT THIS TIME, EVIDENCE SUGGESTS THE DEVICE SYSTEM REMAINS IN-SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12620379 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P910073. One submission often covers many models, so check the product and lot details in each record.
- Z-2326-2025 · Jul 24, 2025 · Open, ClassifiedENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296
- Z-2327-2025 · Jul 24, 2025 · Open, ClassifiedRELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692,
- Z-2325-2025 · Jul 24, 2025 · Open, ClassifiedENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o18
Other Permanent defibrillator electrodes devices
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes333 deaths30,581 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M62Medtronic, Inc. · Permanent defibrillator electrodes221 deaths14,405 reports
- Capsurefix® Novus 5076-52Medtronic, Inc. · Permanent defibrillator electrodes111 deaths9,860 reports
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes80 deaths8,280 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M55Medtronic, Inc. · Permanent defibrillator electrodes116 deaths7,750 reports
- Durata™ 7122Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes151 deaths6,202 reports
- Durata™ 7122Q-65St. Jude Medical, Inc. · Permanent defibrillator electrodes86 deaths4,773 reports
- Capsurefix® Novus 5076-58Medtronic, Inc. · Permanent defibrillator electrodes47 deaths4,587 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026