ENDOTAK RELIANCE® S 640273-103Recalls & reported problems

Boston Scientific Corporation · Integrated Bipolar Pace/Sense and Defibrillation Lead · Class 3 · Implant

All models of ENDOTAK RELIANCE S (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Boston Scientific Endotak Reliance S is an implanted defibrillation lead, a wire that runs into the heart to deliver shocks and pacing. Its 385 reports include 2 deaths and 298 injuries, and many describe infections, sometimes blood infections, that led to the lead being removed. Other reports describe leads moving out of place and high electrical impedance. Reports fell to 15 over the last year from 34 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
391
Deaths reported
2
Injuries reported
303
Recalls naming it
0

Summary

The FDA's MAUDE database contains 391 adverse event reports that list this exact device (DI 00802526601811), received between Jun 18, 2019 and Sep 29, 2026 — classified as 2 deaths, 303 injuries, 86 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (54 reports).

3 other recalls cover devices cleared under the same FDA submission (P910073); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 8 reportsDec 2023: 9 reportsJan 2024: 4 reportsFeb 2024: 9 reportsMar 2024: 7 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 8 reportsJul 2024: 5 reportsAug 2024: 1 reportsSep 2024: 3 reportsOct 2024: 9 reportsNov 2024: 3 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 2 reportsMar 2025: 5 reportsApr 2025: 2 reportsMay 2025: 5 reportsJun 2025: 3 reportsJul 2025: 4 reportsAug 2025: 5 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 4 reportsFeb 2026: 4 reportsMar 2026: 4 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem170
  • High Capture Threshold54
  • High impedance54
  • Device Dislodged or Dislocated45
  • Over-Sensing40
  • Signal Artifact/Noise31
  • Failure to Capture27
  • Pacing Problem21

Patient problems

  • Unspecified Infection156
  • No Clinical Signs, Symptoms or Conditions147
  • Sepsis45
  • No Code Available36
  • No Known Impact Or Consequence To Patient17
  • Bacterial Infection10
  • Pocket Erosion10
  • Electric Shock6

Recent adverse event reports

The 25 most recent of 391 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2026Insufficient Device Problem Information

    It was reported that this right ventricular (RV) lead exhibited an unknown product performance anomaly. A revision procedure was performed. This RV lead was surgically abandoned and replaced with a new lead successfully. No additional adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927562 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this right ventricular (RV) lead was part of a system revision due to infection. There were no additional adverse patient effects reported. The RV lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50919653 ↗
  • InjurySep 16, 2026Insufficient Device Problem Information

    It was reported that this right ventricular (RV) lead was surgically abandoned due to unknown reasons. This RV lead was successfully replaced and no additional adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50789024 ↗
  • MalfunctionSep 11, 2026Signal Artifact/Noise; Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50736139 ↗
  • InjuryJul 28, 2026High impedance

    It was reported that this right ventricular (RV) lead exhibited high out of range shock impedance measurements. Technical Services (TS) was consulted as the lead was scheduled for replacement but for an unknown reason. TS reviewed the lead and determined there were multiple red alerts for out of range spikes with daily variability in testing. TS noted that the lead was stable since implant in impedance measurements and only noted a few spikes. Subsequently, the lead was explanted, and a new lead was successfully implanted. No additional adverse patient effects were reported. This RV lead is no longer in service.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50060477 ↗
  • InjuryJul 20, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this right ventricular (RV) lead was part of a system revision due to infection. There were no additional adverse patient effects reported. The RV lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 25883619 ↗
  • InjuryMay 29, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS IMPLANTABLE LEAD WAS PART OF THE SYSTEM REVISION DUE TO INFECTION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE IMPLANTABLE LEAD WAS EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 25327494 ↗
  • InjuryMar 12, 2026High impedance; Insufficient Device Problem Information

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS EXPLANTED DUE TO A PRODUCT PERFORMANCE ISSUE. THIS RV LEAD WAS SUCCESSFULLY EXPLANTED AND REPLACED WITH ANOTHER LEAD. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED OUTSIDE THE REVISION PROCEDURE. ADDITIONAL INFORMATION WAS RECEIVED THAT THIS RV LEAD WAS EXPLANTED DUE TO HIGH OUT OF RANGE PACING IMPEDANCE MEASUREMENTS. AS THIS LEAD WAS DAMAGED DURING THE EXTRACTION, IT IS NOT EXPECTED TO RETURN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED OUTSIDE THE REVISION PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24585991 ↗
  • InjuryMar 10, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS PART OF A SYSTEM REVISION DUE TO INFECTION. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED. THIS RV LEAD WAS EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 24565031 ↗
  • InjuryMar 9, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD POKED HOLE IN THE TRICUSPID AREA. THIS RV LEAD WAS EXPLANTED AND REPLACED WITH A DIFFERENT MODEL. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. ADDITIONAL INFORMATION INDICATED THAT THIS RV LEAD EXHIBITED NO PERFORATION. THE PATIENT HAD MILD SYMPTOMS. THIS RV LEAD WAS KEPT BY THE HOSPITAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24552626 ↗
  • InjuryMar 3, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS IMPLANTABLE LEAD WAS PART OF THE SYSTEM REVISION DUE TO INFECTION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE IMPLANTABLE LEAD WAS EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 24494015 ↗
  • InjuryFeb 20, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS PART OF A SYSTEM REVISION DUE TO INFECTION. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED. THIS RV LEAD WAS EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 24405051 ↗
  • InjuryFeb 17, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS PART OF A SYSTEM REVISION DUE TO INFECTION. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED. THIS RV LEAD WAS EXPLANTED. ADDITIONAL INFORMATION INDICATED THAT THE PATIENT HAD BLOOD INFECTION. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED.

    Patient problems: Sepsis; Unspecified Infection

    MAUDE report 24364628 ↗
  • MalfunctionFeb 11, 2026Failure to Capture

    IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) HAD A DECREASE ON ITS PACING PERCENTAGE, SO A REVIEW WAS REQUESTED. TECHNICAL SERVICES (TS) WAS CONSULTED AND REVIEWED STORED DATA. TS DISCUSSED HOW PREMATURE VENTRICULAR CONTRACTIONS (PVC) RIGHT AFTER ATRIAL PACING WERE SUSPECTED. THIS RESULTED IN THE PATIENTS RIGHT VENTRICULAR SIGNALS BEEN BLANKED AND PACING BEEN DELIVERED DURING THE RIGHT VENTRICLE REFRACTORY PERIOD. LOSS OF CAPTURE (LOC) WAS SUSPECTED FOR THE RV BASED ON AVAILABLE INFORMATION. TS DISCUSSED PROGRAMMING OPTIONS FOR THIS DEVICE. THIS DEVICE REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24314524 ↗
  • InjuryFeb 5, 2026Under-Sensing

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING. THE PATIENT HAD A LEFT VENTRICULAR ASSIST DEVICE (LVAD) IMPLANTED RECENTLY WHICH WAS WHEN THE UNDERSENSING WAS NOTED. SENSITIVITY WAS INITIALLY INCREASED BUT DID NOT IMPROVE SENSING. THE RV LEAD WAS SURGICALLY ABANDONED AND REPLACED WITH A NON-BOSTON SCIENTIFIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LEAD FOR BIPOLAR SENSING. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. THIS RV LEAD IS NO LONGER IN SERVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24272450 ↗
  • InjuryJan 28, 2026Device Sensing Problem

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD EXHIBITED A SENSING ISSUE. DEVICE DATA REVIEW DETERMINED THERE WAS SEVERAL SHOCK AND ROUND OF ANTI-TACHYCARDIA PACING (ATP) DELIVERED. ALL DELIVERED WAS CONFIRMED TO BE APPROPRIATE. THIS LEAD WAS SUBSEQUENTLY DEACTIVATED AND REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24200895 ↗
  • InjuryJan 26, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS PART OF A SYSTEM REVISION DUE TO INFECTION. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED. THIS RV LEAD WAS EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 24179470 ↗
  • MalfunctionJan 20, 2026High impedance

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) DEFIBRILLATION LEAD EXHIBITED HIGH OUT-OF-RANGE SHOCK IMPEDANCE MEASUREMENTS GREATER THAN 125 OHMS. REVIEW OF SHOCK IMPEDANCE TRENDS SHOWED AN INITIAL RISE IN MEASUREMENTS FOLLOWED BY A PLATEAU, CONSISTENT WITH NORMAL LEAD MATURATION. THE AVERAGE SHOCK IMPEDANCE RANGE WAS 95-100 OHMS, WITH A FEW OUTLIERS. THIS RV LEAD REMAINS IN SERVICE. NORMAL FOLLOW UP AND CONTINUED MONITORING WAS RECOMMENDED AT THIS TIME. NO ADVERSE PATIENT EFFECTS OR INTERVENTIONS WERE REPORTED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24123455 ↗
  • InjuryJan 5, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS PART OF A SYSTEM REVISION DUE TO INFECTION. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED. THIS RV LEAD WAS EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 23960783 ↗
  • MalfunctionDec 30, 2025High impedance; High impedance

    IT WAS REPORTED THAT THE PATIENT WITH THIS RIGHT VENTRICULAR (RV) LEAD WAS EXHIBITING OUT OF RANGE SHOCK IMPEDANCE MEASUREMENTS, MEASURING GREATER THAN 125 OHMS. TECHNICAL SERVICES (TS) WAS CONSULTED AND OBSERVED THAT THE MEASUREMENT THAT WAS OUT OF RANGE APPEARS TO BE AN OUTLIER, AS THE OTHER SHOCK IMPEDANCE MEASUREMENTS WERE OBSERVED TO BE STABLE AND WITHIN RANGE, AND NOTED THAT THERE IS NO EVIDENCE OF NOISE. TS RECOMMENDED PERFORMING ISOMETRICS, POCKET MANIPULATION, AND SERIAL IMPEDANCE TESTS, AND IF NO NOISE IS OBSERVED, THE PATIENT MAY CONTINUE TO BE MONITORED. REPROGRAMMING CAN ALSO BE CONSIDERED. THIS RV LEAD REMAINS IN SERVICE AND NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23927439 ↗
  • InjuryDec 19, 2025Failure to Read Input Signal; Failure to Read Input Signal

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS EXPLANTED DUE TO LOW INTRINSIC AMPLITUDE. A NEW LEAD WAS IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23849452 ↗
  • InjuryOct 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS PART OF A SYSTEM REVISION DUE TO INFECTION. THERE WAS NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. THIS RV LEAD WAS EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 23351978 ↗
  • InjurySep 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS PART OF A SYSTEM REVISION DUE TO INFECTION. THERE WAS NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. THIS RV LEAD WAS EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 22986962 ↗
  • MalfunctionAug 26, 2025High impedance; Off-Label Use; High impedance; Off-Label Use

    IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) DISPLAYED A MESSAGE FOR OUT OF RANGE IMPEDANCE MEASUREMENTS WHEN ATTEMPTED TO BE PROGRAMMED INTO MAGNETIC RESONANCE IMAGING (MRI) PROTECTION MODE. TECHNICAL SERVICES (TS) WAS CONSULTED AND DISCUSSED AVAILABLE DATA. TS DISCUSSED HOW THE DEVICE SHOULD BE EVALUATED IN PERSON BEFORE PROCEEDING. THIS DEVICE REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22895287 ↗
  • InjuryAug 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS PART OF A SYSTEM REVISION DUE TO INFECTION. THERE WAS NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. THIS RV LEAD WAS EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 22836621 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P910073. One submission often covers many models, so check the product and lot details in each record.

  • Z-2325-2025 · Jul 24, 2025 · Open, Classified
    ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o18
  • Z-2327-2025 · Jul 24, 2025 · Open, Classified
    RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692,
  • Z-2326-2025 · Jul 24, 2025 · Open, Classified
    ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296

Other Permanent defibrillator electrodes devices

All Permanent defibrillator electrodes devices →