JuggerKnot CM-99115BRecalls & reported problems

Riverpoint Medical, LLC · Soft Anchor · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00810020087154), received between Jun 16, 2022 and Jan 30, 2025 — classified as 18 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 3 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated7
  • Fracture7
  • Ejection Problem2
  • Mechanical Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions18
  • Foreign Body In Patient1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 30, 2025Fracture; Fracture

    ACCORDING TO THE REPORTER, "THE TIP OF THE 1 PRODUCT'S INSERTER(LOT: 23120524) WAS FRACTURED, AND THE FRACTURED PIECE OF INSERTER REMAINS IN THE PATIENT'S BODY."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21277512 ↗
  • MalfunctionDec 4, 2024Fracture; Fracture

    ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE TIP OF THE COMPLAINT PRODUCT'S INSERTER WAS FRACTURED WHILE THE SURGEON TRIED TO INSERT THE ANCHOR INTO THE PATIENT' BURR HOLE, AND THE FRACTURED PIECE OF INSERTER HAS BEEN REMAINED IN THE PATIENT'S BODY. AND THE ANCHOR COULDN'T BE IMPLANTED. THE ABOVE SAME INCIDENT OCCURRED ON THE 2 SAME PRODUCTS. THEREFORE, THE SURGEON USED A BACK UP ANOTHER PRODUCT FOR THE SURGERY."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20847459 ↗
  • MalfunctionDec 4, 2024Fracture; Fracture

    ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE TIP OF THE COMPLAINT PRODUCT'S INSERTER WAS FRACTURED WHILE THE SURGEON TRIED TO INSERT THE ANCHOR INTO THE PATIENT BURR HOLE, AND THE FRACTURED PIECE OF INSERTER HAS BEEN REMAINED IN THE PATIENT'S BODY. AND THE ANCHOR COULDN'T BE IMPLANTED. THE ABOVE SAME INCIDENT OCCURRED ON THE 2 SAME PRODUCTS. THEREFORE, THE SURGEON USED A BACK UP ANOTHER PRODUCT FOR THE SURGERY."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20846833 ↗
  • MalfunctionNov 26, 2024Fracture; Fracture

    ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE TIP OF THIS COMPLAINT PRODUCT'S INSERTER WAS FRACTURED DURING USE. AND THE SURGERY WAS FINISHED WITHOUT REMOVING THE FRACTURED PIECE OF INSERTER, SO IT HAS BEEN REMAINED IN THE PATIENT'S BODY."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20789711 ↗
  • MalfunctionNov 26, 2024Fracture; Fracture

    ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE TIP OF THE COMPLAINT PRODUCT'S INSERTER WAS FRACTURED DURING USE, AND THE ANCHOR COULDN'T BE IMPLANTED. SO, THE SURGEON MADE ANOTHER BURR HOLE AND IMPLANTED A BACK UP OF THE SAME PRODUCT'S ANCHOR INTO IT. THE SURGERY WAS FINISHED WITHOUT REMOVING THE FRACTURED PIECE OF INSERTER, SO IT HAS BEEN REMAINED IN THE PATIENT'S BODY."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20788820 ↗
  • MalfunctionNov 11, 2024Fracture; Fracture

    ACCORDING TO THE REPORTER, "THE TIP OF INSERTER WAS FRACTURED DURING USE AND THE SURGEON DECIDED TO FINISHED THE SURGERY WITHOUT REMOVING THE FRACTURED PIECE OF INSERTER FROM THE PATIENT'S BODY, SO IT HAS BE REMAINED IN THE PATIENT'S BODY."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20658849 ↗
  • MalfunctionFeb 23, 2024Fracture; Fracture

    ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE TIP OF THE INSERTER WAS FRACTURED WHILE THE SURGEON TRIED TO IMPLANTED THE ANCHOR. AND THE SURGERY WAS FINISHED WITH THE FRACTURED PIECE OF INSERTER REMAINING IN THE PATIENT'S BODY."

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18766876 ↗
  • MalfunctionApr 25, 2023Mechanical Problem; Mechanical Problem

    ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE ANCHOR OF THIS COMPLAINT PRODUCT WAS PULLED OUT FROM THE PATIENT'S BURR HOLE ALTHOUGH THE SURGEON IMPLANTED IT FOLLOWING THE NORMAL USAGE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16810284 ↗
  • MalfunctionMar 22, 2023Mechanical Problem; Mechanical Problem

    ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE ANCHOR PULLED OUT FROM THE PATIENT'S BURRHOLE ALTHOUGH THE SURGEON IMPLANTED THE ANCHOR OF THIS COMPLAINT PRODUCT FOLLOWING THE NORMAL USAGE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16594811 ↗
  • MalfunctionNov 30, 2022Ejection Problem; Ejection Problem

    ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE ANCHOR WAS PULLED OUT FROM THE PATIENT'S BURR HOLE WHEN THE SURGEON APPLIED TENSION TO THE SUTURE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15888097 ↗
  • MalfunctionNov 30, 2022Ejection Problem; Ejection Problem

    ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE ANCHOR PULLED OUT FROM THE PATIENT'S BURR HOLE WHEN THE SURGEON APPLIED TENSION TO THE SUTURE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15888047 ↗
  • MalfunctionJun 16, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ACCORDING TO THE REPORTER, "JUGGERKNOT WITH BROADBAND TAPE FAILED, PULLED OUT OF BONE WHILE TIEING KNOTS FOR SLAP REPAIR."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14718561 ↗
  • MalfunctionJun 16, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ACCORDING TO THE REPORTER, "JUGGERKNOT WITH BROADBAND TAPE FAILED, PULLED OUT OF BONE WHILE "TIEING" KNOTS FOR SLAP REPAIR."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14718554 ↗
  • MalfunctionJun 16, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ACCORDING TO THE REPORTER, "JUGGERKNOT WITH BROADBAND TAPE FAILED, PULLED OUT OF BONE WHILE TIEING KNOTS FOR SLAP REPAIR."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14718518 ↗
  • MalfunctionJun 16, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ACCORDING TO THE REPORTER, "JUGGERKNOT WITH BROADBAND TAPE FAILED, PULLED OUT OF BONE WHILE TIEING KNOTS FOR SLAP REPAIR."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14718510 ↗
  • MalfunctionJun 16, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ACCORDING TO THE REPORTER, "JUGGERKNOT WITH BROADBAND TAPE FAILED , PULLED OUT OF BONE WHILE TIEING KNOTS FOR SLAP REPAIR."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14718473 ↗
  • MalfunctionJun 16, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ACCORDING TO THE REPORTER, "JUGGERKNOT WITH BROADBAND TAPE FAILED, PULLED OUT OF BONE WHILE TIEING KNOTS FOR SLAP REPAIR."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14718470 ↗
  • MalfunctionJun 16, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ACCORDING TO THE REPORTER, "JUGGERKNOT WITH BROADBAND TAPE FAILED, PULLED OUT OF BONE WHILE TIEING KNOTS FOR SLAP REPAIR."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14718412 ↗

Other Fastener, Fixation, Nondegradable, Soft Tissue devices

All Fastener, Fixation, Nondegradable, Soft Tissue devices →