Skip to content

JuggerKnot CM-99529BRecalls & reported problems

Riverpoint Medical, LLC · Soft Anchor · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00810020087338), received between Dec 21, 2022 and Mar 21, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem3
  • Mechanical Jam2
  • Device Dislodged or Dislocated1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 21, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE ANCHOR OF THIS COMPLAINT PRODUCT DIDN'T BE IMPLANTED INTO THE PATIENT'S BURR HOLE. THEREFORE, THE SURGEON USED AND IMPLANTED A BACK UP OF THE SAME PRODUCT INTO THE SAME BURR HOLE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18949238 ↗
  • MalfunctionAug 10, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE ANCHOR WAS PULLED OUT FROM THE PATIENT'S BURR HOLE. THEREFORE, THE SURGEON MADE ANOTHER BURR HOLE AND IMPLANTED A BACK UP OF THE SAME PRODUCT'S ANCHOR INTO IT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17502624 ↗
  • MalfunctionJun 19, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE ANCHOR WAS PULLED OUT FROM THE PATIENT'S BURR HOLE UNINTENTIONALLY.".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17159576 ↗
  • MalfunctionMay 19, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE ANCHOR WAS PULLED OUT UNINTENTIONALLY ALTHOUGH THE SURGEON IMPLANTED IT FOLLOWING THE NORMAL USAGE. THEREFORE, THE SURGEON USED A BACK UP ANOTHER PRODUCT FOR THE SURGERY.".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16960683 ↗
  • MalfunctionMar 22, 2023Mechanical Problem; Mechanical Problem

    ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE ANCHOR DIDN'T BECOME TO THE SPHERICAL FORM, SO ANCHOR PULLED OUT FROM THE PATIENT'S BURR HOLE. THEREFORE, THE SURGEON USED A BACK UP PRODUCT FOR THE SURGERY.".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16594763 ↗
  • MalfunctionDec 21, 2022Mechanical Jam; Mechanical Jam

    ACCORDING TO THE REPORTER, "WHEN INSERTING THE JUGGERKNOT BB TRIPLE LOADED, THE ANCHOR JAMMED INSIDE THE SETTING SLEEVE. AFTER REGULAR PREDRILLING AND INSERTION OF THE JUGGERKNOT ANCHOR, AN ATTEMPT WAS MADE TO PULL THE INSERTER OF THE ANCHOR OUT OF THE SETTING SLEEVE. DUE TO THE SMALLER INNER LUMEN OF THE SETTING SLEEVE AND THE MANY THREADS, THIS WAS NOT POSSIBLE. THE USER HAD TO HIT THE SETTING SLEEVE WITH A HAMMER TO RELEASE THE INSERTER AND THE SETTING SLEEVE. THE SOFT ANCHOR BROKE OUT WITH ONE ANCHOR."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16029463 ↗
  • MalfunctionDec 21, 2022Mechanical Jam; Mechanical Jam

    ACCORDING TO THE REPORTER, "WHEN INSERTING THE JUGGERKNOT BB TRIPLE LOADED, THE ANCHOR JAMMED INSIDE THE SETTING SLEEVE. AFTER REGULAR PREDRILLING AND INSERTION OF THE JUGGERKNOT ANCHOR, AN ATTEMPT WAS MADE TO PULL THE INSERTER OUT OF THE SETTING SLEEVE. DUE TO THE SMALL INNER LUMEN OF THE SETTING SLEEVE AND THE MANY THREADS, THIS WAS NOT EASILY POSSIBLE. THE USER HAD TO IT THE SETTING SLEEVE WITH A HAMMER TO RELEASE THE INSERTER AND THE SETTING SLEEVE. THE SOFT ANCHOR BROKE OUT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16029362 ↗

Other Fastener, Fixation, Nondegradable, Soft Tissue devices

All Fastener, Fixation, Nondegradable, Soft Tissue devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page