JuggerKnot CM-99529BRecalls & reported problems
Riverpoint Medical, LLC · Soft Anchor · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00810020087338), received between Dec 21, 2022 and Mar 21, 2024 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Patient Device Interaction Problem3
- Mechanical Jam2
- Device Dislodged or Dislocated1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 21, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE ANCHOR OF THIS COMPLAINT PRODUCT DIDN'T BE IMPLANTED INTO THE PATIENT'S BURR HOLE. THEREFORE, THE SURGEON USED AND IMPLANTED A BACK UP OF THE SAME PRODUCT INTO THE SAME BURR HOLE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18949238 ↗MalfunctionAug 10, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE ANCHOR WAS PULLED OUT FROM THE PATIENT'S BURR HOLE. THEREFORE, THE SURGEON MADE ANOTHER BURR HOLE AND IMPLANTED A BACK UP OF THE SAME PRODUCT'S ANCHOR INTO IT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17502624 ↗MalfunctionJun 19, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE ANCHOR WAS PULLED OUT FROM THE PATIENT'S BURR HOLE UNINTENTIONALLY.".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17159576 ↗MalfunctionMay 19, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE ANCHOR WAS PULLED OUT UNINTENTIONALLY ALTHOUGH THE SURGEON IMPLANTED IT FOLLOWING THE NORMAL USAGE. THEREFORE, THE SURGEON USED A BACK UP ANOTHER PRODUCT FOR THE SURGERY.".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16960683 ↗MalfunctionMar 22, 2023Mechanical Problem; Mechanical Problem
ACCORDING TO THE REPORTER, "IT WAS REPORTED THAT DURING THE SURGERY, THE ANCHOR DIDN'T BECOME TO THE SPHERICAL FORM, SO ANCHOR PULLED OUT FROM THE PATIENT'S BURR HOLE. THEREFORE, THE SURGEON USED A BACK UP PRODUCT FOR THE SURGERY.".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16594763 ↗MalfunctionDec 21, 2022Mechanical Jam; Mechanical Jam
ACCORDING TO THE REPORTER, "WHEN INSERTING THE JUGGERKNOT BB TRIPLE LOADED, THE ANCHOR JAMMED INSIDE THE SETTING SLEEVE. AFTER REGULAR PREDRILLING AND INSERTION OF THE JUGGERKNOT ANCHOR, AN ATTEMPT WAS MADE TO PULL THE INSERTER OF THE ANCHOR OUT OF THE SETTING SLEEVE. DUE TO THE SMALLER INNER LUMEN OF THE SETTING SLEEVE AND THE MANY THREADS, THIS WAS NOT POSSIBLE. THE USER HAD TO HIT THE SETTING SLEEVE WITH A HAMMER TO RELEASE THE INSERTER AND THE SETTING SLEEVE. THE SOFT ANCHOR BROKE OUT WITH ONE ANCHOR."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16029463 ↗MalfunctionDec 21, 2022Mechanical Jam; Mechanical Jam
ACCORDING TO THE REPORTER, "WHEN INSERTING THE JUGGERKNOT BB TRIPLE LOADED, THE ANCHOR JAMMED INSIDE THE SETTING SLEEVE. AFTER REGULAR PREDRILLING AND INSERTION OF THE JUGGERKNOT ANCHOR, AN ATTEMPT WAS MADE TO PULL THE INSERTER OUT OF THE SETTING SLEEVE. DUE TO THE SMALL INNER LUMEN OF THE SETTING SLEEVE AND THE MANY THREADS, THIS WAS NOT EASILY POSSIBLE. THE USER HAD TO IT THE SETTING SLEEVE WITH A HAMMER TO RELEASE THE INSERTER AND THE SETTING SLEEVE. THE SOFT ANCHOR BROKE OUT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16029362 ↗
Other Fastener, Fixation, Nondegradable, Soft Tissue devices
- TRUESPAN ORTHOCORD 228151Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,686 reports
- TRUESPAN ORTHOCORD 228152Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,398 reports
- OMNISPAN ORTHOCORD 228142Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE932 reports
- FOOTPRINT 72202901Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE721 reports
- RIGIDLOOP 232447Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE633 reports
- Arthrex® AR-3636Arthrex, Inc. · Fastener, fixation, nondegradable, soft tissue624 reports
- OMNISPAN ORTHOCORD 228141Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE609 reports
- TRUESPAN ORTHOCORD 228162Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE460 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page