Skip to content

Vision Express BURV53H-0Recalls & reported problems

Mortara Instrument, Inc. · Holter Analysis System · Class 2

All models of Vision Express (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Oct 11, 2024Open, Classified

    There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.

    Z-0323-2025 · Root cause: Software design · Recall details · FDA record ↗

Other Computer, Diagnostic, Programmable devices

All Computer, Diagnostic, Programmable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page