NA CSK-TC5Recalls & reported problems

Boston Scientific Neuromodulation Corporation · CSK TC Electrode, 5cm · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00813250010091), received between Jan 26, 2018 and Aug 12, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Separation” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 29, 2026Open, Classified

    Reusable electrodes may not meet expected performance levels.

    Z-2402-2026 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Material Separation2
  • Break1
  • Structural Problem1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 12, 2024Structural Problem; Structural Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A RADIOFREQUENCY ABLATION (RFA) PROCEDURE DURING WHICH AN ELECTRODE BECAME DETACHED FROM THE CONNECTOR WHILE INSIDE THE PATIENT. THE OUTCOME OF THE PROCEDURE WAS UNABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19966839 ↗
  • MalfunctionApr 30, 2018Break; Material Separation; Break; Material Separation

    THE DEVICE ANALYSIS PERFORMED ON THE ELECTRODE (CSK-T5, LOT M229) REVEALED THAT A PORTION OF THE SHAFT OF THE ELECTRODE WAS BROKEN OFF DUE TO EXCESSIVE EXTERNAL MECHANICAL FORCE. THE DEVICE ANALYSIS PERFORMED ON FIVE ELECTRODES (CSK-TC10, ALL WITH LOT K239) FOUND THAT THE ELECTRODES WERE COMPLETELY SEPARATED FROM THE HUB DUE TO EXCESSIVE EXTERNAL MECHANICAL FORCE. THE DEVICE ANALYSIS PERFORMED ON THE ELECTRODE (CSK-10TC, LOT K509) REVEALED THAT THE ELECTRODE WAS COMPLETELY SEPARATED FROM THE HUB DUE TO EXCESSIVE EXTERNAL MECHANICAL FORCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7474156 ↗
  • MalfunctionJan 26, 2018Material Separation; Material Separation

    A REPORT WAS RECEIVED THAT THE ELECTRODE FELL OUT FROM ITS FIXING HEAD WHILE TAKING OFF THE GUIDE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7222181 ↗