InHealth Technologies; Blom-Singer; Classic IN 1606-NSRecalls & reported problems
Freudenberg Medical, LLC · Class 2
All models of InHealth Technologies; Blom-Singer; Classic (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00813307011071), received between Sep 8, 2025 and Sep 8, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Difficult to Insert” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Insert1
- Migration1
Patient problems
- Airway Obstruction1
- Aspiration/Inhalation1
- Choking1
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 8, 2025Difficult to Insert; Migration; Difficult to Insert; Migration
INCIDENT DESCRIPTION: " A 67-YEAR-OLD PATIENT WITH A TRACHEOSTOMY AND A VOICE PROSTHESIS FOLLOWING A TOTAL LARYNGECTOMY FOR RECURRENT INVASIVE SQUAMOUS CELL CARCINOMA OF THE RIGHT GLOTTIC CAVITY EXPERIENCED A COMPLICATION DURING IMPLANT REPLACEMENT ON (B)(6). THE REPLACEMENT ATTEMPT FAILED, AND DURING THE PROCEDURE, THE VOICE IMPLANT FLANGE WAS ASPIRATED. A SHILEY 8 TRACHEOSTOMY TUBE WITH AN INFLATED CUFF WAS INSERTED SUBSEQUENTLY." PATIENT IMPACT: " THE PATIENT EXPERIENCED CHOKING DURING THE INCIDENT. THERE WAS NO BUZZER IN THE CUBICLE, AND THE SURGEON HAD TO CALL FOR HELP VOCALLY. THE FLANGE WAS NOT VISUALIZED IN THE ESOPHAGUS DURING ENDOSCOPY. AN ATTEMPT TO INSERT A LONGER 8 MM IMPLANT FAILED, AS THE FLANGE COULD NOT BE DEPLOYED. THE PATIENT UNDERWENT REPLACEMENT SURGERY UNDER GENERAL ANESTHESIA WITH A 10 MM BLOM-SINGER IMPLANT ON (B)(6) AND WAS DISCHARGED WITHOUT FURTHER COMPLICATIONS ON (B)(6). NO RETAINED EVIDENCE WAS KEPT BY THE FACILITY."
Patient problems: Airway Obstruction; Aspiration/Inhalation; Choking; Foreign Body In Patient
MAUDE report 22991909 ↗
Other Prosthesis, Laryngeal (Taub) devices
- InHealth Technologies; Blom-Singer; Classic IN 1610-LFFreudenberg Medical, LLC · Prosthesis, Laryngeal (Taub)1 recall0 reports
- InHealth Technologies; Blom-Singer; Classic IN 1610-LEFFreudenberg Medical, LLC · Prosthesis, Laryngeal (Taub)1 recall0 reports
- Provox2 Voice Prosthesis 8mm 7218Atos Medical AB · Prosthesis, Laryngeal (Taub)5 reports
- Provox ActiValve Strong 6 mm 7161Atos Medical AB · Prosthesis, Laryngeal (Taub)3 reports
- Provox Vega 22.5Fr 6mm 8283Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
- Provox Vega 17Fr 6mm 8271Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
- Provox ActiValve Strong 4.5 mm 7160Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
- Provox Vega 22.5Fr 10mm 8285Atos Medical AB · Prosthesis, Laryngeal (Taub)2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page