Impella Heart Pump recallZ-0210-2027 · Abiomed, Inc. · initiated Sep 18, 2026
Class I Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abiomed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.
FDA root cause category: Under Investigation by firm
What the company is doing
Abiomed notified consignees on about 09/18/2026 via "URGENT MEDICAL DEVICE RECALL (REMOVAL)" letter. Consignees were informed of the issue and that affected units can continue to be used while awaiting service from Abiomed to ensure continuity of care. Consignees were instructed to identify if affected units are on hand, Abiomed will contact each consignee to coordinate the removal, update and return of affected devices, complete and return the provided business reply form by email to: [email protected] (or Fax to (866) 201-9339). If you have questions regarding returning this form, please contact (866) 482-2893. Inform all applicable personnel within your facility of the recall, and if any of the affected devices have been forwarded to another facility, to notify those facilities. If you have questions or concerns regarding this notice, please contact [email protected] or your local clinical field staff or call the Manager Product Issue Escalation & Field Actions at 978-646-1400.
Recall details
- Recall number
- Z-0210-2027
- Recall class
- Class I (most serious)
- Status
- Open, Classified
- Initiated by the company
- Sep 18, 2026
- Posted by the FDA
- Oct 28, 2026
- Recalling firm
- Abiomed, Inc.
- Quantity
- 293 units
- Product code
- OZD
- FDA submissions
- P140003, P170011
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries: Belgium, Brazil, Canada, Colombia, Croatia, Czech Republic, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom
Product description
Impella Connect Optical AIC, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
Product Code:0042-0040-US, 0042-0040-UK, 0042-0040-EU, 0042-0040-CA; UDI: 00813502011401, 00813502012484, 00813502012217, 00813502012958; Serial No. IC8024, IC7912, IC8179, IC8216, IC8349, IC8415, IC8418, IC8529, IC8617, IC8808, IC8852, IC8906, IC9081, IC9082, IC9198, IC9227, IC9657, IC9987, IC10097, IC10098, IC10294, IC10327, IC10530, IC10531, IC10383, IC10578, IC10607, IC10549, IC10557, IC10717, IC10722, IC10769, IC10879, IC10877, IC10900, IC10926, IC11012, IC11068, IC11158, IC11168, IC11147, IC11203, IC11353, IC13018, IC13602, IC13606, IC13609, IC13610, IC13605, IC13498, IC13509, IC13506, IC13512, IC13514, IC13515, IC13517, IC13519, IC13513, IC13516, IC13534, IC13536, IC13537, IC13538, IC13535, IC13456, IC13495, IC13485, IC13575, IC13578, IC13551, IC13554, IC13548, IC13552, IC13553, IC13556, IC13558, IC13543, IC13549, IC13555, IC13559, IC13562, IC13564, IC13557, IC13563, IC13649, IC13651, IC13589, IC13591, IC13592, IC13598, IC13599, IC13600, IC13597, IC13593, IC13607, IC13608, IC13571, IC13572, IC13573, IC13596, IC13622, IC13620, IC13621, IC10025, IC9990, IC10240, IC10241, IC10592, IC10601, IC11198, IC11640, IC11642, IC11645, IC11646, IC13696, IC11123, IC11282, IC11264, IC11266, IC10146, IC10180, IC10371, IC10396, IC10397, IC10721, IC10772, IC10828, IC10835, IC11119, IC11072, IC11878, IC11958, IC11967, IC11969, IC11978, IC11986, IC11992, IC11997, IC12260, IC12261, IC12446, IC12454, IC12458, IC12459, IC12544, IC12546, IC12566, IC12586, IC12595, IC12596, IC12666, IC12670, IC12749, IC12744, IC12747, IC12768, IC10164, IC10165, IC12315, IC12318, IC9726, IC9730, IC9729, IC9830, IC10392, IC11217, IC11220, IC11215, IC11219, IC11281, IC11289, IC11952, IC11959, IC11960, IC11966, IC11970, IC11973, IC12309, IC12310, IC12307, IC12319, IC12455, IC12456, IC12467, IC12485, IC12484, IC10394, IC10400, IC10938, IC11715, IC11722, IC12220, IC12221, IC12671, IC12672, IC12561, IC12563, IC13338, IC13143, IC13144, IC13146, IC13148, IC13150, IC13151, IC13156, IC13157, IC13158, IC13142, IC8821, IC9076, IC9427, IC9766, IC10541, IC10639, IC12696, IC12700, IC12702, IC12703, IC12704, IC12753, IC12754, IC12755, IC12757, IC13281, IC10224, IC10225, IC10788, IC10789, IC10790, IC10792, IC10793, IC11723, IC11724, IC12256, IC13235, IC8836, IC8839, IC8902, IC9014, IC9022, IC10140, IC10141, IC10408, IC10425, IC10949, IC10954, IC10947, IC12320, IC12558, IC12568, IC12590, IC12597, IC12541, IC10560, IC10886, IC10892, IC10920, IC11074, IC11229, IC11998, IC12218, IC12288, IC12443, IC12444, IC12445, IC12452, IC12545, IC12552, IC12587, IC12762, IC10168, IC10125, IC11999, IC12278, IC12480, IC12543, IC12766, IC10312, IC11171, IC11172, IC12604, IC12605, IC12891, IC12895, IC12892, IC13266, IC13267, IC13503, IC13504, IC9399, IC9402, IC9626, IC10403.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Impella 0042-0040-USAbiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump9 recalls41 deaths1,306 reports
Product lines
Other devices under the same FDA submission
14 devices cleared under P140003, P170011 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Impella 0048-0045Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump5 recalls1,268 deaths11,110 reports
- Impella 1000100Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump5 recalls840 deaths6,088 reports
- Impella 1000080Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump7 recalls967 deaths5,076 reports
- Impella 0042-0000-USAbiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump13 recalls120 deaths2,466 reports
- Impella 1000323Abiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump6 recalls184 deaths925 reports
- Impella 0550-0008Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump4 recalls87 deaths1,005 reports
- Impella 004334Abiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump3 recalls61 deaths408 reports
- Impella 0046-0035Abiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump4 recalls37 deaths258 reports
- Impella 0048-0024Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump2 recalls39 deaths298 reports
- Impella 0048-0032Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump3 recalls3 deaths42 reports
- Impella 0048-0003Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump6 deaths124 reports
- Impella 1000200Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump9 deaths90 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99778).
- Z-0208-2027 · Open, ClassifiedAutomated Impella Controller, Packaged. Primary user control interface for the Impella Catheter, catheter-based heart pump.
- Z-0209-2027 · Open, ClassifiedImpella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page