M3 Gold M3-30ARecalls & reported problems
Gyrus ACMI, LLC · M3 Gold Autoclavable Foroblique Telescope 30, Color Coded White · Class 2
All models of M3 Gold (5)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00821925005679), received between Nov 11, 2020 and Aug 26, 2026 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Sharp Edges” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
6 other recalls cover devices cleared under the same FDA submission (K890328); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 24, 2026Class IIOpen, Classified
There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.
Z-3299-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Sharp Edges5
- Break2
- Crack1
- Degraded1
- Environmental Particulates1
- Failure to Clean Adequately1
- Material Integrity Problem1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- No Patient Involvement1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 26, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50456819 ↗MalfunctionAug 18, 2026Sharp Edges
The customer reported sharp edge on the lens part during an inspection for use involving an Olympus Hysteroscope (And Accessories). There was no patient injury reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50301613 ↗MalfunctionApr 6, 2026Crack
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FOROBLIQUE TELESCOPE HAD CRACKED OBJECTIVE COVER GLASS AND BROKEN COVER GLASS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24792894 ↗MalfunctionMar 31, 2026Break; Environmental Particulates
IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE HYSTEROSCOPE AND ACCESSORIES EXHIBIT CHIPPED OBJECTIVE LENS AND BROKEN COVER GLASS ALSO DUST PARTICLES WERE FOUND UNDER THE EYEPIECE WINDOW. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24747503 ↗MalfunctionMar 6, 2026Sharp Edges
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HYSTEROSCOPE (AND ACCESSORIES) EXHIBITED DISTAL END DENTS, SHARP. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24533481 ↗MalfunctionFeb 10, 2026Degraded
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24306495 ↗MalfunctionJan 28, 2026Failure to Clean Adequately
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24202230 ↗MalfunctionNov 19, 2025Sharp Edges; Sharp Edges
IT WAS REPORTED THAT THE DISTAL TIP OF THE RIGID SCOPE WAS SHARP. THE ISSUE WAS FOUND DURING RECEIPT OF THE DEVICE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23596119 ↗MalfunctionJul 26, 2024Material Separation; Sharp Edges; Material Separation; Sharp Edges
IT WAS REPORTED THE HYSTEROSCOPE HAD BURNT TIP AND SHARP EDGE CAUSING SEPARATION ON THE LENS. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19838120 ↗MalfunctionJun 11, 2024Material Integrity Problem; Sharp Edges; Material Integrity Problem; Sharp Edges
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE HYSTEROSCOPE DISTAL TIP WAS FOUND TO BE BURNT AND SHARP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19509881 ↗MalfunctionNov 11, 2020Material Twisted/Bent; Material Twisted/Bent
OLYMPUS WAS INFORMED THAT DURING PREPARATION FOR USE, THE AUTOCLAVABLE FOROBLIQUE 30 DEGREE TELESCOPE'S OUTER TUBE WAS FOUND BENT. THERE WAS NO PATIENT INVOLVEMENT, AND NO USER INJURY REPORTED.
Patient problems: No Patient Involvement
MAUDE report 10825715 ↗
FDA clearance and classification
Cleared through 510(k) K890328 on Feb 6, 1990 (decision: Substantially Equivalent). Submitted by Circon Video as “Adult Resectoscope”. The FDA review took 379 days. FDA record ↗
The FDA regulates product code FAS (electrode, electrosurgical, active, urological) under 21 CFR 876.4300 as a Class II device. 21 CFR 876.4300 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K890328. One submission often covers many models, so check the product and lot details in each record.
- Z-3300-2026 · Aug 24, 2026 · Open, ClassifiedOlympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.
- Z-2720-2026 · May 5, 2026 · Open, ClassifiedOlympus Ellik Evacuator Includes Glass Body, Male Connecting Tubing & Latex Bulb. Model Number: 194.
- Z-2139-2025 · Jun 18, 2025 · Open, ClassifiedBlue Silicone, Reusable Silicone Seal 6 - 12 FR. Model Number: CS-B612. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the working channel. Some
- Z-2140-2025 · Jun 18, 2025 · Open, ClassifiedGreen Silicone Seal for Accessories, Reusable Silicone Seal 0 - 7 FR. Model Number: CS-G7. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the work
- Z-0568-2024 · Oct 2, 2023 · Open, ClassifiedACMI Roller Bar Electrodes bearing incorrect pouch label of ACMI USA Elite System and USA Series Right Angle Cutting Loop Electrode Stabilized, 0.012 Wire 24 Fr (8mm), Model: REF M
- Z-1952-2018 · Apr 17, 2018 · TerminatedGyrus ACMI Reusable Active Cord for Bovie/ValleyLab Generators, Catalog No C650-129A (00821925008700) The cord consists of a single conductor stranded wire terminated with plugs t
Common questions about the M3 Gold M3-30A
Has the M3 Gold M3-30A been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-3299-2026 (Class II), began on Aug 24, 2026: There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the M3 Gold M3-30A?
The FDA's MAUDE database lists 11 adverse event reports for this device, received between Nov 11, 2020 and Aug 26, 2026. In those reports, the most common specific device problems are “Sharp Edges” (5), “Break” (2) and “Crack” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the M3 Gold M3-30A?
None of the reports are classified as deaths, 0 as injuries and 11 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the M3 Gold M3-30A FDA approved?
It was cleared through 510(k) K890328 on Feb 6, 1990. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the M3 Gold M3-30A, and is it still sold?
The FDA's device database lists Gyrus ACMI, LLC as the company responsible for it (brand M3 Gold, model M3-30A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL devices
- USA Elite MLE-24-012Gyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall9 reports
- Olympus WA22603SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall7 reports
- Olympus WA22621SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall4 reports
- Olympus WA22305SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall3 reports
- Olympus WA22657SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
- Olympus WA22332DOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
- Olympus WA22301DOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
- Olympus WA22305DOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page