Olympus WA22301DRecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 14042761051658), received between Jun 25, 2018 and Feb 13, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Jun 8, 2023Open, Classified
Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment, implants or stents. As a result, fragments can fall inside the patient and must be retrieved. (2) The second Caution refers to the risk of sparkover when getting in contact with metal parts as this can cause uncontrolled heating of the HF-resection electrode and may result in damage and breakage to the HF-resection electrode tip
Z-2413-2023 · Root cause: Process design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break2
- Fracture2
- Material Fragmentation2
- Mechanical Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 13, 2023Break; Fracture; Material Fragmentation; Mechanical Problem; Break; Fracture; Material Fragmentation; Mechanical Problem
OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC TRANSCERVICAL RESECTION (TCR) PROCEDURE, THE LOOP WIRE AT THE DISTAL END OF THE HF RESECTION ELECTRODE BROKE OFF AND FELL INTO THE PATIENT. HOWEVER, NO FRAGMENT REMAINED INSIDE THE PATIENT SINCE A CT SCAN WAS PERFORMED WHERE NO FOREIGN OBJECTS WERE FOUND INSIDE THE PATIENT. THE INTENDED PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A SIMILAR DEVICE AND THERE WAS NO REPORT ABOUT PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16361336 ↗MalfunctionJun 25, 2018Break; Fracture; Material Fragmentation; Mechanical Problem; Break; Fracture; Material Fragmentation; Mechanical Problem
OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC TRANSCERVICAL RESECTION OF THE ENDOMETRIUM IN SALINE (TCRIS) PROCEDURE, THE LOOP WIRE AT THE DISTAL END OF THE HF RESECTION ELECTRODE DETACHED AND FELL INSIDE THE PATIENT. THE LOOP WIRE WAS REPORTEDLY NOT RETRIEVED BUT AN EXAMINATION BY X-RAY SHOWED NO FOREIGN OBJECTS INSIDE THE PATIENT. THE INTENDED PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER HF RESECTION ELECTRODE AND THERE WAS NO REPORT ABOUT AN ADVERSE EVENT OR PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7633196 ↗
Other Electrode, Electrosurgical, Active, Urological devices
- M3 Gold M3-30AGyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall11 reports
- USA Elite MLE-24-012Gyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall9 reports
- Olympus WA22603SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall7 reports
- Olympus WA22621SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall4 reports
- Olympus WA22305SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall3 reports
- Olympus WA22657SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
- Olympus WA22332DOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
- Olympus WA22305DOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page