Skip to content

USA Elite MLE-24-012Recalls & reported problems

Gyrus ACMI, LLC · ACMI USA Elite System and USA Series Right Angle Cutting Loop Electrode Stabilized, 0.012 Wire 24 Fr (8mm) (12/PK) · Class 2

All models of USA Elite (24)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00821925036413), received between Apr 2, 2020 and Dec 14, 2023 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Arcing of Electrodes” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

6 other recalls cover devices cleared under the same FDA submission (K890328); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2023Class IIOpen, Classified

    Incorrect pouch labels, the contents within the mislabeled pouch were Roller Bar electrodes.

    Z-0568-2024 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Arcing of Electrodes4
  • Defective Device3
  • Connection Problem2
  • Component Missing1
  • Fracture1
  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 14, 2023Fracture; Fracture

    THE CUSTOMER REPORTED TO OLYMPUS THE CUTTING LOOP WIRE LOOP WAS BREAKING THROUGH THE INSULATION BEFORE BREAKING ALTOGETHER. THE ISSUE WAS FOUND DURING AN UNIDENTIFIED PROCEDURE WHICH WAS COMPLETED WITH A SIMILAR DEVICE. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, MISHANDLING OF THE DEVICE LEADING TO THE BREAK WAS FOUND WHICH CONSTITUTES A REPORTABLE MALFUNCTION. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18326290 ↗
  • MalfunctionDec 21, 2022Connection Problem; Connection Problem

    THE CUSTOMER REPORTS DURING AN UNSPECIFIED PROCEDURE USING A CUTTING LOOP, THE CUTTING LOOP WOULD NOT MEET CONNECTION TO CORD IN WORKING ELEMENT. THE USER HAD TO OPEN ANOTHER. THERE WAS NO RISK TO STAFF OR PATIENT. THE PROCEDURE WAS PROLONGED LONG ENOUGH TO GET NEW EQUIPMENT WITH NO FURTHER CONSEQUENCES TO THE PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16028016 ↗
  • MalfunctionDec 21, 2022Connection Problem; Connection Problem

    THE CUSTOMER REPORTS DURING AN UNSPECIFIED PROCEDURE USING A CUTTING LOOP, THE CUTTING LOOP WOULD NOT MEET CONNECTION TO CORD IN WORKING ELEMENT. THE USER HAD TO OPEN ANOTHER. THERE WAS NO RISK TO STAFF OR PATIENT. THE PROCEDURE WAS PROLONGED LONG ENOUGH TO GET NEW EQUIPMENT WITH NO FURTHER CONSEQUENCES TO THE PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16027653 ↗
  • MalfunctionJul 22, 2022Sparking; Sparking

    AS REPORTED, DURING A HYSTEROSCOPIC RESECTION THERAPEUTIC PROCEDURE, UPON ACTIVATION IN THE PATIENT, COULD HEAR SPARKING FROM INSIDE THE PATIENT AND COULD SMELL BURNING OF PLASTIC. THE ELECTRODE WAS IMMEDIATELY REMOVED AND REPLACED. THE CASE WAS SUCCESSFULLY COMPLETED WITH A NEW ELECTRODE. THERE WAS NO PATIENT HARM, NO USER INJURY REPORTED DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15089876 ↗
  • MalfunctionApr 22, 2021Arcing of Electrodes; Defective Device; Arcing of Electrodes; Defective Device

    QUANTITY OF 3 ELECTRODES WITH THE SAME MODEL AND LOT NUMBER WERE USED DURING A TRANSURETHRAL RESECTION OF THE PROSTATE (TURP) PROCEDURE. THE FIRST TWO (2 FAILED) AND THE THIRD (3RD) ONE WAS USED TO COMPLETE THE INTENDED PROCEDURE. THE ELECTRODES WERE REPORTED TO BE DEFECTIVE (ARCED). THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED. THIS COMPLAINT IS RELATED TO REPORTS PATIENT IDENTIFIER (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11711753 ↗
  • MalfunctionApr 22, 2021Arcing of Electrodes; Defective Device; Arcing of Electrodes; Defective Device

    QUANTITY OF 3 ELECTRODES WITH THE SAME MODEL AND LOT NUMBER WERE USED DURING A TRANSURETHRAL RESECTION OF THE PROSTATE (TURP) PROCEDURE. THE FIRST TWO (2 FAILED) AND THE THIRD (3RD) ONE WAS USED TO COMPLETE THE INTENDED PROCEDURE. THE ELECTRODES WERE REPORTED TO BE DEFECTIVE (ARCED). THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED. THIS COMPLAINT IS RELATED TO REPORT PATIENT IDENTIFIER (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11711748 ↗
  • MalfunctionApr 22, 2021Arcing of Electrodes; Defective Device; Arcing of Electrodes; Defective Device

    QUANTITY OF 3 ELECTRODES WITH THE SAME MODEL AND LOT NUMBER WERE USED DURING A TRANSURETHRAL RESECTION OF THE PROSTATE (TURP) PROCEDURE. THE FIRST TWO (2 FAILED) AND THE THIRD (3RD) ONE WAS USED TO COMPLETE THE INTENDED PROCEDURE. THE ELECTRODES WERE REPORTED TO BE DEFECTIVE. THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED. THIS COMPLAINT IS RELATED TO REPORTS PATIENT IDENTIFIER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11711689 ↗
  • MalfunctionFeb 26, 2021Arcing of Electrodes; Arcing of Electrodes

    IT WAS REPORTED THAT DURING A TRANSURETHRAL RESECTION OF THE PROSTATE (TURP) PROCEDURE, THE 4 ELECTRODES ARCED WHILE USING IN THE PATIENT. THERE WERE NO FURTHER DETAILS PROVIDED REGARDING THE EVENT. THERE WAS NO PATIENT INFECTION, INJURY OR HARM REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11385507 ↗
  • MalfunctionApr 2, 2020Component Missing

    DURING A HYSTEROSCOPY/MYOMECTOMY, A RIGHT ANGLE CUTTING LOOP ELECTRODE WAS USED. THE CUTTING LOOP DISAPPEARED AS IF IT DISINTEGRATED OR BURNED OFF. AT THE TIME IT WAS BEING USED AT 220 CUT, 100 COAG. BOTH THE UTERUS AND SPECIMEN WERE SEARCHED FOR ANY PARTS. PHYSICIAN WAS CONFIDENT THAT NOTHING WAS LEFT BEHIND. DEVICE WAS GIVEN TO MATERIAL'S MANAGEMENT FOR RETURN TO THE MANUFACTURER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9915849 ↗

FDA clearance and classification

Cleared through 510(k) K890328 on Feb 6, 1990 (decision: Substantially Equivalent). Submitted by Circon Video as “Adult Resectoscope”. The FDA review took 379 days. FDA record ↗

The FDA regulates product code FAS (electrode, electrosurgical, active, urological) under 21 CFR 876.4300 as a Class II device. 21 CFR 876.4300 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K890328. One submission often covers many models, so check the product and lot details in each record.

  • Z-3300-2026 · Aug 24, 2026 · Open, Classified
    Olympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.
  • Z-3299-2026 · Aug 24, 2026 · Open, Classified
    Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.
  • Z-2720-2026 · May 5, 2026 · Open, Classified
    Olympus Ellik Evacuator Includes Glass Body, Male Connecting Tubing & Latex Bulb. Model Number: 194.
  • Z-2139-2025 · Jun 18, 2025 · Open, Classified
    Blue Silicone, Reusable Silicone Seal 6 - 12 FR. Model Number: CS-B612. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the working channel. Some
  • Z-2140-2025 · Jun 18, 2025 · Open, Classified
    Green Silicone Seal for Accessories, Reusable Silicone Seal 0 - 7 FR. Model Number: CS-G7. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the work
  • Z-1952-2018 · Apr 17, 2018 · Terminated
    Gyrus ACMI Reusable Active Cord for Bovie/ValleyLab Generators, Catalog No C650-129A (00821925008700) The cord consists of a single conductor stranded wire terminated with plugs t

Common questions about the USA Elite MLE-24-012

Has the USA Elite MLE-24-012 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0568-2024 (Class II), began on Oct 2, 2023: Incorrect pouch labels, the contents within the mislabeled pouch were Roller Bar electrodes. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the USA Elite MLE-24-012?

The FDA's MAUDE database lists 9 adverse event reports for this device, received between Apr 2, 2020 and Dec 14, 2023. In those reports, the most common specific device problems are “Arcing of Electrodes” (4), “Defective Device” (3) and “Connection Problem” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the USA Elite MLE-24-012?

None of the reports are classified as deaths, 0 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the USA Elite MLE-24-012 FDA approved?

It was cleared through 510(k) K890328 on Feb 6, 1990. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the USA Elite MLE-24-012, and is it still sold?

The FDA's device database lists Gyrus ACMI, LLC as the company responsible for it (brand USA Elite, model MLE-24-012). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL devices

All ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page