Olympus WA22332DRecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 14042761051702), received between Aug 29, 2018 and Jul 25, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Jun 8, 2023Open, Classified
Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment, implants or stents. As a result, fragments can fall inside the patient and must be retrieved. (2) The second Caution refers to the risk of sparkover when getting in contact with metal parts as this can cause uncontrolled heating of the HF-resection electrode and may result in damage and breakage to the HF-resection electrode tip
Z-2413-2023 · Root cause: Process design · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Fracture1
- Material Fragmentation1
- Mechanical Problem1
Patient problems
- Device Embedded In Tissue or Plaque1
- Perforation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 25, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS REVIEWED THE FOLLOWING LITERATURE TITLED: "EN BLOC RESECTION VERSUS CONVENTIONAL TURBT FOR T1HG BLADDER CANCER: A PROPENSITY SCORE-MATCHED ANALYSIS." THIS STUDY AIMED TO CLARIFY THE PERIOPERATIVE, ONCOLOGIC, AND PATHOLOGIC OUTCOMES OF EN BLOC RESECTION OF BLADDER TUMORS (ERBT) FOR PATIENTS WITH PT1 BLADDER CANCER (BCA) COMPARED WITH CONVENTIONAL TRANSURETHRAL RESECTION OF BLADDER TUMORS (CTURBT) IN A LARGE MULTICENTER RETROSPECTIVE COHORT USING PROPENSITY SCORE-MATCHED (PSM) ANALYSIS. THE PSM COHORT CONSISTED OF 202 PATIENTS: 101 IN CTURBT GROUP AND 101 IN ERBT GROUP. COMPLICATIONS WERE DETECTED IN SIX PATIENTS IN THE CTURBT GROUP AND FOUR PATIENTS IN THE ERBT GROUP. THE 3-YEAR RFS ESTIMATES WERE 59.3% [95% CONFIDENCE INTERVAL (CI) 48.2¿68.8] IN THE CTURBT GROUP, AND 65.8% (95% CI 54.0¿75.2) IN THE ERBT GROUP (P = 0.7). THERE WERE NO DIFFERENCES IN THE 3-YEAR MFS AND THE 3-YEAR CSS ESTIMATES BETWEEN THE TWO GROUPS. THE DETECTION RATE FOR MM¿VP WAS SIGNIFICANTLY HIGHER IN PATIENTS WHO UNDERWENT ERBT (CTURBT: 77% VERSUS ERBT: 99%, P\0.001). ERBT IMPROVES THE QUALITY OF RESECTION AND SPECIMEN, YIELDING LESS RESIDUE ON RETURN AND YIELDING SUPERIOR HISTOPATHOLOGIC INFORMATION SU
Patient problems: Perforation
MAUDE report 17393586 ↗MalfunctionAug 29, 2018Break; Fracture; Material Fragmentation; Mechanical Problem; Break; Fracture; Material Fragmentation; Mechanical Problem
OLYMPUS WAS INFORMED THAT DURING AN UNSPECIFIED HYSTEROSCOPY PROCEDURE, A PART OF THE LOOP WIRE AT THE DISTAL END OF THE HF RESECTION ELECTRODE BROKE OFF AND A FRAGMENT FELL INSIDE THE PATIENT'S BODY AND REMAINED THERE, SINCE IT COULD NOT BE LOCATED. NO FURTHER INFORMATION WAS PROVIDED AND THERE WAS ALSO NO REPORT ABOUT AN ADVERSE EVENT OR PATIENT INJURY.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 7825979 ↗
Other Electrode, Electrosurgical, Active, Urological devices
- M3 Gold M3-30AGyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall11 reports
- USA Elite MLE-24-012Gyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall9 reports
- Olympus WA22603SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall7 reports
- Olympus WA22621SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall4 reports
- Olympus WA22305SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall3 reports
- Olympus WA22301DOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
- Olympus WA22657SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
- Olympus WA22305DOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page