UroPass 61046BXRecalls & reported problems
Gyrus ACMI, LLC · UroPass Ureteral Access Sheath 10/12 Fr (3.3/4 mm) x 46 cm (5/PK) · Class 2
All models of UroPass (15)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00821925035331), received between Jan 25, 2021 and Aug 3, 2021 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K051593); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Sep 11, 2024Class IIOpen, Classified
Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures.
Z-0364-2025 · Root cause: Storage · Recall details · FDA record ↗
Most reported problems
Device problems
- Break2
- Detachment of Device or Device Component1
- Device Contamination with Chemical or Other Material1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 3, 2021Break; Break
UPDATES ON THE EVENT REPORTED: FURTHER COMMUNICATION WITH THE CUSTOMER CONVEYED THE FOLLOWING INFORMATION: CUSTOMER STATED THAT THE SUBJECT DEVICE IS NOT RETURNING FOR EVALUATION. THE SUBJECT DEVICE WAS DISPOSED. THE INTENDED PROCEDURE WAS COMPLETED (DID NOT PROVIDE THE TYPE/SPECIFIC PROCEDURE). CUSTOMER STATED NO PATIENT HARM, NO USER INJURY. NO FURTHER DETAILS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12268615 ↗MalfunctionJun 30, 2021Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT USER FOUND A FOREIGN OBJECT INSIDE THE STERILE PACKAGE BEFORE ITS USE. NO PATIENT HARM OR INJURY OCCURRED DUE TO THIS MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12094870 ↗MalfunctionMay 5, 2021Break; Break
A REPORT OF DILATOR TIP WAS BROKEN AND THE BROKEN PIECES FELL INTO THE PATIENT DURING A RETROGRADE INTRARENAL SURGERY PROCEDURE. THE USER INSERTED THE DEVICE AND FELT SOME RESISTANCE FROM THE PRODUCT WHICH WAS DIFFERENT FROM THE NORMAL USE. THE END DILATOR WAS CHECKED AND FOUND BROKEN PIECES FELL INTO THE PATIENT. EACH BROKEN PIECE WAS FOUND AND REMOVED FROM THE PATIENT BLADDER AND LOWER URETER. THE BROKEN PIECES WERE RETRIEVED USING FORCEPS WITH RIGID CYSTOSCOPE AND RIGID URETEROSCOPE. THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11776471 ↗InjuryJan 25, 2021Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component
AS REPORTED, DURING THE REMOVAL OF A STONE BY THE UROPASS DEVICE, THE MATERIAL WAS REMOVED FROM THE UROPASS AND FELL INTO THE PATIENT. ALL THE PIECES OF THE STONE WERE SUCCESSFULLY REMOVED FROM THE PATIENT. THERE WAS NO PATIENT HARM REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11224348 ↗
FDA clearance and classification
Cleared through 510(k) K051593 on Sep 2, 2005 (decision: Substantially Equivalent). Submitted by Acmi Corporation as “Uropass Ureteral Access Sheath, Model 61224”. The FDA review took 78 days. FDA record ↗
The FDA regulates product code KNY (accessories, catheter, g-u) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K051593. One submission often covers many models, so check the product and lot details in each record.
- Z-2401-2023 · May 19, 2023 · Open, ClassifiedUroPass Ureteral Access Sheaths, 5 pieces/box
Common questions about the UroPass 61046BX
Has the UroPass 61046BX been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0364-2025 (Class II), began on Sep 11, 2024: Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the UroPass 61046BX?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jan 25, 2021 and Aug 3, 2021. In those reports, the most common specific device problems are “Break” (2), “Detachment of Device or Device Component” (1) and “Device Contamination with Chemical or Other Material” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the UroPass 61046BX?
None of the reports are classified as deaths, 1 as an injury and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the UroPass 61046BX FDA approved?
It was cleared through 510(k) K051593 on Sep 2, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the UroPass 61046BX, and is it still sold?
The FDA's device database lists Gyrus ACMI, LLC as the company responsible for it (brand UroPass, model 61046BX). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ACCESSORIES, CATHETER, G-U devices
- UroPass 61238BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall10 reports
- UroPass 61138BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall5 reports
- UroPass 61024BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall5 reports
- UroPass 61224BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61146BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61254BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61054BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61324BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page