Brand Name: UroPass Ureteral Access Sheaths recallZ-0364-2025 · Olympus Corporation of the Americas · initiated Sep 11, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures.
FDA root cause category: Storage
What the company is doing
****UPDATE Nov 8, 2024*** Olympus mailed a revision letter-URGENT: MEDICAL DEVICE CORRECTION UPDATED INFORMATION to consignees on November 8, 2024. Letter states; To reduce the risk of UV exposure to your device(s), Olympus instructs users to implement the following actions: Minimize the exposure of UroPass Ureteral Access Sheaths to ultraviolet light. Exposure to UV light during storage periods may cause embrittlement of this device, increasing the potential for breakage and patient injury, including retained device fragments, tissue damage, or perforation of the ureter. Updated Information: " As instructed in IFU Section 4.0, inspect the device prior to use. If any damage to the device is noted, DO NOT USE. " Inspect the device upon removal or completion of the procedure. If damage is noted, the patient should be evaluated for a potentially retained foreign body. " Sharp edges of a broken sheath have been reported to have caused ureteral tissue damage or perforation. If device damage is noted, patients should be monitored intraoperatively and in the immediate post-operative period for these types of injuries. Olympus is not requiring the return of your UroPass device(s) as a result of this action. However, if you want to return the UroPass device(s) in your inventory, please contact Customer Service at 1-800-848-9024, option 2 to obtain a Return Material Authorization. Olympus will issue a credit to your facility upon return of your affected product. 4. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification, including the new information regarding UV exposure. Olympus is in the process of updating the Instructions for Use with this information. 5. Olympus requests that you acknowledge receipt of this letter. Access the Olympus recall portal to indicate that you have received this notification. a. Go to https://olympusamerica.com/recall b. Enter the recall number "0458 c. Complete the form as instructed. 6. If
Recall details
- Recall number
- Z-0364-2025
- Status
- Open, Classified
- Initiated by the company
- Sep 11, 2024
- Posted by the FDA
- Nov 12, 2024
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 14,093 units
- Product code
- KNY
- FDA submissions
- K051593
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Canada, China, Germany, Hong Kong, Japan, Republic of South Korea, Singapore.
Product description
Brand Name: UroPass Ureteral Access Sheaths Product Name: Ureteral Access Sheaths Model/Catalog Number: 61024BX, 61038BX, 61046BX, 61054BX, 61124BX, 61138BX, 61146BX, 61154BX, 61224BX, 61238BX, 61246BX, 61254BX, 61324BX, 61338BX, 61346BX, 61354BX Software Version: N/A Product Description: The Olympus UroPass Ureteral Access Sheath Set consists of a hydrophilic coated outer sheath and an inner tapered dilator intended to establish a conduit for the passage of endoscopes and retrieval devices into the ureter. The hydrophilic coating on the UroPass Ureteral Access Sheath eases passage and placement. Both the outer sheath and inner dilator are radio-opaque for ease of viewing radiographically. Intended for single use only. Component: N/A
Code information (models, lots, serial numbers)
Model/UDI-DI: 61024BX/00821925035317, 61038BX/00821925035324, 61046BX/00821925035331, 61054BX/00821925035348, 61124BX/00821925035355, 61138BX/00821925035362, 61146BX/00821925035379, 61154BX/00821925035386, 61224BX/00821925035393, 61238BX/00821925035409, 61246BX/00821925035416, 61254BX/00821925035423, 61324BX/00821925035430, 61338BX/00821925035447, 61346BX/00821925035454, 61354BX/00821925035461.
Devices named in this recall
16 devices in the FDA device registry whose identifier appears in the recall's code information.
- UroPass 61238BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall10 reports
- UroPass 61024BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall5 reports
- UroPass 61138BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall5 reports
- UroPass 61046BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61054BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61146BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61224BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61254BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61324BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall3 reports
- UroPass 61338BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall2 reports
- UroPass 61154BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall1 report
- UroPass 61246BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall1 report
- UroPass 61346BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall1 report
- GYRUS ACMI, INC 61038BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall0 reports
- UroPass 61124BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall0 reports
- UroPass 61354BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page