Skip to content

UroPass 61146BXRecalls & reported problems

Gyrus ACMI, LLC · UroPass Ureteral Access Sheath 11/13 Fr (3.7/4.3 mm) x 46 cm (5/PK) · Class 2

All models of UroPass (15)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00821925035379), received between Jan 7, 2021 and Oct 15, 2024 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Device Fell” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K051593); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 11, 2024Class IIOpen, Classified

    Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures.

    Z-0364-2025 · Root cause: Storage · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Fell1
  • Material Frayed1
  • Material Integrity Problem1
  • Material Separation1
  • Sharp Edges1
  • Structural Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 15, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE TIP OF THE UROPASS ACCESS SHEATH WAS DAMAGED WHEN IT WAS TAKEN OUT OF THE BOX. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20457753 ↗
  • MalfunctionSep 21, 2021Structural Problem; Sharp Edges; Structural Problem; Sharp Edges

    AS REPORTED, THE USER FOUND THE SHAPE OF THE DILATOR TIP END ABNORMAL, IT HAS STRAIGHT SECTION HOWEVER, THE SUBJECT DEVICE HAS A LITTLE DIAGONAL CROSS SECTION. AS THE USER FELT RESISTANCE WHEN INSERTING THE UROPASS ON THE PROCEDURE, IT TOOK IT OUT FROM THE PATIENT BODY AND CHECKED THIS ABNORMALITY. THE ISSUE FOUND DURING A THERAPEUTIC RIRS (RETROGRADE INTRARENAL SURGERY) PROCEDURE. THE SUBJECT DEVICE WAS NOT USED ON THE PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED USING THE SAME SET OF EQUIPMENT WITH NO INJURY OR HARM TO THE PATIENT. THERE WAS NO USER INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12509457 ↗
  • MalfunctionSep 16, 2021Material Frayed; Device Fell; Material Frayed; Device Fell

    AS REPORTED, THE USER WERE USING THE DEVICE DURING A STONE MANIPULATION PROCEDURE. WHEN THE DEVICE WAS PUT OVER THE GUIDE WIRE AND THEN PUT INTO THE PATIENT THE USER NOTICED AN ISSUE. THE USER NOTICED THAT THE VERY TIP OF THE PRODUCT, THE WHITE PIECE, HAD FRAYED AND FELL OFF INTO THE PATIENT. THE PIECE THAT FELL OFF WAS RETRIEVED FROM THE PATIENT WITH A FLEXIBLE GRASPER WITH NO REPORTED ADVERSE EVENT TO THE PATIENT. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER SIMILAR DEVICE. THERE IS NO PATIENT HARM OR INJURY ASSOCIATED ON THIS REPORTED EVENT. NO USER INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12484542 ↗
  • InjuryJan 7, 2021Material Separation; Material Separation

    IT IS REPORTED DURING A CYSTOSCOPY USING A UROPASS ACCESS SHEATH, THE DISTAL FIVE INCHES OF THE INNER CANNULA SEPARATED AND WAS RETAINED INSIDE THE PATIENT. EVENT DETAILS AS FOLLOWS: THE DOCTOR STATED THAT THE PROCEDURE STARTED NORMALLY, WITH INTRODUCING THE CYSTOSCOPE TO SURVEY THE BLADDER. HE THEN PLACED A GLIDEWIRE UP THE URETER, THEN PLACED A SECOND WIRE UP. NEXT, THE ACCESS SHEATH WAS INTRODUCED. WHEN THE DOCTOR REMOVED THE INNER CANNULA OF THE ACCESS SHEATH, IT WAS APPARENT THE DISTAL 5 INCHES OF THE INNER CANNULA HAD DETACHED. THE PROCEDURE WAS NOT COMPLETED. A 5-INCH SEGMENT OF THE DISTAL END OF THE INNER CANNULA WAS RETAINED INSIDE THE PATIENT. THE PATIENT REPORTED PAIN DUE TO THE RETAINED DEVICE FRAGMENT. ON AN UNSPECIFIED DATE, THE PATIENT UNDER WENT A SECOND PROCEDURE DURING WHICH THE DEVICE FRAGMENT WAS REMOVED WITH NO FURTHER CONSEQUENCES REPORTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 11131694 ↗

FDA clearance and classification

Cleared through 510(k) K051593 on Sep 2, 2005 (decision: Substantially Equivalent). Submitted by Acmi Corporation as “Uropass Ureteral Access Sheath, Model 61224”. The FDA review took 78 days. FDA record ↗

The FDA regulates product code KNY (accessories, catheter, g-u) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K051593. One submission often covers many models, so check the product and lot details in each record.

  • Z-2401-2023 · May 19, 2023 · Open, Classified
    UroPass Ureteral Access Sheaths, 5 pieces/box

Common questions about the UroPass 61146BX

Has the UroPass 61146BX been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0364-2025 (Class II), began on Sep 11, 2024: Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the UroPass 61146BX?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jan 7, 2021 and Oct 15, 2024. In those reports, the most common specific device problems are “Device Fell” (1), “Material Frayed” (1) and “Material Integrity Problem” (1); the most common patient problem is “Device Embedded In Tissue or Plaque” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the UroPass 61146BX?

None of the reports are classified as deaths, 1 as an injury and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the UroPass 61146BX FDA approved?

It was cleared through 510(k) K051593 on Sep 2, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the UroPass 61146BX, and is it still sold?

The FDA's device database lists Gyrus ACMI, LLC as the company responsible for it (brand UroPass, model 61146BX). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other ACCESSORIES, CATHETER, G-U devices

All ACCESSORIES, CATHETER, G-U devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page