Skip to content

UroPass 61138BXRecalls & reported problems

Gyrus ACMI, LLC · UroPass Ureteral Access Sheath 11/13 Fr (3.7/4.3 mm) x 38 cm (5/PK) · Class 2

All models of UroPass (15)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00821925035362), received between Apr 14, 2021 and Dec 28, 2022 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K051593); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Sep 11, 2024Class IIOpen, Classified

    Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures.

    Z-0364-2025 · Root cause: Storage · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break3
  • Material Separation1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 28, 2022Material Separation; Material Separation

    (B)(6) MODEL# 61154BX / LOT # 09J1800111.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16062697 ↗
  • MalfunctionNov 16, 2022Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED, AT PREPARATION FOR USE, THE USER TRIED TO USE THE PRODUCT AFTER OPENING , HOWEVER, IT WAS STATED THAT IT WAS DIFFICULT TO USE DUE TO THE PRODUCT WAS BENT SIGNIFICANTLY COMPARED TO THE NORMAL PRODUCT. ACCORDING TO THE REPORT, IT WAS STATED THAT THE PRODUCT ITSELF TOO STIFF AND CURVED, MAKING IT DIFFICULT TO USE FOR PATIENTS. THE DEVICE WAS REPLACED . THE INTENDED THERAPEUTIC RIRS (RETROGRADE INTRARENAL SURGERY) PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. NO PATIENT HARM , NO USER INJURY REPORTED DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15807169 ↗
  • MalfunctionApr 15, 2021Break; Break

    DURING A THERAPEUTIC URETEROSCOPY PROCEDURE THE USER ATTEMPTED 3 DIFFERENT ACCESS SHEATHS AND EACH TIME, THE END TIP OF THE INNER INTRODUCER CRUMBLED. FOR THE TWO SHEATHS THE ISSUE WAS FOUND BEFORE IT WAS INSERTED/USED ON THE PATIENT. THE PARTS FELL INSIDE THE PATIENTS¿ KIDNEY WHEN THE THIRD SHEATH WAS USED. THE DEVICE WAS VISUALLY INSPECTED PRIOR TO BEING INSERTED AND THE DEVICE APPEARED TO BE INTACT WHEN IT WAS INSERTED. UNDER THE ENDOSCOPIC IMAGE THE PARTS THAT FELL WERE REMOVED USING A COOK BASKET WITH MODEL NUMBER UNKNOWN. THE INTENDED PROCEDURE WAS COMPLETED USING ANOTHER SHEATHS WITH MODEL NO.61146BX, LOT NO. 09L1900159. THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. THERE WAS NO USER INJURY REPORTED. THIS REPORT IS RELATED TO REPORTS PATIENT IDENTIFIER (B)(6) AND (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11670708 ↗
  • MalfunctionApr 14, 2021Break; Break

    DURING A THERAPEUTIC URETEROSCOPY PROCEDURE THE USER ATTEMPTED 3 DIFFERENT ACCESS SHEATHS AND EACH TIME, THE END TIP OF THE INNER INTRODUCER CRUMBLED. FOR THE TWO SHEATHS THE ISSUE WAS FOUND BEFORE IT WAS INSERTED/USED ON THE PATIENT. THE PARTS FELL INSIDE THE PATIENTS¿ KIDNEY WHEN THE THIRD SHEATH WAS USED. THE DEVICE WAS VISUALLY INSPECTED PRIOR TO BEING INSERTED AND THE DEVICE APPEARED TO BE INTACT WHEN IT WAS INSERTED. UNDER THE ENDOSCOPIC IMAGE THE PARTS THAT FELL WERE REMOVED USING A COOK BASKET WITH MODEL NUMBER UNKNOWN. THE INTENDED PROCEDURE WAS COMPLETED USING ANOTHER SHEATHS WITH MODEL NO.61146BX, LOT NO. 09L1900159. THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. THERE WAS NO USER INJURY REPORTED. THIS REPORT IS RELATED TO REPORTS PATIENT IDENTIFIER (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11667842 ↗
  • MalfunctionApr 14, 2021Break; Break

    DURING A THERAPEUTIC URETEROSCOPY PROCEDURE THE USER ATTEMPTED 3 DIFFERENT ACCESS SHEATHS AND EACH TIME, THE END TIP OF THE INNER INTRODUCER CRUMBLED. FOR THE TWO SHEATHS THE ISSUE WAS FOUND BEFORE IT WAS INSERTED/USED ON THE PATIENT. THE PARTS FELL INSIDE THE PATIENTS¿ KIDNEY WHEN THE THIRD SHEATH WAS USED. THE DEVICE WAS VISUALLY INSPECTED PRIOR TO BEING INSERTED AND THE DEVICE APPEARED TO BE INTACT WHEN IT WAS INSERTED. UNDER THE ENDOSCOPIC IMAGE THE PARTS THAT FELL WERE REMOVED USING A COOK BASKET WITH MODEL NUMBER UNKNOWN. THE INTENDED PROCEDURE WAS COMPLETED USING ANOTHER SHEATHS WITH MODEL NO.61146BX, LOT NO. 09L1900159. THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. THERE WAS NO USER INJURY REPORTED. THIS REPORT IS RELATED TO REPORTS PATIENT IDENTIFIERS (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11667814 ↗

FDA clearance and classification

Cleared through 510(k) K051593 on Sep 2, 2005 (decision: Substantially Equivalent). Submitted by Acmi Corporation as “Uropass Ureteral Access Sheath, Model 61224”. The FDA review took 78 days. FDA record ↗

The FDA regulates product code KNY (accessories, catheter, g-u) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K051593. One submission often covers many models, so check the product and lot details in each record.

  • Z-2401-2023 · May 19, 2023 · Open, Classified
    UroPass Ureteral Access Sheaths, 5 pieces/box

Common questions about the UroPass 61138BX

Has the UroPass 61138BX been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0364-2025 (Class II), began on Sep 11, 2024: Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the UroPass 61138BX?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Apr 14, 2021 and Dec 28, 2022. In those reports, the most common specific device problems are “Break” (3), “Material Separation” (1) and “Material Twisted/Bent” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the UroPass 61138BX?

None of the reports are classified as deaths, 1 as an injury and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the UroPass 61138BX FDA approved?

It was cleared through 510(k) K051593 on Sep 2, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the UroPass 61138BX, and is it still sold?

The FDA's device database lists Gyrus ACMI, LLC as the company responsible for it (brand UroPass, model 61138BX). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other ACCESSORIES, CATHETER, G-U devices

All ACCESSORIES, CATHETER, G-U devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page