Skip to content

UroPass 61054BXRecalls & reported problems

Gyrus ACMI, LLC · UroPass Ureteral Access Sheath 10/12 Fr (3.3/4 mm) x 54 cm (5/PK) · Class 2

All models of UroPass (15)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00821925035348), received between Nov 17, 2022 and Mar 6, 2023 — classified as 4 injuries.

The most frequently reported specific device problem is “Break” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K051593); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Sep 11, 2024Class IIOpen, Classified

    Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures.

    Z-0364-2025 · Root cause: Storage · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break3
  • Detachment of Device or Device Component1

Patient problems

  • Pulmonary Edema3
  • Foreign Body In Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 6, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS THAT DURING A RETROGRADE INTRARENAL SURGERY, THREE UROPASS URETERAL ACCESS SHEATH BROKE AT THE DISTAL END OF THE DILATOR INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED BY REPLACING WITH ANOTHER SHEATH HOWEVER IT WAS SHORTER. THE FRAGMENTS THAT FELL INTO THE PATIENT HAVE BEEN REMOVED BUT CAUSED EXTENSION OF THE PROCEDURE FOR ONE AND A HALF HOURS. THE PATIENT WAS UNDER GENERAL ANESTHESIA. THE ANESTHETIST REPORTED TO THE PHYSICIAN THAT THE PATIENT HAD AN EARLY PULMONARY EDEMA AND WAS GIVEN A DIURETIC WHICH SEEMED TO HAVE NOT MADE THE PROBLEM MORE SERIOUS. IT WAS THOUGHT TO BE A CONSEQUENCE OF PATIENT BEING UNDER GENERAL ANESTHESIA FOR A TIME LONGER THAN EXPECTED. THERE WERE NO REPORTS OF FURTHER PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. (B)(6) 1 OF 3 UROPASS URETERAL ACCESS SHEATH. (B)(6) 2 OF 3 UROPASS URETERAL ACCESS SHEATH. (B)(6) 3 OF 3 UROPASS URETERAL ACCESS SHEATH.

    Patient problems: Pulmonary Edema

    MAUDE report 16492576 ↗
  • InjuryMar 6, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS THAT DURING A RETROGRADE INTRARENAL SURGERY, THREE UROPASS URETERAL ACCESS SHEATH BROKE AT THE DISTAL END OF THE DILATOR INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED BY REPLACING WITH ANOTHER SHEATH HOWEVER IT WAS SHORTER. THE FRAGMENTS THAT FELL INTO THE PATIENT HAVE BEEN REMOVED BUT CAUSED EXTENSION OF THE PROCEDURE FOR ONE AND A HALF HOURS. THE PATIENT WAS UNDER GENERAL ANESTHESIA. THE ANESTHETIST REPORTED TO THE PHYSICIAN THAT THE PATIENT HAD AN EARLY PULMONARY EDEMA AND WAS GIVEN A DIURETIC WHICH SEEMED TO HAVE NOT MADE THE PROBLEM MORE SERIOUS. IT WAS THOUGHT TO BE A CONSEQUENCE OF PATIENT BEING UNDER GENERAL ANESTHESIA FOR A TIME LONGER THAN EXPECTED. THERE WERE NO REPORTS OF FURTHER PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. (B)(6) 1 OF 3 UROPASS URETERAL ACCESS SHEATH. (B)(6) 2 OF 3 UROPASS URETERAL ACCESS SHEATH. (B)(6) 3 OF 3 UROPASS URETERAL ACCESS SHEATH.

    Patient problems: Pulmonary Edema

    MAUDE report 16492538 ↗
  • InjuryMar 6, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS THAT DURING A RETROGRADE INTRARENAL SURGERY, THREE UROPASS URETERAL ACCESS SHEATH BROKE AT THE DISTAL END OF THE DILATOR INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED BY REPLACING WITH ANOTHER SHEATH HOWEVER IT WAS SHORTER. THE FRAGMENTS THAT FELL INTO THE PATIENT HAVE BEEN REMOVED BUT CAUSED EXTENSION OF THE PROCEDURE FOR ONE AND A HALF HOURS. THE PATIENT WAS UNDER GENERAL ANESTHESIA. THE ANESTHETIST REPORTED TO THE PHYSICIAN THAT THE PATIENT HAD AN EARLY PULMONARY EDEMA AND WAS GIVEN A DIURETIC WHICH SEEMED TO HAVE NOT MADE THE PROBLEM MORE SERIOUS. IT WAS THOUGHT TO BE A CONSEQUENCE OF PATIENT BEING UNDER GENERAL ANESTHESIA FOR A TIME LONGER THAN EXPECTED. THERE WERE NO REPORTS OF FURTHER PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. (B)(6)1 OF 3 UROPASS URETERAL ACCESS SHEATH. (B)(6) 2 OF 3 UROPASS URETERAL ACCESS SHEATH. (B)(6) 3 OF 3 UROPASS URETERAL ACCESS SHEATH.

    Patient problems: Pulmonary Edema

    MAUDE report 16492483 ↗
  • InjuryNov 17, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTS DURING A PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE USING A UROPASS SHEATH, THE DEVICE BROKE OFF AND FELL INTO THE PATIENT. ADDITIONAL DETAILS REGARDING THE PATIENT AND EVENT HAVE BEEN REQUESTED. AT THIS TIME, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 15818790 ↗

FDA clearance and classification

Cleared through 510(k) K051593 on Sep 2, 2005 (decision: Substantially Equivalent). Submitted by Acmi Corporation as “Uropass Ureteral Access Sheath, Model 61224”. The FDA review took 78 days. FDA record ↗

The FDA regulates product code KNY (accessories, catheter, g-u) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K051593. One submission often covers many models, so check the product and lot details in each record.

  • Z-2401-2023 · May 19, 2023 · Open, Classified
    UroPass Ureteral Access Sheaths, 5 pieces/box

Common questions about the UroPass 61054BX

Has the UroPass 61054BX been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0364-2025 (Class II), began on Sep 11, 2024: Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the UroPass 61054BX?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Nov 17, 2022 and Mar 6, 2023. In those reports, the most common specific device problems are “Break” (3) and “Detachment of Device or Device Component” (1); the most common patient problems are “Pulmonary Edema” (3) and “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the UroPass 61054BX?

None of the reports are classified as deaths, 4 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the UroPass 61054BX FDA approved?

It was cleared through 510(k) K051593 on Sep 2, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the UroPass 61054BX, and is it still sold?

The FDA's device database lists Gyrus ACMI, LLC as the company responsible for it (brand UroPass, model 61054BX). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other ACCESSORIES, CATHETER, G-U devices

All ACCESSORIES, CATHETER, G-U devices →

Sources and updates