Skip to content

USA Elite RBRecalls & reported problems

Gyrus ACMI, LLC · ACMI USA Elite System and USA Series Roller Bar Electrode 24/26 Fr (8/8.667 mm) (6/PK) · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00821925036444), received between Nov 17, 2023 and Nov 17, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Sparking” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

6 other recalls cover devices cleared under the same FDA submission (K890328); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2023Open, Classified

    Incorrect pouch labels, the contents within the mislabeled pouch were Roller Bar electrodes.

    Z-0568-2024 · Root cause: Labeling Change Control · FDA record ↗

Most reported problems

Device problems

  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 17, 2023Sparking; Sparking

    THE CUSTOMER REPORTED TO OLYMPUS THAT THE ROLLER BAR (6/PK) HAD A SPARK/LIGHT AND THEN SMELLED DIFFERENT. THE EVENT OCCURRED DURING A PROCEDURE. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18161150 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K890328. One submission often covers many models, so check the product and lot details in each record.

  • Z-3299-2026 · Aug 24, 2026 · Open, Classified
    Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.
  • Z-3300-2026 · Aug 24, 2026 · Open, Classified
    Olympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.
  • Z-2720-2026 · May 5, 2026 · Open, Classified
    Olympus Ellik Evacuator Includes Glass Body, Male Connecting Tubing & Latex Bulb. Model Number: 194.
  • Z-2139-2025 · Jun 18, 2025 · Open, Classified
    Blue Silicone, Reusable Silicone Seal 6 - 12 FR. Model Number: CS-B612. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the working channel. Some
  • Z-2140-2025 · Jun 18, 2025 · Open, Classified
    Green Silicone Seal for Accessories, Reusable Silicone Seal 0 - 7 FR. Model Number: CS-G7. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the work
  • Z-1952-2018 · Apr 17, 2018 · Terminated
    Gyrus ACMI Reusable Active Cord for Bovie/ValleyLab Generators, Catalog No C650-129A (00821925008700) The cord consists of a single conductor stranded wire terminated with plugs t

Other ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL devices

All ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.